the calibration percentage. LHS counts will increase w.r.t to
0000 => 0000 d
increment of the mV from Tare mV. After study state of counts enter the “d” counts in RHS w.r.t to calibration
CALIBRATION LOAD 2756 => 15000 d
Keep the test weight on the bin
percentage. Press
to complete the calibration calibrati on & close the
calibration menu.
Note: 1% calibration = 20 d counts
SYSTEM DATA MAINPROGRAM
ISSUE: 1.0 - 10/03/17
Press
to go to main program.
Press
to run the main program while main program menu
in bottom line.
mV CALIBRATION PROCEDURE FOR TUC-6 (BLMS)
TUC DISPLAY
ACCESS KEYS
DESCRIPTION
X
= 00,00 t
Press
L
= 63,56 %
program.
&
Press
to enter the OFF LINE menu from the main
to enter the PASSWORD menu
Password
Password Enter password :
Enter 4 digit password which is stored in the configuration and press
to access the OFF LINE menu
Password Enter assword : Y
If TUC shows “Y” the password is correct. Password is wrong it shows “N”. If password is wrong again go to password menu to enter the correct password. If password is correct press to go to CALIBRATION menu.
Analog Input Calibration
CALIBRATION_ MENU RE-CALIBRATE? ->
TARE MV 2.125
SPAN MV 16.125
SYSTEM DATA MAINPROGRAM
ISSUE: 1.0 - 10/03/17
Press
to proceed the calibration. If the calibration menu in
bottom
PRESS 1 2 3 & press
to enter the calculated mV as a
calibration.
Enter the Tare mV, which is generated by the load cell w.r.t bin dead weight. Press
proceed the span mV calibration.
Enter the calculated the span mV as per calculation table. Press