Multi-Slice CT Scanner System Brand Type Application Origin
Neusoft Neuviz 64i Applicable for imaging diagnostic incl. Cardiac Analysis CHINA
Standard Configuration Components Gantry
Qty. 1
Low voltage Slip-ring Slip-ring rotating gantry 50 KW High Voltage Generator Dual focus oil-cooled Tube High sensitivity sensitivity GOS Solid-state Detector Patient Table Operating Console 19" LCD Monitor Neuviz 64 software Cardiac Scan System (software) ClearView Iterative Recon Workstation (Hardware) (Hardware) Workstation (Software) 2D/3D Viewer MPR Viewer CPR Viewer MIP/MinIP Volume Rendering Vessel Analysis Calcium Scoring Cardiac Function Analysis Coronary Analysis Tumor Evaluation Report 3D Software Package Accessories Operation Manual Main AC Cable Computer Table 64 Slice Cradle Cushion NMS Head Holder Assembly Head Holder Cushion Mindray PM-9000 Express Connection Cable Tool Box UPS (console)
1 1
1 1
1 1
Technical Specification Item
S p ecification
I Compone Component nt Perfor Performa mance nce 1 Gantr Gantry y Gantry aperture Scan Field Gantry tilt R otation Time (360°rotation) Partial Scan Time (240°) Focus-to-isocenter Distance Focus-to-detector D istance
70 cm 50 cm ± 30° 0.5s, 0.6s, 0.8s, 1s, 1.5s, 2s 0.32s, 0.39s, 0.52s, 0.65s, 0. 97s, 1.3s 570 mm 1040 mm
Multi-Slice CT Scanner System Brand Type Application Origin 2 Data Acquisiti Acquisition on System Max. number of slices/rotation Number of Detector Rows Number of detector electronic channels (DAS) utilized for 64 slices/rotation acquisition Number of Detector Elements Total Channels per Slice Number of Projections Sequence Acquisition Modes Spiral Acquisition Modes
Neusoft Neuviz 64i Applicable for imaging diagnostic incl. Cardiac Analysis CHINA
64 32 64 672 x 32 1344 4640 64x0.625, 32x0.625, 16x0. 625, 8x0. 625, 4x0. 625, 2x0. 625 64x0.625, 32x0.625, 16x0.625
3 X-ray X-ray Tube Assembly Assembly Tube Tube Current Rang e Tube Voltag e Tube Anode H eat Storage Capacity Cooling Rate Focal Spot Size
CTR 2250 30mA ~ 420mA 80 kV, 100 kV, 120 kV, 140 kV 5.0 MHU 815 KHU/min 0.6 x 1.2 (small) / 1.1 x 1.2 (larg e)
4 Filte Filterr System Equivalent B eam Limiting Device
Al Equivalent Tube: 1.5mm Al Equivalent to 6.68mm Al
5 X-Ray X-Ray Generator Generator Max. Power
50 kW
6 Patie Patient nt table Max. table load Table feed speed Vertical table / Travel range Vertical travel speed Scannable range
205 kg/ 452 lbs 1 mm/s - 160 mm/s 430 mm - 970 mm 9 mm/s - 15 mm/s 1750 mm
7 Host computer computer system system The host computer workplace provides an intelligent and reliable workflow for data acquisition, image reconstruction, and routine post processing at the CT scanner. High-performance Computer Host : 1 x Quad Core Intel (R) Xeon (R) Recon : 1 x Quad Core Intel Xeon 2.40 GHz processor Standard Monitor Flat Screen Monitor 19" (48 cm) 1,280 x 1, 024 Resolution 1,024 x 1,024 Image Display Matrix 0.29 mm Pixel Size R AM Storag e 36 GB I mage Storage 250 + 500 GB; 960, 000 Uncompressed I mages Additional Additional Storage CD-R 700 MB 1,100 Images DVD DICOM Drive 4.7 GB DVD Media 8.400 Images Write-RW/ +RW/ -DL/ Read DICOM Viewer Included on each CD; Automatically started on the viewer's PC
Multi-Slice CT Scanner System Brand Type Application Origin
Neusoft Neuviz 64i Applicable for imaging diagnostic incl. Cardiac Analysis CHINA
8 AVW AVW Workplace Workplace system AVW workplace provides the unique advantage of an efficient multi-modality diagnostic workflow at a single workplace. It manages the clinical diagnostic workflow anywhere within the clinical environment. High-performance computer Dell Precision Standard monitor Flat Screen Monitor 19" R AM storage > 16 GB I mage storag e > 700 GB ; 1, 400, 000 Uncompressed images Additional Additional storage CD-R 700 MB 1,100 Images DVD DICOM Drive 4.7 GB DVD Media 8,4000 Images Write-RW/ +RW/ - DL/ Read DICOM Viewer Included on each CD; Automatically started on the viewer's PC
II System System Perf Perfor orman mance ce 1 System Softwa Software re Protocols Topogram Leng th Scan times Views R eal-time topog ram Sequence Acquisition Acquisition R econstructed slice widths Dynamic Multi-Scan:
Multi-slice Spiral Acquisition Reconstructed Slice Widths Slice Increment Spiral Scan Time Scan Length Pitch Factor Automatic clustering of scans.
Up to 10, 000 protocols can be edited, modified, and stored
50 - 1700 mm 1 . 5 - 18 s A.P., Lateral, Dual Yes
0.625, 1. 25, 2.5, 5, 10 mm Multiple (continuous) sequence scanning wi without ta table movement fo for fast dynamic contrast studies with maximum slice thickness of 20 mm
0.625, 0.8, 1, 1.25, 1.5, 2, 2.5, 3, 4, 5, 6, 7, 8, 9, 10 mm 0.1 – 20 mm Max. 100 s Max. 1700 mm 0.13 - 2.0
2 Image Image reconstruction reconstruction R eal-Time Display Scan Field R econ Field R econ Time R econ Matrix HU Scale
Real-time imag e display during spiral acquisition. 50 cm 5 – 50 cm Up to 20 images/s with full cone beam reconstruction 512x512, 768x768, 1024x1024 – 3, 2768 to + 3, 2767
3 CINE CINE Display Display Max. Image Rate
30 frames/s frames/s
4 Image transfer/N transfer/Networki etworking ng Interface for transfer of medical images and information using the DICOM standard. 5 Raw Raw Data Data Capacity
800 GB
Multi-Slice CT Scanner System Brand Type Application Origin
Neusoft Neuviz 64i Applicable for imaging diagnostic incl. Cardiac Analysis CHINA
6 Applica Applicatio tion n SSD (Surface Shaded Display) Volume Rendering (VR) Maximum Intensity Projection (MIP) Minimum Intensity Projection (MinIP) Average I ntensity Projection Projection (AIP) Multi-Planar Multi-Planar Reconstruction Reconstruction (MPR) Curve Multi-Planar Reconstruction (CPR) Auto Voice Auto Film CT Angiography (CTA), Virtual Endoscopy (VE), Pediatric Protocols Networking Bolus Tracking CD Viewer ClearView Remote Service System Volume Calculation Vessel Measurable Analysis Bone Remove 7 Heart Heart Scan System Heart-scan CT with ECG-synchronized true isotropic volume acquisition using prospective ECG triggered or retrospective ECG-gating mode. The ECG signal used for gating the CT images is acquired by an ECG device. The ECG signal is displayed on the ECG device and the scan interface. Temporal resolution can reach down to 65 High/ 83 Low ms. 8 Installat Installation ion Outline Dimension & Weight Gantry Dimensions Gantry Weight Gantry Package Dimensions Couch Dimensions Couch Weight Couch Packag e Dimensions Console Table Dimensions Power Supply Requirements Power Capacity Input Voltage
Voltage Variation 3-phase Unbalance Frequency Frequency Grounding Resistance Min. Area of Scanning room Min. Area of Operating Room Operating Operating Room Recommended Room Size Min. Height of Ceiling Ceiling Temperat ure of Scanning Room Humidity of Scanning room Atmospheric Pressure Temperature of of Tr Transportation an and St Storage H um umidit y of Tra ns nsport at at io ion and St or orag e Running Noise
2244mm (L) x 890mm (W) x 1920mm (H) 1800 Kg 2370 mm (L) x 1030 mm (W) x 2250 mm (H)
2400mm (L) x 610mm (W) x 1055mm (H) 350 Kg 2570mm (L) x 970mm (W) x 1230mm (H) 1400mm (L) x 800mm (W) x 760mm (H)
80 KVA (Low config uration) 380 / 400 VAC 3-phase 4-line Power supply from below opons : 190/ 200/ 208/ 220/ 230/ 240/ 380/ 400/ 415/ 440/ 460/ 480 VAC ± 10% ≤ 5% 50/60Hz ± 1Hz 4Ω (independent grounding system); 1Ω (complex grounding system) 5550 mm × 3650 mm 1700 mm × 3650 mm
Opera ng Room : 3000 mm × 4600 mm Scanning Room : 6000 mm × 4600 mm 2010 mm Scan room 18 ℃ ~ 24 ℃; Control room 18 ℃ ∼ 28 ℃ Scan room 30 % ~ 60 %; Control room 20 % ~ 80 % 70 kPa ~ 106 kPa -20 ℃ ~ +55 ℃ 10 % ~ 90 %, no no-condensing No more than 70 dBA