BS-300 Chemistry hemist ry Ana An alyz ly zer
Service Manual
© 2006 Shenzhen Mindray Bio-medical Electronics Co., Ltd. All rights Reserved. For this Service Manual, the issued Date is 2006-06 (Version: 1.0).
Intellectual Property Statement tatement SHENZHEN MINDRAY MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. LTD. (hereinafter (hereina fter called Mindray) owns the intellectual property rights to this Mindray product and this manual. This manual may refer to information protected by copyrights or patents and does not convey any license under the patent rights of Mindray, nor the rights of others. Mindray does not assume any liability arising out of any infringements of patents or other rights of third parties. Mindray intends to maintain the contents of this manual as confidential information. Disclosure of the information in this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. Release, amendment, reproduction, distribution, rent, adaption and translation of this manual in any manner whatsoever without the written permission of Mindray is strictly forbidden. , , , trademarks or trademarks owned by Mindray in other trademarks that appear in this manual are without the intention of improperly using them. respective owners.
, are the registered China and other countries. All used only for editorial purposes They are the property of their
Respons Respons ibility ibil ity on the Manufacturer Manufacturer Party Party Contents of this manual are subject to changes without prior notice. All information contained in this manual is believed to be correct. Mindray shall not be liable for errors contained herein nor for incidental or consequential damages in connection with the furnishing, performance, or use of this manual. Mindray is responsible for safety, reliability and performance of this product only in the condition that:
all installation operations, expansions, changes, modifications and repairs of this product are conducted by Mindray authorized personnel; the electrical installation of the relevant room complies with the applicable national and local requirements; the product is used in accordance with the instructions for use.
WARNING: It is important for the hospital or organization that employs this equipment to carry out a reasonable service/maintenance plan. Neglect of this may result in machine breakdown or injury of human health.
NOTE: This equipment is to be operated only by medical professionals trained and authorized by Mindray or Mindray-authorized distributors.
I
Warranty THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE. Exemptions Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel. This warranty shall not extend to:
any Mindray product which has been subjected to misuse, negligence or accident; any Mindray product from which Mindray's original serial number tag or product identification markings have been altered or removed; any product of any other manufacturer.
Return Return Policy olic y Return Return Procedure In the event that it becomes necessary to return this product or part of this product to Mindray, the following procedure should be followed: 1
Obtain return authorization: Contact the Mindray Service Department and obtain a Customer Service Authorization (Mindray) number. The Mindray number must appear on the outside of the shipping container. Returned shipments will not be accepted if the Mindray number is not clearly visible. Please provide the model number, serial number, and a brief description of the reason for return.
2
Freight policy: The customer is responsible for freight charges when this product is shipped to Mindray for service (this includes customs charges).
3
Return address: Please send the part(s) or equipment to the address offered by Customer Service department
Company Contact Manufacture:
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Ad dress dr ess::
Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, P.R.China, 518057
Phone:
+86 755 26582479 26582888
Fax:
+86 755 26582500 26582501
II
Preface Who Shoul d Read Read This Manual Manual This manual is written for service professionals authorized by Mindray.
Conventions Used Used i n This Manual Manual Safety Safety Symbols This chart explains the symbols used in this manual. When you see …
Then …
WARNING:
Read the statement following the symbol. The statement is alerting you to an operating hazard that can cause personal injury.
BIOHAZARD:
Read the statement following the symbol. The statement is alerting you to a potentially biohazardous condition.
CAUTION:
Read the statement following the symbol. The statement is alerting you to a possibility of system damage or unreliable results.
NOTE:
Read the statement following the symbol. The statement is alerting you to information that requires your attention.
Graphics All graphics, including screens and printout, are for illustration purpose only and must not be used for any other purposes.
i
Contents
Contents Preface....................................................................................................................................... Pref ace........................................................................................................................................... .... i Who Should Read This Manual .............................................................................................. i Conventions Used in This Manual .......................................................................................... i 1
Specifications....................................................................................................................1-1 1.1 1.2 1.3 1.4
2
System Feature ............................................................................................... ................................................................................................... .... 1-1 Loading System Feature .....................................................................................1-1 Analysis System Feature..................................................................................... 1-2 Others ....................................................................................... .................................................................................................................. ........................... 1-2
System Sys tem Ins tall ati on ......................................................................................... ........................................................................................................... .................. 2-1 2.1 2.2
3
System Descriptions ........................................................................................................3-1 3.1
3.2
3.3
3.4 3.5 3.6 4
Main Control Board..............................................................................................4-1 Power Drive Board ..............................................................................................4-2 A/D Conversion Board.........................................................................................4-2 Board .........................................................................................4-2 Reagent Refrigeration Board...............................................................................4-2 Level Detection Boards .......................................................................................4-3 Feeder Connection Board ...................................................................................4-3 Manipulator Connection Board............................................................................4-3 Probes Connection Board ...................................................................................4-4 Power Supply Assembly ......................................................................................4-4
Maintenance and Service.................................................................................................5-1 5.1 5.2 5.3 5.4 5.5
6
Dispensing System..............................................................................................3-1 3.1.1 Probe assemblies.................................................................................3-1 3.1.2 Disk assemblies ...................................................................................3-2 Feeder .................................................................................................................3-3 3.2.1 Feeder assemblies ...............................................................................3-3 3.2.2 Manipulator...........................................................................................3-4 Temperature Control System...............................................................................3-5 3.3.1 Temperature control assembly ............................................................. 3-5 3.3.2 Reagent preheating..............................................................................3-6 3.3.3 Reagent refrigeration............................................................................3-7 Photometric System ............................................................................................3-8 Fluid System ....................................................................................... ........................................................................................................ ................. 3-8 ISE Module (optional) ........................................................................................3-10
Funct Fun ct ions io ns of Board Bo ards s .......................................................................................... ......................................................................................................... ............... 4-1 4.1 4.2 4.3 4.4 4.5 4.6 4.7 4.8 4.9
5
Check before Installation .....................................................................................2-1 Installation Procedure..........................................................................................2-1
Replacing Light Filter Assembly .......................................................................... 5-1 Replacing Optical Fiber .......................................................................................5-4 Adjusting Reaction Disk, Manipulator and Feeder .............................................. 5-7 Adjusting Probes and Disks.................................................................................5-8 Replacing Components of ISE Unit (optional)...................................................5-10 5.5.1 Replacing Tubing................................................................................5-10 5.5.2 Replacing Pumps ...............................................................................5-13 5.5.3 Replacing ISE Module........................................................................5-13
Software
Introduction.......................................................................................................6-1
1
Contents 6.1
6.2 7
System Software..................................................................................................6-1 6.1.1 System initialization..............................................................................6-1 6.1.2 Shutdown processing ...........................................................................6-2 Control Software ................................................................................................. ................................................................................................... 6-2
Servic Serv ic e Flow ......................................................................................... ...................................................................................................................... ............................. 7-1 7.1 7.2 7.3 7.4 7.5
7.6 7.7 7.8 7.9
7.10
Fluid Level Detection Failure of Reagent Probe..................................................7-1 Fluid Level Detection Failure of Sample Probe...................................................7-2 Liquid Dropping From Probes..............................................................................7-3 Failing to Detect Level of Water for Washing Exteriors.......................................7-4 Abnormal Results ........................................................................................ ................................................................................................ ........ 7-5 7.5.1 All Results Being Abnormal .................................................................. 7-5 7.5.2 Some Results Being Abnormal.............................................................7-5 7.5.3 Several Results Being Abnormal..........................................................7-6 Insufficient Light Intensity of Lamp ...................................................................... 7-7 Temperature Control Failure................................................................................7-8 Bar Code Scanner (optional) Failure...................................................................7-9 Feeder Failure .................................................................................................. .................................................................................................... 7-11 7.9.1 Transducer Distribution of the Feeder................................................ 7-11 7.9.2 Feeder Failure ....................................................................................7-12 7.9.3 Manipulator Failure.............................................................................7-13 Troubleshooting of ISE Unit (optional)...............................................................7-13
8
Mechanical Structure .......................................................................................................8-1
9
Tools and Parts .................................................................................................................9-1 9.1 9.2
10
Service Tools ....................................................................................................... ....................................................................................................... 9-1 Parts ....................................................................................................................9-1
Maintenance And Test Software....................................................................................10-1 10.1 10.2
10.3 10.4 10.5
General..............................................................................................................10-1 Command ........................................................................................ .......................................................................................................... .................. 10-3 10.2.1 Single Command Area .......................................................................10-3 10.2.2 Macroinstruction Area.......................................................................10-10 PARA and Speed Speed .............................................................................................10-13 Temperature.....................................................................................................10-19 Photoelectric....................................................................................................10-20
Ap pend ix A Board Bo ards s Co nnect nn ect io n Diagr Di agrams ams .............................................................................. A-1 Ap pend ix B Test Po ints in ts of Board Bo ards s ............................................................................................ B-1
2
Specifications
1
1.1 1.1
Specifications
System Syst em Featur Feature e
Dimension: 980mm × 710mm × 1200mm (W × D × H)
Weight: 175kg
Power supply: AC100-130V ± 10% or 200-240V ± 10%
Input power: 1000VA
System: optional, multi-channel, multi-channel, multi-test
Scope: Clinical chemistries
Test types: end-point, kinetic and fixed-time. All support double-reagent and double-wavelength
Calibration type: Linear (single-point, two-point and multi-point), Logistic 4P, Logistic 5P, Exponential 5P, Polynominal 5P, Parabola, Spline −
1.2 1.2
X -R,
QC rules: Westgard multi-rule, Twin-plot
Cumulative sum check, Cumulative error,
Tests analyzed simultaneously: 48 (single-reagent) / 24 (double-reagent); if the ISE unit (optional) is connected, 3 (Na, K and Cl included) or 4 (Na, K, Cl and Li included) tests are added
Throughput: maximum 300tests/h; if the ISE (optional) is connected, maximum 420tests/h (Na, K and Cl included) or maximum 480 tests/h (Na, K, Cl and Li included)
Loading Loadi ng System Feature Feature
Sample volume: 3µl~45µl; Precision: 0.5µl; for the ISE (optional), 70µl serum, 70µl plasma, 140µl diluted urine
Sample disk: general sample disk, including the inner circle and the outer circle
Sample tube position: 60 positions, including 6 calibrator positions, 3 control positions, 5 for STAT sample positions; 5 virtual disks for maximum 300 samples
Sample probe: with a built-in level detector; equipped with auto safeguard; capable of tracking sample level
Washing function: automatically washing interior and exterior of sample probe; carryover no more than 0.1%
Pre-dilution: Pre-dilution: 4 ≤ dilution rate ≤ 150, taking reaction cuvettes as the container
Reagent volume: 30-450ul; Precision: 1ul
Reagent disk: including the inner circle and the outer circle
Reagent position number: 25/50 reagent positions. Each reagent position is available for containing one Hitachi 7060 bottle, one Hitachi 7170 bottle, one Mindray inner-circle bottle or one Mindray outer-circle bottle.
Reagent probe: One independent probe which has a built-in level detector; is equipped with auto safeguard and capable of tracking reagent level
Washing Washing function: automatically washing interiors and exteriors of reagent probes; carryover no more than 0.1%
1-1 1-
Specifications
1.3 1.3
Analysis Analy sis System Feature Feature
Lamp house: 50w lamp
Light splitting mode: Splitting by optical fiber, filtering by an interference filter.
Half band-width: 10±2nm
Wavelength: 340, 405, 450, 510, 546, 578, 630, 670, 700nm
1.4
Mixing bar: for single-reagent tests, it functions after sample dispensing; for double-reagent tests, it functions after the dispensing of the sample and the second reagent.
Absorbance range: range: -0.1~5, -0.1~5, 10mm optical optical path conversion conversion
Reaction cuvette: 5 × 6 × 25mm, optical path 5mm. Material: PP, disposable. Volume: 750uL
Reaction liquid volume: 180-500µl
Max. reaction time: 20 minutes
Reaction temperature: 37±0.3 ℃ with fluctuation of ±0.1℃
Others
Operating system: Windows 2000 or Windows XP
Display: optional
System interface: RS-232
Printer: optional
Built-in bar code scanner: optional
ISE module: optional
1-2
System Installation
2 System Installation NOTE: The analyzer should be installed or moved to another place by Mindray-authorized personnel only.
2.1
Check before Installation The user should provide the environment that meets the requirements mentioned in the Operation Manual. Check if the environment meets the requirements before installing the analyzer. Refer to the chapter 2 of the Operation Manual for details.
2.2
Installation Procedure 1
Ensure available installation fields and environments in hospitals.
2
Confirm the reagents and calibrators.
3
Go to the installation field, and then check the delivery list for acceptance.
4
Install the four handles on the four angles of the analyzer. Move the analyzer to the installation field, fix the casters, and then remove the handles.
5
Insta ll the computer, display and printer.
6
Open the front plate, and check whether cable connections are loose. Open the top plate, check whether the probe assemblies, reagent disk and sample disk are intact and in good performance.
7
Connect the communication cable, power cable, grounding wire, waste tank and deionized water tank. Install the used-cuvette bucket, reagent probe, sample probe and mixing bar.
8
Top up the deionized water tank with deionized water.
9
Put reaction cuvettes in the feeder. Remember to check whether the surfaces of the cuvettes are smooth. In case of any bump, remove it before loading the cuvette to the compartment. Do not touch the light transmission part of the cuvette in which the colorimetric reading is taken.
10
Load acid and alkaline detergents to positions 46 and 47, and distilled water to positions 49 on the reagent disk. Load distilled water to position 60 on the sample disk.
11
Switch on the analyzer as follows: POWER → ANALYZING UNIT POWER →display → computer → printer.
12
After Windows is started, started, double-click the icon of BS-300 on the desktop to start the system software. The system program will automatically finish the self-test, become online and warm up the reaction cuvettes within about 30 minutes.
13
Select the [System/Status] System/Status] menu, and then observe the system status and record it in the table below:
2-1
System Installation Feeder Feeder status status
Unconne Unconnecte cted d
Reaction temperature
disk
Reagent temperature
disk
Full Full
Half Half full
Empty
Ambient temperature Waste tank status
Abnormal (full)
Normal (not full)
Detergent status
Abnormal (empty)
Normal (available)
Printer status
No printer
Normal
Main Main contr control ol unit unit
Unco Unconn nnect ected ed
Reac Reacti tion on disk disk unit unit
Unco Unconn nnec ecte ted d
Reag Reagen entt disk disk unit unit
Unco Unconn nnec ecte ted d
Samp Sample le disk disk unit unit
Unco Unconn nnec ecte ted d
Loadi Loading/ ng/unl unloa oadi ding ng unit unit
Unco Unconn nnect ected ed
Temperature unit
Unconnected
control
Idle Idle
Runni Running ng
Lamp Netw Networ ork k conn connec ecti tion on
Unco Unconn nnec ecte ted d
Wavelength
Dark current
Light source base
340 405 450 510 546 578 630 670 700 Refe eferenc ence lilight sy system 14
Select the [System/Maintenance] System/Maintenance] menu. Then select the Motion Motion tab page, and implement all sub-steps of each unit to see whether they are normal. In case of any exception, adjust it.
15
Wash the interiors and exteriors of the sample probe, reagent probe and mixing bar for several times to make the fluid circuit filled.
16
Set parameters for test, reagent and calibration under theParameters theParameters menu.
2-2
System Installation 17
Request for calibration and samples, run and then debug the results.
18
After debugging the results, fill them in the table below: Test est
ALT
CREA REA
BUN BUN
Target value 2sd range Test value 1 Test value 2 Test value 3 Test value 4 Test value 5 Test value 6 Test value 7 Test value 8 Test value 9 Test value 10 19
Training Can the user complete daily tests?
Yes □
No □
Is the user familiar with the analytical methods such as kinetic, two-point, endpoint?
Yes □
No □
Is the user familiar with the daily, weekly and monthly maintenance and relevant maintenance methods?
Yes □
No □
Is the user skilled in washing dust screens?
Yes □
No □
Is the user skilled in cleaning and replacing the probes and the mixing bar?
Yes □
No □
Is the user skilled in replacing the plunger assemblies of syringes?
Yes □
No □
Is the user skilled in replacing the lamp?
Yes □
No □
Is the user skilled in maintenance of built-in bar code Yes □ scanner?
No □
Is the user skilled in maintenance of ISE unit?
Yes □
No □
Does the user know the positions, roles and Yes □ preparation methods of distilled water and acid and alkaline detergents?
No □
2-3
System Descriptions
3 System Descriptions The BS-300 analyzer consists of the analyzing unit, operation unit and output unit. The analyzing unit consists of the dispensing system, feeder, temperature control system, photometric system and fluid system.
3.1
Dispensing System The dispensing system consists of the probe assemblies (including the reagent probe assembly, sample probe assembly and mixing bar assembly), reagent disk, sample disk and reaction disk.
3.1.1 Probe assemblies Among the probe assemblies, the mixing bar assembly is the same as the reagent probe assembly and the sample probe assembly, except that the knurled axis is 30cm shorter.
Every probe assembly has a horizontal photoelectric switch and a vertical photoelectric switch. These switches are used for defining horizontal and vertical
3-1
System Descriptions initial positions of probe assemblies. The horizontal and vertical step motors precisely control the horizontal and vertical movements of the probe assemblies, and the synchronizing belts serve as the gearing. The shaft and the bushing must corporate with each other precisely, so they cannot be used confusedly.
3.1.2 Disk assemblies The three disk assemblies are different in their coders. The coder corresponds to the position where disks should stop. There is an initial-position mark under every coder. The three coders of the three disks have three coder transducers. Each transducer has two photoelectric switches for inducing the rotation and initial position of the disk.
The step motors control the disk assemblies, and the synchronizing belts serve as the gearing.
3-2
System Descriptions There is a build-in bar code scanner to the left of the sample disk. The scanner is optional.
Build-in bar code scanner
3.2
Sample probe assembly
Feeder The feeder consists of the feeder assemblies and the manipulator. It is designated to send cuvette segments to the reaction disk, take out the used ones and abandon them to the used-cuvette bucket.
3.2.1 Feeder Feeder assembl ies The feeder assemblies include the gearing assembly, cuvette compartment assembly, cuvette-pushing assembly and no-cuvette detection assembly (see the following figure). The supporting plate of the feeder assemblies is a square piece of steel that is connected to the analyzing unit by its four poles, which are secured by four nuts. Unscrewing the nuts, you can disassemble the feeder assemblies from the analyzing unit easily.
3-3
System Descriptions
There are five transducers that are shown in the figure below. The no-cuvette transducer is used to detect whether there is a cuvette segment at the loading position. The insufficient-cuvette transducer is used for determine whether there are less than 10 reaction cuvettes in the compartment or not. If yes, the analyzer will give a prompt.
Pressure Transducer
No-Cuvette Transducer
Cuvette-Pushing Limit Transducer
Insufficient-Cuvette Transducer
Cuvette-taking limit transducer
3.2.2 Manipulator Two step motors (horizontal and vertical) supply power for horizontal and vertical movements of the manipulator. The upper finger and lower finger are same in their structures. They work together to replace used cuvette segments with new ones. The manipulator runs in a relatively complicated way. There are four transducers on it: vertical transducer, transducer, horizontal transducer and two finger transducers.
3-4
System Descriptions
3.3
Temperature Control System
3.3.1 Temperature control assembly The temperature control assembly of the reaction disk consists of the temperature-controlled pot, heat-insulating sheath/plate, top heater, bottom heater, reaction disk/cuvettes, photoelectric seat, temperature transducer, fan and control circuit.
3-5
System Descriptions
1: Temperature transducer and the support 2: Fan 3: photoelectric seat 4: Heat-insulating sheath 5: top heater 6: Cover 7: temperature-controlled pot 8: bottom heater
Upper heater: square in shape, 220/110VAC, 125W Lower heater: ring in shape, 220/110VAC, 350W Total power: pow er: 475W. The function of heaters is to compensate the heat for incubating the reagent and for maintaining the temperature of the temperature-controlled chamber. Fans are used in series in the temperature-controlled chamber. It makes the air circulating in the chamber, and enhances the convective heat exchange. There are four fans in the chamber. All have the alarm function. The temperature transducer feeds back the air temperature at the position several millimeters from the bottom of the reaction cuvette. The overheat protection switch is to switch off the power when the temperature controller does work and the temperature-controlled chamber reaches 55 ℃, so as to avoid overheat or fire. When the temperature-controlled chamber becomes 35 ℃, this switch will automatically be reset.
3.3. 3.3.2 2 Reagent Reagent preheating p reheating The preheating assembly consists of two aluminum plates, a Teflon tube having nine loop sections, heating components, transducer, temperature protection switch, thermal conductive colloid, a section of tube and the reagent probe. The temperature of the thermal source of the preheater is controlled at 45 ℃. The initial temperature of the reagent is 4 ~ 10℃ 10 ℃ when it is taken out of the refrigeration refrigeration
3-6
System Descriptions chamber. When the reagent passes the heater, its temperature increases to 35 ℃. Then the reagent is added into the reaction cuvette and the preheating process is finished.
Reagent preheating assembly
3.3. 3.3.3 3 Reagent Reagent refrig r efrig eration The refrigeration module consists of refrigeration cabin, PU heat-insulating sheath, reagent disk, reagent bottle, temperature transducer, refrigeration flakes, heat sinking component, fan and control circuit. The refrigeration module is shown in the following figure.
The refrigeration assembly consists of fan, hot-end radiator, POM connector, cold-end heat-conductive aluminum block, and PELTIER refrigeration flake. Each analyzer has two such refrigeration assemblies, as shown in the figure below. The cold-side of the refrigeration flake clings to the refrigeration compartment, and the hot-side clings to the radiator (The side having letters should cling to the refrigeration refrigeration aluminum block).
3-7
System Descriptions
Each refrigeration flake corresponds to a heat-sinking block and a cooling fan. It should be installed with the cold side upward.
3.4
Photometric System The photometric system consists of a measurement photometric system and a reference photometric system. The measurement photometric system provides 9 monochromatic lights to measure the absorbance of the reacting liquid in the rotating reaction cuvettes. The reference photometric system compensates the measurement measurement photometric system to make the measurement more accurate.
Tungsten-halogen lamp Biconvex 卤 钨灯 聚 光 镜lens
Plano Main fiber convex v et e杯 tte 小透镜 反u应 传光束 lens C
Plano Fil光 ter片 convex 滤 小透镜 光电管 lens Photodiode
Fibers 传光束小端
共 路
. . .
测量光路 Measurement lights
Photodiode 光 电管 参 考 光 路 light Reference
3.5
Fluid System The fluid system is shown in the following figure.
3-8
System Descriptions
Sample syringe
Reagent syringe Mixing bar
As shown in the figure above, the fluid system consists of interior washing and exterior washing. The syringe assembly controls the aspiration volume by controlling the travel of the sample/reagent sample/reagent syringe. It is the core part of the fluid system.
3-9
System Descriptions
3.6 3.6
ISE ISE Module Modu le (optio (opt ional) nal) +
+
-
+
The ISE module that is used to measure the concentration of K , Na , Cl and Li in serum, plasma and urine consists of ion-selective electrodes, peristaltic pumps and calibrants.
ISE
Calibrant B Pump W
Pump A
Calibrant A
Pump B
Waste
3-10
System Descriptions
3-11
Functions of Boards
4 Functi ons of Boards The analyzer is integrated with the following boards:
Main control board
Power drive board
Sample level detection board
Reagent level detection board
A/D conversion board
Ten photoelectric conversion boards (340nm, 405nm, 450nm, 510nm, 546nm, 578nm, 630nm, 670nm, 700nm and reference light)
Reagent refrigeration board
Power supply assembly
Manipulator connection board
Feeder connection board
Probes connection board PC Power supply assembly PFC board 12V&5V board 24V board ISE power supply board
Level detection boards Transducers
Main control board
Reagent refrigeration board
Power drive board A/D conversion board 10 photoelectric conversion boards
4.1 4.1
Build-in sample bar code scanner
ISE module Step motors DC motors
Heaters of reaction disk
Pumps, valves
Reagent pre-heater
Lamp
Peltiers transducers
Main Main Contro l Board The main control board is the control center of the whole hardware system. It consists of the control circuits of 6 functional units (including main control unit, photoelectric unit, reagent unit, sample unit, loading/mixing unit, temperature control unit). Each functional unit has an MCU. They communicate in the multi-unit mode and thus compose the whole control system. The functions of the main control board are:
communicating with the PC through the RS232, receiving and analyzing instructions from the PC and sending data to the PC.
controlling photoelectric conversion through the interface with the A/D conversion board, reading and saving data from the A/D conversion board.
outputting control signals of each unit through the interface with the power drive board.
4-1
Functions of Boards
4.2 4.2
receiving signals of fluid level detection and bump collision through the interface with the level detection boards.
detecting signals from temperature transducers and controlling temperature of the reaction disk and reagent preheating.
receiving signals from position transducers, deionized water transducer and waste transducer and controlling the transducers.
controlling the built-in sample bar code scanner, reading the data and sending it to the PC.
controlling the ISE module, reading the results and sending them to the PC.
Power Drive Board The main functions of the power drive board are to receive the control signals from the main control board and control drive components. The block diagram of the power drive board is shown in the figure below.
Control signals from Main Control Board
to 13 st step motors
to 2 DC DC motors
to 2 so s olenoid valves
to 3 pumps
to 2 heaters of of reaction disk
4.3 4.3
to lamp power
to 2 magnets to reagent preheating
A/D Conversi on Board The 10 photoelectric conversion circuits convert the intensity signals of the lights transmitting through the reaction cuvettes to electric signals, and then transmit them to the A/D conversion board through a 5-core shielded cable. Photoelectric conversion boards for different wavelengths have different gains and cannot be replaced by each other. The A/D conversion board filters and amplifies the 10 channels of electric signals output from the photoelectric conversion boards, transmits them through the multi-way gating switch to the A/D converter and then sends them to the main control board for further processing.
4.4 4.4
Reagent Reagent Refri Refrigeration geration Board The circuits of the reagent refrigeration board include the refrigeration circuit and the fan circuit. The refrigeration circuit is needed to work continuously, so it is powered separately. The control objects of the reagent refrigeration board include: Reagent refrigeration: 2 PELTIER components, 4 fans.
4-2
Functions of Boards Heat sink system for the whole device: 3 lamp-cooling fans, 4 temperature control fans of the reaction disk.
4.5 4.5
Level Detectio Detectio n Boards The level detection boards that include sample level detection board and reagent level detection board are used to detect the fluid level of sample and reagent separately. When the analyzer aspirates reagents/samples, the probes dip into the liquid to a specific depth, so as to avoid carryovers that have impacts on test results, and to avoid air aspiration when the reagent/sample is insufficient.
4.6 4.6
Feeder Feeder Connecti on Board The feeder connection board transfers the signals between the feeder transducer and the main control board, and connects the power drive board and the loading motor (DC). Connections:
4.7 4.7
Connector Connector
Connected Connected wit h
J91
Front transducer
J92
Back transducer
J93
Intermediate transducer
J94
No-cuvette transducer
J95
Pressure transducer
J96
Motor control wire (connected with the power drive board)
J97
Motor control wire (connected the motors)
J98
Cuvette-loading button
J99
Connection wire of the main control board
Mani Manipul pul ator Connecti on Board The manipulator connection board transfers the signals between the manipulator transducer and the main control board, and connects the power drive board and the electromagnet. Connections: Connector Connector
Connected Connected wit h
J101
Horizontal loading position transducer
J102
Vertical loading position transducer
J103
Electromagnet-closing transducer of the lower hand
4-3
Functions of Boards
4.8 4.8
Connector Connector
Connected Connected wit h
J108
Electromagnet-closing transducer of the upper hand
J104
Drive wire of the electromagnet of the upper hand
J105
Drive wire of the electromagnet of the lower hand
J106
Drive wire of the electromagnet (connected with the power drive board)
J107
Connection wire of the main control board
J109
Safeguard transducer (reserved)
J110
Safeguard transducer (reserved)
Probes Connecti on Board The probes connection board transfers the signals between the sample/reagent level detection board and the main control board, inputs the temperature signals, outputs the reagent preheating signals, and transfers the signals between the power drive board and the mixing motor (DC).
4.9 4.9
Connector Connector
Connected Connected wit h
J200
Interface of the sample level detection board
J201
Interface of the reagent level detection board
J202
Interface of the mixing motor (DC)
J203
Sample detection signal interface of the main control board
J204
Reagent detection signal interface of the main control board
J205
Temperature control signal interface of the reaction disk
J206
Reagent preheating signal interface
J207
Interface of the power drive board
Power Supply Suppl y Ass embly The power supply assembly consists of three boards: PFC board, 24V board, 12V&5V board and ISE (optional) power supply board. The functions of the PFC board include:
AC/DC conversion;
Supplying the +390V and VDD voltage to the 24V board and the 12V&5V board;
Supplying stable 12V voltage for the lamp;
Supplying control signals of the analyzing unit switch to control the C12V, 5V and 24V outputs.
The 24V board converts the 390VDC current transmitted from the PFC board to the separated 24VDC outputs through the forward converter.
4-4
Functions of Boards The 12V&5V board converts the 390VDC voltage from the PFC board to B12V, C12V and 5V voltages through the forward converter. The ISE power supply board converts the 12V of the reagent refrigeration board to 24V that provides power supply for the ISE unit (optional).
4-5
Maintenance and Service
5 Main Maintenance tenance and Servi Service ce WARNING: Before disassembling or assembling the analyzing unit, ensure the POWER is placed to OFF. The probe tip is sharp and can cause puncture wounds. To prevent injury, exercise caution when working around the probe.
BIOHAZARD: Wear gloves and lab coat and, if necessary, goggles. Dispose of the waste in accordance with your local or national guidelines for biohazard waste disposal.
CAUTION: Please use Mindray-recommended consumables. Other consumables may decrease the system performance.
Refer to the BS-300 Chemistry Analyzer Operation Manual for details about
5.1 5.1
unclogging the sample probe
unclogging the reagent probe
replacing the sample probe
replacing the reagent probe
replacing the mixing bar
replacing the plunger assembly of syringe
replacing the lamp
Replacing Replacing Ligh t Filter Filter Assembly The light filter and optical assembly are fixed in the supporting sleeve. The back end is compacted and enclosed with the photoelectric amplification board and the screen gland. Generally, the supporting sleeve is replaced together with the filter and optical assembly.
5-1
Maintenance and Service Figure 5-1 Light filter assembly
Lens Screw Lens Seat Gasket Filter
Gasket
Shield Box
Photoelectric Conversion Board
Flat Spring Supporting Sleeve
WARNING: Before operating, ensure to place the POWER (main switch) to OFF.
1
Unscrew (counter clockwise) the two cap screws on the screen gland whose wavelength is to be replaced.
2
Open the cover of the A/D conversion board, and unplug the plug corresponding to certain wavelength.
5-2
Maintenance and Service 3
Take out the photoelectric conversion board and the supporting sleeve.
4
Keep the photoelectric conversion board upward, and loosen the two retaining screws on it.
5
Keep the photoelectric conversion board upward, and pull the photoelectric amplification board out of the supporting sleeve.
6
Unpack the new supporting sleeve containing the optical assembly. assembly. Be sure to keep the top side (where to assemble the photoelectric amplification board) of the supporting sleeve upward.
7
Install the original photoelectric conversion board onto the new supporting sleeve, and then fasten the two retaining screws.
8
Install the screen gland, and fasten the retaining screws.
5-3
Maintenance and Service 9
Connect the photoelectric conversion board to the A/D conversion board, and assemble the cover.
WARNING: When replacing the light filter assembly, do not touch the optical assembly in the supporting sleeve and the photoelectric receiving tube of the photoelectric conversion board by hand. The light filters and the photoelectric conversion boards are in one-to-one relationship. Do not disarrange them.
5.2 5.2
Replaci Replacing ng Optical Fiber WARNING: Before operating, ensure to place the POWER (main switch) to OFF.
1
Unscrew the four screws on the supporting pillars of the cuvette feeder, and remove the cuvette feeder.
5-4
Maintenance and Service 2
Unscrew the four screws on the reaction disk cover, and open the reaction disk cover. Attention should be paid to the power cable ca ble of the upper heater when wh en the reaction disk cover is being opened. Power cable of upper heater
3
Take out two cuvette segments in a diagonal with the needle-nose pliers to make two spaces for disassemble the colorimetric disk.
4
Unscrew three M3 cap screws, and then disassemble the colorimetric disk.
5-5
Maintenance and Service 5
Use an M3 cap screwdriver to loosen the retaining screws of the optical fibers on the colorimetric clamp and the reference light support.
6
Take out the optical fibers one by one, and fix the optical components in the colorimetric clamp by fastening the screws slightly. Screw of the reference light
7
Draw out all the nine optical fibers from the reaction compartment.
8
Loosen the M3 cap screw (used for retaining optical fibers) on the lamp housing, and then draw out the optical fibers.
9
Put nine of the ten branches of the new optical fiber into the reaction disk from its bottom one by one, loosen the retaining screw, insert the optical fiber to the end, and then fasten the retaining screw.
10
Fix the reference light optical fiber.
5-6
Maintenance and Service 11
Fix the optical fiber of the lamp housing.
12
Put the colorimetric disk back and fix it.
13
Put the reaction disk cover back and fix it.
14
Install the cuvette feeder and fix it.
CAUTION: When replacing the optical fiber, ensure that its bending radius is no less than 20cm. Otherwise, the optical fiber will be damaged.
5.3 5.3
Adj ust ing Reactio Reaction n Disk, Manip Manip ulator ulato r and Feeder Feeder NOTE: Debug the lower arm first (The relation between the lower arm position and the reaction cuvette position is very important.), and the n the upper arm. When debugging the lower arm, move the cuvette compartment aside.
1
Disassemble the cuvette feeder, and adjust the circular position of the reaction disk (through the initial position transducer of the reaction disk) to the standard position (The finger of the lower arm points to the center of the slot between reaction cuvette segments).
2
Horizontally adjust the manipulator to the standard position (The finger of the lower arm can work on reaction cuvettes well, and it should be 0.2mm away from the nearest point of the reaction cuvette).
5-7
Maintenance and Service
5.4 5.4
3
Vertically adjust the manipulator the standard position (The finger of the lower arm can work on reaction cuvette well, and the finger support of the lower arm should be 0.15mm above the reaction cuvette.).
4
Run the manipulator to the position for taking new cuvettes, and horizontally adjust the reaction cuvettes in the feeder. When catching a cuvette, ensure a clearance of 0.4mm ~ 0.6mm between the finger of the upper arm and the cuvette, and center them.
Adju sting Probes and Disks NOTE: Adjust the positions of the three probes and reaction cuvettes, and then others.
1
Adjust the working position of the reagent probe (To (To minimize the cumulative error, ensure that the reagent probe return to the initial position before each adjustment.) as follows: A
For the reagent discharging position, ensure the th e reagent reagen t probe is in the center of the reaction cuvette and the probe tin is 2 ~ 3mm away from the bottom of the reaction cuvette.
B
Adjust the position of the initial-position transducer of the reagent disk. Ensure the mouths of reagent cuvettes in the inner and outer circles fit the reagent disk cover well.
C
Adjust the washing position of the reagent probe. Ensure the reagent probe is in the center of the wash well and the probe pin is 5mm away from the bottom of the wash well. If necessary, adjust the position of the wash well.
D
Adjust the reagent probe’s position on the outer circle of the reagent disk. Ensure the reagent probe is in the center of the hole of the outer circle.
5-8
Maintenance and Service E
2
3
4
Adjust the reagent probe’s position on the inner circle of the reagent disk. Ensure the reagent probe is in the center of the hole of the inner circle.
Adjust the working position of the sample probe (To (To minimize the cumulative error, ensure that the sample probe return to the initial position before each adjustment.) as follows: A
For the sample discharging position, ensure the sample probe is in the center of the reaction cuvette and the probe tin is 2 ~ 3mm away from the bottom of the reaction cuvette.
B
Adjust the position of the initial-position transducer of the sample disk. Ensure the mouths of sample cuvettes in the inner and outer circles fit the reagent disk cover well.
C
Adjust the washing position of the sample probe. Ensure the sample probe is in the center of the wash well and the probe pin is 5mm away from the bottom of the wash well. If necessary, adjust the position of the wash well.
D
Adjust the sample probe’s position on the outer circle of the sample disk. Ensure the sample probe is in the center of the hole of the outer circle.
E
Adjust the sample probe’s position on the inner circle of the sample disk. Ensure the sample probe is in the center of the hole of the inner circle.
Adjust the working position of the mixing bar. A
Switch off the analyzing unit first, and then rotate the mixing arm ar m to see if its rotation radius is proper. If the mixing bar cannot reach the center of the reaction cuvette, adjust the position of the mixing bar on the mixing arm properly.
B
Disassemble the mixing arm, and then switch on the analyzing unit to run the reaction disk and stop it at any cuvette. Hold the mixing arm by hand, and fix it at the center of the reaction cuvette.
C
Vertically adjust the mixing bar’s position in the reaction cuvette. Ensure the mixing bar tip is 1 ~ 2mm away from the bottom of the reaction cuvette.
D
Adjust the washing position of the mixing bar. Ensure the mixing bar is in the center of the wash well and about 5mm away from the bottom of the wash well. If necessary, adjust the position of the wash well.
In the engineering adjustment software, select the Debug instruction menu, and rotate the reaction disk for one lap. Dive probes and mix the reactant at every position or every several positions to see if the three probes interfere with any reaction cuvette (Command 61). If the mixing bar knocks the side or the bottom of any cuvette, repeat Step 3.
5-9
Maintenance and Service
5.5 5.5
Repl Replacing acing Components Compo nents of ISE ISE Unit (opti onal)
5.5.1 5.5.1 Replaci Replaci ng Tub Tubing ing NOTE: (1) Fluidic tubing can be divided into the following six types. W1 with two adapters connects the ISE module and the pump W. Pump module
ISE module
Reagent module to be placed here
Pump tubing adapter Pump tubing Pump head
Pump B
Pump A
Pump W
Fluidic tubing adapter
W2 connects the reagent module and the pump W. A1 (B1) with one adapter connects the sample entry port and the pump A (pump B). A2 (B2) with two adapters connects reagent module and the pump A (pump B). They are different in length and W1, A2 and B2 are not directional when they are installed. Usually the adapter has not been mounted to the fluidic tubing, so you have to do it by hand. The tubing should be mounted to a short metallic tube on the adapter. For convenience, the connecting end of the tubing can be dipped into hot water for several seconds. (2) Pump tubing is used around a pump head. It has one adapter on each end that makes connection with the fluidic tubing easy.
5-10
Maintenance and Service
5.5.1. 5.5.1.1 1 Replacin g Fluid ic Tubi Tubi ng W1 and W2 1
Place the POWER to OFF. OFF.
2
Open the ISE unit door.
3
If you want to replace the W1, unscrew the screw of the ISE module and take off the cover.
4
Put out the fluidic tubing W2 and insert it into a container such as a cup that is used to contain the waste solution flowing from the W2.
5
Start the analyzing unit and the system software.
6
Enter the ISE Maintenance screen Maintenance screen of the system software.
7
Click the Maintenance Maintenance button several times until there is no solution flowing out from the W2.
8
Place the POWER to OFF. OFF.
9
If you want to replace W1, pull out its two adapters directly from the waste pump tubing adapter and the right angle adapter that is fixed to the compression plate. Right angle adapter
The side matches the adapter of fluid tubing W1 The side with a recess matches the compression plate of the ISE module.
Note that when pulling out the fluidic tubing adapter, in order not to release the right angle adapter, you can hold the right angle adapter with a finger. After that, install a new W1. For W2 exchange, it just needs to replace the W2 with a new one. 10
If you want to replace the W2, replace it with a new new one. Otherwise, insert the W2 back to the reagent module.
11
Place the POWER to ON.
12
Enter the System Maintenance screen Maintenance screen of the system software.
13
Select the Others tab Others tab and click the Download Settings button. Settings button.
14
Enter the ISE Maintenance screen. Maintenance screen.
15
Click the A the A p urge ur ge button button to observe if there is solution leaking out. If there is solution leaking out, repeat from the step 4 to install the tubing again.
16
Install the cover of the ISE module if it has been taken off.
17
Close the ISE unit door.
5-11
Maintenance and Service
5.5.1. 5.5.1.2 2 Replacin g Fluid ic Tubi Tubi ng A1, A2, B1 and and B2 1
Place the POWER to OFF. OFF.
2
Take off the panel under the sample disk and you can see the shielding box and the peristaltic pumps.
3
Open the ISE unit door.
4
Take out the reagent module.
5
Put the fluidic tubing W2 in a container such as a cup that is used to contain the fluid flowing out from the W2.
6
Start Start the analyzing unit and the system software.
7
Enter the ISE Maintenance screen Maintenance screen of the system software.
8
If you want to replace A1 or A2, click the A the A purg pu rge e button. If you want to replace B1 or B2, click the B purge button. purge button. Repeat this step for several times until the received data indicates that the Calibrant A or Calibrant B has air bubble in it.
9
Place the POWER to OFF. OFF.
10
Replace the tubing with a new one. Note that before installing the A1 or B1 to the sample entry port, you can dip the connecting end of the A1 or B1 into hot water for several minutes to make the following procedures easily performed.
11
Start the analyzing unit.
12
If the system software is not running, start it. Otherwise, enter theSystem the System Maintenance Maintenance screen of the system software, select the Others Others tab and click the Download Settings button. Settings button.
13
Enter the ISE Maintenance screen Maintenance screen of the system software.
14
If you have replaced A1 or A2, click the A the A purg pu rge e button. If you have replaced B1 or B2, click the B purge button. purge button. Repeat this step for several times until the received data indicates that the Calibrant A or Calibrant B has no air bubble in it.
5.5.1. 5.5.1.3 3 Replacin g Pump Tubi ng 1
Place the POWER to OFF. OFF.
2
Open the ISE unit door.
3
Pinch one adapter of the pump tubing and take it out from the tubing shelf. Then take off the whole tubing.
Tubing shelf
5-12
Maintenance and Service 4
Put one adapter of the new pump tubing onto the tubing shelf and make the tubing around the pump ahead, then put the other adapter onto the shelf with a little strength.
5
Start Start the analyzing unit and the system software.
6
Enter the ISE Maintenance screen. Maintenance screen.
7
Click the Pumps button Pumps button to see if the calibration of pump is correct. If not, repeat the upper steps.
5.5.2 5.5.2 Repl Replacing acing Pumps 1
Place the POWER to OFF. OFF.
2
Take off the pump tubing around the pump head.
3
Disconnect the cable and the pump motor.
Cable
Pump motor
4
Unscrew the four screws around the pump head.
5
Pull out the pump directly. directly.
6
Put a new pump on the pump shelf, tighten the screws, connect the calbe and pump motor, and then put the pump tubing around the pump head.
7
Start Start the analyzing unit and the system software.
8
Enter the ISE Maintenance screen. Maintenance screen.
9
Click the Pumps button Pumps button to see if the calibration of pump is correct. If not, repeat the upper steps.
5.5.3 Replacing ISE Module Some times there is something wrong with the components in the ISE module such as sample entry port, bubble detector, PCB and so on. In these cases, the ISE module should be replaced according to the following procedure. 1
Place the POWER to OFF. OFF.
2
Take off the panel under the sample disk and you can see the shielding box and the pumps easily. easily.
3
Open the ISE unit door.
4
Start Start the analyzing unit and the system software.
5
Enter the ISE the ISE Maintenance screen. Maintenance screen.
5-13
Maintenance and Service 6
Click the Maintenance button. Maintenance button.
7
Place the POWER to OFF. OFF.
8
Unscrew the screw of the ISE module and take off the cover.
9
Take off all electrodes from above to below.
10
Disconnect the right angle adapter and the compression plate.
11
Disconnect the cable.
12
Unscrew the four screws that fix the ISE module to the shielding box. Then take out the module from the shielding box.
13
Put a new ISE module to the shielding box. Then connect the cable and install the electrodes sequentially. At last mount the right angle adapter to the compression plate.
14
Install the electrodes sequentially.
15
Enter the System Maintenance screen Maintenance screen of the system software.
16
Select the Others tab Others tab and click the Download Settings button.
17
Enter the ISE the ISE Maintenance screen. Maintenance screen.
18
Click the A the A purg pu rge e button to see if there is solution leaking out. If there is, take off the electrodes and install them again.
19
Check if the new ISE module can work normally. normally.
5-14
Software Introduction
6 Software Introduction The software of BS-300 analyzer is composed of the system software and the control software.
6.1 6.1
System Syst em Software Softw are The system software can, according to the requirements and inputs of users, generate a work schedule (instruction sequence), control the units of analyzing unit in the sequence of instructions in the work schedule, receive photoelectric data, response messages or execution results from the analyzing unit, output them to the PC screen or the printer. printer. With these outputs, users can obtain correct test results. According to different different functions, functions, the entire PC software system can be divided divided into the following parts:
System Initialization
This part includes the initialization of the PC operating system, the initialization of the communication communication between the PC and analyzing unit, and the controls of analyzing unit reset.
Control system
This part includes the formation of the work schedule, instruction data sending and receiving.
GUI
This part includes the requests for tests (routine tests, emergency tests, calibration tests and QC tests), the observation of test status (status of the reaction disk, sample disk and reagent disk), test management, calibration management, QC management, result query, alarm management and the help system.
Shutdown Processing
This part includes the resets of sub-units.
6.1.1 System initialization To initialize the PC operating system,
Check the PC operating system. The system software must run under Windows 2000 or Windows XP. Otherwise, the system will prompt that the system software cannot run under other operating system, and then the system software system exits automatically.
Check the current screen resolution for the operating system. The system software must run under the resolution of 1024 × 768. Otherwise, the system prompts to reset the resolution before restarting the system software and exit the system software.
Disable the screen protection program. The system software must keep displayed while running. To prevent the screen protection program from disturbing users in the operating and observing processes, disable it.
Lock the keyboard. When running, the system software will lock some key combinations to prevent users from starting any other program and conducting any
6-1
Software Introduction other operation. In this case, users cannot switch over to any other program or print the current screen.
Check whether there is any username and password of the maintenance/debugging personnel in the registry. The system software has the debugging function. Only the authorized maintenance/debugging personnel can enter the debugging window and maintain or debug the system and the host. For the confidentiality, the username and password are saved in the registry. If they are unavailable in the registry, the system will re-write the default username and password into the registry.
To initialize the communication and the units auto-check,
Set serial ports and initialize them, including such parameters as the baud rate, data bit, start bit, stop bit, parity bit, transmitting/receiving buffer, control protocol and so on. In addition, start the serial port receiving thread.
Handshake for communication. Send a handshake instruction to t he analyzing unit. If the analyzing unit responds to this instruction (namely, send back a handshake instruction to the PC), it indicates the PC handshakes handshakes with the host successfully. successfully. If the host fails to respond to handshake instruction, it will re-send the handshake instruction back in a specific period. If it fails for continuously three times, the system will prompt to exit the system software. If you ignore that and continue to enter the system software, all tests cannot be conducted under the system.
Check whether the printer is connected. If not, the system will prompt to connect it.
Send an instruction for querying the auto-check results of the units, check the auto-check result of each unit. (The units are auto-checked when the analyzing unit is started, and the auto-check results are stored in the main unit.) In case of any fault data in the auto-check result of any unit, the system will prompt to switch off the analyzing unit and check the faulty unit.
6.1. 6.1.2 2 Shutdown pr ocessin g In the BS-300 Chemistry Analyzer Control System window, click the Exit button. Exit button. The Confirm dialog Confirm dialog box appears. If you click OK, OK , the system will
6.2 6.2
Switch off the lamp;
Unload all reaction cuvettes;
Wash the fluid tubing;
Reset all units;
Unlock the keyboard;
Prompt to exit the operating system and shut down the PC.
Contro l Software The analyzing unit can be functionally divided into the following units: photoelectric unit, reaction disk unit, reagent disk unit, sample disk unit, reagent probe syringe unit, sample probe syringe unit, mixing bar unit, loading and manipulator unit, temperature control unit, fluid tubing unit and reagent refrigeration unit. The functions of those units are listed in the following table.
6-2
Software Introduction Unit
Function
Main unit
Receives macroinstructions macroinstructions from the PC, decomposes decomposes them into a series of action instructions, and then delivers the action instructions to destination sub-units sub-units (in specific order) at a certain interval. Monitors the status of other units, and transmits data collected by the photoelectric unit to the PC.
Photoelectric unit
Photoelectrically Photoelectrically detects, amplifies and converts the solution in reaction cuvettes, and stores A/D converted data at the twin port FIFO for the main unit to read and transmit them to the PC.
Reaction unit
disk
Contains 80 cuvette No.
Reagent unit
disk
Sample unit
disk
Runs the reaction cuvette with the specified No., following the instruction of the main unit, to the reagent dispensing position, sample dispensing position, mixing position, and photoelectric detection position at all wavelengths. Contains 50 bottle positions. Carries reagents, and runs the reagent bottles with specified No., following the instruction of the main unit, to the reagent aspirating position. Contains 60 tube positions in the inner and outer circles. Carries samples, and runs the sample tubes with specified No., following the instruction of the main unit, to the sample aspirating position.
Reagent probe syringe unit
Receives instructions from the PC, and controls the reagent probe in aspirating a specific volume of reagent from the reagent bottle and dispensing it to the specified reaction cuvette.
Sample probe syringe unit
Receives instructions from the PC, and controls the sample probe in aspirating a specific volume of sample from the sample tube and dispensing it to the specified cuvette.
Mixing bar unit
Receives instructions from the PC, and controls controls the mixing mixing bar in mixing the solution in the cuvette that has been run to the mixing position. Runs the mixing bar to the wash well, and wash it after each mixing process to avoid carryover.
Loading and manipulator unit
Controls the cuvette feeder and manipulator. The feeder assembly is responsible for detecting whether there are enough reaction cuvettes in the cuvettes compartment and pushing reaction cuvettes to the position for the manipulator to take cuvettes. The manipulator is responsible for taking reaction cuvettes from the reaction disk, placing them into the used cuvette bucket, taking new reaction cuvettes from the cuvette compartment, compartment, and putting them on the reaction disk.
6-3
Software Introduction Unit Temperature control unit
Function Controls the reaction disk temperature, reagent probe preheating temperature and fluid tubing. Note: Note: The reaction reaction disk temp temperat erature ure should should be be kept kept at 37 37 ℃, the heating cavity of the reagent probe should be preheated to 45 ℃.
Fluid tubing unit
Controls the t he fluid tubing tubi ng in washing the t he reagent probe, sample probe and mixing bar.
Reagent refrigeration unit
Refrigerates the reagent chamber temp temper erat atur ure e bet betwe ween en 4 ~ 10 10 ℃.
6-4
and
controls
its
Service Flow
7 Service Flow WARNING: Before disassembling or assembling the analyzing unit, ensure the POWER is placed to OFF. The probe tip is sharp and can cause puncture wounds. To prevent injury, exercise caution when working around the probe. When you disassemble or replace any board, ensure to wear antistatic gloves.
BIOHAZARD: Wear gloves and lab coat and, if necessary, goggles.
7.1 7.1
Fluid Level Level Detectio Detectio n Fail Failure ure of Rea Reagent gent Probe Surface detection signal transmission error: The surface detection board does not have the working voltage, or when the probe touches the surface, the indicator of the surface detection board is normal, but there is always a host alarm indicating surface detection failure. Possible cause: The patch cord for the reagent surface detection is disconnected, or the connector is not well connected.
The reagent probe cannot detect the surface on the reagent disk
Surface detection failure of reagent probe(Th e reagent probe does not detect the surface on the reaction disk.)
Surface detection signal processing error: When the probe touches the surface, the indicator of the surface detection board is normal, the connection between the main control board and the surface detection board is normal, and no surface detection signal reaches the main control board, but there is always a host alarm indicating surface detection failure. Possible cause: Main control board. board. Power failure: The working voltage of the surface detection board is 12V. In case of no power supply, the surface detection signal cannot be generated. Possible cause: Disconnection; the connector is not well connected; there is no 12V output from the main control board. Probe failure: The detection terminal has a changeable capacitance. Loose exterior or interior of the probe, the sealing-off of the probe connecting wire and the probe breakage will result in unsteady voltage or obvious voltage changes. In this case, the capacitance signal will be unavailable or exceed the range of the surface detection board. Possible cause: Reagent probe assembly, connector.
The surface detection signal fails to be generated by capacitance change.
Surface detection board failure: When there is no probe failure, but the indicator is always on, or the indicator is not on when the probe touches the surface, there must be a surface detection board failure. Troubleshooting: Replace the surface detection board.
The reagent probe misdetected the surface
When this failure occurs, the sample probe cannot detect the surface at the reaction disk, but there is no alarm at the reagent unit. For details, see Surface Detection Failure: Sample Probe.
7-1
Service Flow
7.2 7.2
Fluid Level Level Detectio Detectio n Fail Failure ure of Sampl Sampl e Probe
The sample probe cannot detect the surface at the reaction disk.
No reagent in the reaction disk: The reagent probe misdetects the surface, aspirates air and dispenses into the reaction cuvette. Therefore, the reagent unit CPU fails to make any judgment. In this case, the sample probe cannot detect the surface when dispensing the sample into this reaction cuvette. Possible cause: Surface detection failure of the reagent probe insufficient reagent: The reagent probe aspirates less reagent than the set volume. As a result, the sample probe cannot detect the surface at the preset height. Possible cause: The reagent is prepared in the bottle, the reagent syringe is not fixed, leakage occurs in the tubing of the reagent probe, or there are bubbles in the reagent. Position correction parameter error of the sample probe: In case of such an error, the sample probe cannot reach the surface in the steps. Troubleshooting: Re-correct the sample probe position. Surface detection signal transmission failure: When the probe touches the surface and stops, the indicator of the surface detection board is on for about 2 seconds. The output signal of the surface detection board switches normally, but there is always a host alarm indicating that no surface is detected. Possible cause: The patch cord for detecting the sample surface is disconnected, or the connector is not well connected.
Surface detection failure of the sample probe
The sample probe cannot detect the surface at the reaction disk.
The sample probe misdetects the surface (In this case, there will be no alarm, but test results will become abnormal. On the reaction curve, you can see that no sample is added.)
Signal processing failure of the main control board: When the probe touches the surface, the indicator works normally. But there is always a host alarm indicating that no surface is detected. Ensure that the sample surface detection signal is transmitted to the relevant connector of the main control board. Possible cause: Main control board failure Surface detection signal failure: The indicator of the surface detection board is not on or always on when the sample probe touches the surface. Surface mis-detection at the sample disk: In this case, there will be no alarm, but the sample probe fails to aspirate the sample, and the test result and reaction curve become abnormal. Possible cause: Sample probe, surface detection board, surface detection patch cord. Surface mis-detection at the reaction disk: In this case, there will be no alarm, but there is residual sample on the sample probe tip, and drops on the table when the probe leaves the reaction disk. Possible cause: The sub-unit software is later than V1.7, or the sample probe is not in the center of the reaction cuvette.
7-2
Service Flow
7.3 7.3
Liqu id Dropping From Probes Probes Drips drop from the sample probe due to fluid tubing leakage. In this case, there will be liquid dropping onto the table widely, and there may be liquid dropping even when the sample probe is static.
Troubleshootin g for the liquid dropping from the reagent probe and sample probe. Drips drop from the sample probe due to surface detection failures. In this case, most liquid drop on the area between the reaction disk and the wash well.
Tubing leakage due to the abraded syringe piston Troubleshooting: Replace the syringe or the syringe piston
The reagent probe takes liquid from the reaction disk. The reagent probe did not detect the surface at the reaction disk, and dispenses the reagent at the corresponding height. If the reagent probe cannot touch the surface at the position for dispensing reagent, the last reagent drop will be taken out by the reagent probe. Possible cause: The probe arm is installed at a too high position; the position of the sample probe has not been corrected; the syringe is loose; there are bubbles in the reagent bottle.
The sample probe takes liquid from th e reaction disk. The sample probe detects the surface and stops immediately. Then it dispenses the sample at 1mm under the surface. If the reagent probe mis-detects the surface, the dispensed sample will fail to be added into the reagent, but taken at the probe tip and drops when the probe is moving. As a result, some test results become abnormal. Possible cause: There are bubbles in the reaction cuvette; the sample probe is not in the center of the reaction cuvette, and its position has not been corrected.
Drips taken from the reagent disk or sample disk and dropping from probes Such failures are usually followed by alarms indicating that no surface is detected at the reagent/sample disk. The reagent/sample probe does not aspirate reagent/sample, but the reagent/sample probe will stretch out to the bottom of the reagent bottle/sample tube due to surface detection failures. As a result, reagent/ sample liquid will be taken on the exterior of the probe and drop while the probe is moving.
Probe being bent Troubleshooting: Replace the probe.
Smudges on the sample probe absorb some liquid and drop it while the probe is moving. Troubleshooting: Cleaning the sample probe. Other failures Probe driving assemblies are not assembled well, resulting in probe jittering and liquid dropping in the vertical movements. Troubleshooting: Adjust or replace the probe driving assemblies.
7-3
Service Flow
7.4 7.4
Faili Faili ng to Detect Detect Level of Wate Waterr for Washin Washing g Exteriors Surface detection signal failure: See relevant resolutions for the surface detection failure.
No water in the wash water tank: Users can see this failure in the alarm information detergent empty . In case of no alarm information, check whether there is any problem in the detergent tank cover assembly.
Failing to Detect the Surface of the Water for Washing Exteriors
Pump damaged: In this case, there will be no water in the three wash wells. Troubleshooting: Replace the pump. No water in the wash well Tubing blocked: In this case, there will be insufficient water for washing exteriors, the three water flows are not the same, or there will be no water in any of the three wash wells. Possible cause: The unidirectional valve is blocked, or the detergent entrance is blocked.
The tube is disconnected. Possible cause: The tube is degraded, or other causes. Troubleshooting: Connect the tubes again.
Reagent or sample probe. Solution: Replace a probe
7-4
Service Flow
7.5 7.5
Abn orm al Resul Resul ts
7.5. 7.5.1 1 A ll Results Being Abn ormal No sample or insufficient sample is added. In this case, it can be seen on the reaction curve that the absorbance has not increased in the period for adding the sample. The reaction curve is even, and many results are 0. Possible cause: Sample probe blocked, tubing leakage, syringe not fixed, valve not shut tightly.
All results are lower than normal results.
The reaction is not thorough. If the sample is not mixed uniformly with the reagent, the reaction will not be thorough, and the light cannot detect the changes of the absorbance. Possible cause: The mixing bar did not mix the reacting liquid.
All test results are abnormal
Some results are higher, and some are lower, with bad repeatability.
The volumes of dispensed sample are not the same. In this case, it can be seen from the reaction curve that the response is proportional to the volume of the sample. Possible cause: Tubing leakage, valve not shut tightly, or cuvette segments not cleaned well.
7.5.2 Some Results Being Abnormal Good repeatability Troubleshooting: Re-calibrate it.
Bad repeatability Observe the reaction curve, and find the possible causes. Some test results are abnormal. All test results are approaching approaching to 0. 0. Observe the reaction curve first. Possible cause: The lamp base of corresponding wavelength exceeds 65535; the photoelectric detection unit fails to detect the change of absorbance, reagent failure and optical fiber disconnection.
7-5
Service Flow
7.5.3 Several Results Being Abnormal Several test results are too low. In this case, observe the reaction curve first. Possible cause: Drips dropping from the sample probe Several test results are abnormal
Several test results suddenly become too high. In this case, observe the reaction curve first. Possible cause: There are bubbles in the current cuvette; the mixing bar knocks the cuvettes.
7-6
Service Flow
7.6 7.6
Insufficient Ligh t Intensity of Lamp Insufficient Light Intensity of the Lamp
Record the lamp background value and gain
Measure and record the lamp background value and gain
Lamp background values of all channels decreases.
Lamp base values of several channels decreases
Replace lamb
Estimate whether the lamp has worked for 2000h (lifetime), replace Lamb
S E Y
Replace the lamp. Check whether the AD value of the channel whose lamp vase value is smaller increases obviously.
YES
O N
Replace the lamp with the original lamp
Adjust the gain gain parameter to make make the lamp base AD values of all channel lamps increase to about 62000.
Replace the photoelectric transducer assembly, and then check whether the lamp base value increases obviously. YES
Replace the photoelectric transducer assembly.
Remove the fault alarm
Replace the optical fiber
Clean the photometric system (such as the lens)
NO
YES
Change the positions with the smaller fiber end, and check whether the lamp base value increases obviously. O N
YES
Check whether the photometric system (such as the lens) corresponding corresponding to the fiber channel is covered with dirt or the like. (If yes, clean it.) Then check whether the lamp base value increases obviously
Note: Do not adjust the photoelectric gain parameter in case you have no idea about the cause for the alarm indicating insufficient light intensity of the lamp.
7-7
Service Flow
7.7 7.7
Temperatu re Contro Cont roll Fail Failur ure e Temperature control failure
Query and record the fault log codes and descriptions.
Monitor the temperature curve.
Obtain the current temperature control parameter, and check whether it is the same as the factory parameter.
Replace the temperature transducer.
Enable and disable the temperature control function repeatedly for 2 ~ 3 times
NO
Check whether the temperature transducer is normal.
s e y
Replace the heating assembly.
NO NO
Check whether the heating resistance is normal.
S E Y
Restart the analyzer and then check whether it is normal.
1
To monitor the temperature curve A
In the BS-300 Chemistry Analyzer Control System window, select the [System/Temperature [System/Temperature]] menu. A dialog box appears. Enter BS300DEBUG BS300DEBUG and MINDRAY MINDRAY respectively in the User Name Name and Password Password text boxes. Then click OK OK to enter the Temperature Control Curve window. Curve window.
B
Set a proper temperature range for observing the change of the temperature curve, and then click the Start button. Start button.
7-8
Service Flow
2
3
C
Minimize the Temperature Control Curve Curve window for future observation at any time.
D
After that, print or snap the temperature curve and then return it to the headquarters together with the service work report.
To measure the impedance of the temperature transducer A
Switch off the BS-300 analyzer.
B
Open the left side plate of the BS-300 analyzing unit to expose the main control board.
C
Unplug J45 (reagent preheating temperature transducer), and measure the resistance between PIN1 and PIN3 with a multimeter. The relation between the resistance and the temperature is: 25℃ /1100Ω, 38.5℃/1145Ω, 45℃/1170Ω, 50℃/1170Ω.
D
Unplug J44 (reaction disk temperature transducer), and measure the resistance between PIN1 and PIN3 with a multimeter. The relation between the resistance and the temperature is: 25℃/1100Ω, 38.5℃ /1145Ω, 45℃/1170Ω, 50℃/1170Ω.
To measure the resistance of the heater A
Measure the impedance of the reagent preheater in the following procedures: Unplug J24 (power driving board) Measure the impedance between PIN1 and PIN2. Reagent preheater (BA30-10-06626)/24.6 to 28.5Ω.
B
Measure the impedances of the upper and lower heaters of the reaction disk in the following procedures: Disconnect the connecting wire from J113 of the power patching board. Measure the impedances of the upper and lower heater: upper temperature-sensitive heater (BA30-10-06624)/437.8 to 483.8Ω, lower temperature-sensitive heater (BA30-10-06625)/131.4 to 145.2Ω.
7.8 7.8
Bar Code Scanner Scanner (optio nal) Failu Failu re WARNING: Light sent by the bar code scanner may hurt your eyes. Do not stare into the laser beam from the bar code scanner.
If the bar code scanner cannot emit laser, you can service it according to the following steps. 1
Place sample tubes with bar code labels that meet the requirements mentioned in the operation manual.
2
Restart the analyzer.
7-9
Service Flow 3
Enter the Scanner Setup Setup & Maintenance screen Maintenance screen of the system software.
4
Enter the Maintenance tab. Maintenance tab.
5
Enter the sample position range in the Sample Position field Position field and click the Step Scan button. Scan button. Observe if the scanner can emit laser. If not, click the Laser on button. on button. If the scanner still cannot emit laser, go to the next step.
If the bar code scanner emits laser normally but is failed in reading, you can service it according to the following steps. 1
Place the ANALYZING UNIT POWER to OFF. OFF.
2
Place sample tubes with bar code labels that meet the requirements mentioned in the operation manual.
3
See if the window on the sample department is stained. If so, clean the window with absolute alcohol-dipped gauze and go to the next step. If not, go to the step 8.
4
Restart the analyzer.
5
Enter the Scanner the Scanner Setup Setup & Maintenance screen Maintenance screen of the system software.
6
Enter the Maintenance tab. Maintenance tab.
7
Enter the sample position range in the Sample Position field Position field and click the Step Scan button. Scan button. Observe the scanning procedures to see if the scanner can read the bar codes. If not, go to the next step.
8
Enter the Setup tab. Setup tab.
9
Select all symbologies but do not select any Check or Check or Length. Length .
10
Enter the Maintenance tab. Maintenance tab.
11
Click the Scan button Scan button to see if the scanner can read the bar codes. If not, go to the next step.
12
Place the test fixture to the position 28 on the sample disk.
13
Click the Mechanism button Mechanism button at the System Maintenance screen Maintenance screen to reset all mechanical parts. If the laser from the scanner cannot reach the bar code through the slit on the test fixture, adjust the screws on the brackets until the laser from the scanner can reach the bar code.
7-10
Service Flow
Bracket 1
Bracket 2
14
7.9
If the scanner cannot read the bar codes successfully, replace the bar code scanner. If the new scanner cannot read the bar code, you need to check the bar code labels the user used.
Feeder Feeder Failu re
7.9. 7.9.1 1 Transd Transd ucer Distr ibut ion of the Feede Feederr Pressure Transducer
No-Cuvette Transducer
Cuvette-Pushing Limit Transducer
Insufficient-Cuvette Transducer
7-11
Cuvette-taking limit transducer
Service Flow
7.9.2 Feeder Failure
The trolley moves backward to the end, but does not move forward. Possible causes:
Feeder failure symptom
1. The cuvette-taking limit signal has not been detected. 2. The belt is too tight, so the motor cannot run. Troubleshooting: 1. Check whether the voltage is 0 when the cuvette-taking limit optical coupler is blocked. 2. Adjust the position of the motor to loosen the belt.
The trolley moves forward and stops before compacting the cuvette segment
1. The moving resistance is too large, so the trolley fails to compact the cuvette segment, but the pressure transducer has been blocked. 2. The sliding block falls off, resulting in large friction. 3. The motor cannot work normally. 4. The belt is too tight. Troubleshooting: 1. Clean the guiding axis of the trolley. 2. Replace the sliding block assembly. 3. Replace the motor. 4. Loosen the belt.
The trolley keeps moving backward and forward
1. The no-cuvette transducer has not been pressed completely, so the system deems that the cuvette compartment is empty, and the trolley moves back for cuvette segments. Troubleshooting: 1. Adjust the position of the no-cuvette transducer. 2. Replace the feeder assemblies.
7-12
Service Flow
7.9.3 Manipulator Failure
Manipulator verticalmovement error: Cannot reach/leave the initial position.
The signal of the vertical optical coupler has not been detected after the motor runs for specified steps. Possible causes: 1. The resistance of the lead screw is too large (frequently). 2. The optical couplers for the vertical initial position cannot work (scarcely). 3. The vertical motor of the manipulator cannot work (scarcely). 4. The optical coupler connector is not connected well (infrequently). 5. The power driving board cannot work (infrequently). 6. The main control board cannot work (scarcely).
Troubleshooting: 1. Lubricate the lead screw after cleaning it. 2. Replace relevant assemblies.
The upper hand of the manipulator cannot close/open. Direct cause: The finger signal has not been detected after the electromagnet of the manipulator is switched on.
1. The snap ring of the manipulator is not flexible (infrequently). 2. The coupler of the upper finger cannot work any more (scarcely). 3. The upper hand of the manipulator is not arranged well with the position of the feeder/reaction disk (frequently). 4. The coupler connector is not connected well (infrequently).
1. Adjust the positions of the upper hand and the feeder/ reaction disk. 2. Replace the manipulator (Its position must be adjusted).
1. The operator has added too many cuvette segments into the feeder (behind the trolley). When the trolley moves backward, the jaw is pressed back by the cuvette segments; when it moves forward again, the hand fails to open in time, so the jaw may press one side of the trolley only. As a result, the cuvette segment at the front fails to corporate with the upper hand of the manipulator well (frequently). 2. The lower hand of the manipulator fails to corporate with the reaction disk position well (frequently). 3. The snap ring of the manipulator is not flexible (infrequently). 4. The lower hand coupler of the manipulator cannot work any more (scarcely).
1. Press the cuvette-taking key after adding cuvette segments into the feeder. 2. Adjust the positions of the lower hand and reaction disk. 3. Replace the manipulator (Its position must be adjusted)
1. The snap ring of the manipulator is not flexible (frequently). 2. The finger coupler of the manipulator cannot work any more (scarcely).
Troubleshooting: 1. Replace the manipulator (Its position must be adjusted).
Manipulator Failure
The lower hand of the manipulator cannot close/open. Direct cause: The finger signal has not been detected after the electromagnet of the manipulator is switched on.
The upper/lower hand of the manipulator accidentally closes/ opens. Direct cause: After the manipulator is switched on, the finger coupler signal is received, but the signal changes suddenly while the manipulator is moving.
7.10 7.10 Trou bleshoot bl eshoot ing o f ISE ISE Unit (optio nal) When there is something wrong with the ISE unit, an error code will be displayed at the screen and saved in the log. The error code of the ISE unit has a format of 7 ASCII characters, the format of which is <1234567>. Each digit represents a certain error.
Digit 1: Air in sample, Calibrant A, Calibrant B, Bubble Detector, mV range.
Digit 2: Sample/Calibrant B mV range.
7-13
Service Flow
Digit 3: mV out of range for Calibrant A in Sample Cycle, or Calibrant B in Urine Cycle.
Digit 4: Noise in mV of Calibrant B in Calibration Cycle, or Sample mV during a Sample Cycle.
Digit 5: mV Noise in Calibrant A of the Calibration Cycle, or Sample Cycle, mV Noise in Calibrant B of the Urine Cycle.
Digit 6: mV Drift of calibrant A.
Digit 7: Slope of Sample value is out of range.
Each row represents where the error locates in. You can find all the seven numbers in the table below digit by digit and then know what is wrong with the ISE unit. 0 means no error. For example, when you see <0001000>, you only need to find where the “1” locates in the Digit 4 column and you will see it locates in the Na+ row. So you will know the + cause of this error is the noise in mV of Calibrant B or sample related to the Na electrode. Error
Digit 1
Digit 2
Digit 3
Digit 4
Digit 5
Digit 6
Digit 7
Air/ Hardwa re
mV Out
mV Out
Calib/S ample
Cal A in Calib/S ample Mode, Cal B in Urine Mode
mV Noise
mV Noise
Cal Drift
Cal B/Samp le
Cal A in Calib/S ample Mode, Cal B in Urine Mode
in Sample Mode, Cal B in Urine Mode
Out of Slope/ Machin e Ranges
S
0
0
0
0
0
0
Air in Calibrant A
A
0
0
0
0
0
0
Air in Calibrant B
B
0
0
0
0
0
0
Air in Cleaner
C
0
0
0
0
0
0
Air in Segment
M
0
0
0
0
0
0
Pump Calibration
P
0
0
0
0
0
0
No Flow
F
0
0
0
0
0
0
Bubble Detector
D
0
0
0
0
0
0
Reagent Read
Chip
R
0
0
0
0
0
0
Reagent Write
Chip
W
0
0
0
0
0
0
Store Calibration Value Error
Q
0
0
0
0
0
0
Command Error
T
0
0
0
0
0
0
No Error
0
0
0
0
0
0
0
Na+
0
1
1
1
1
1
1
Air Sample/Urine
in
7-14
A
Service Flow Error
Digit 1
Digit 2
Digit 3
Digit 4
Digit 5
Digit 6
Digit 7
Air/ Hardwa re
mV Out
mV Out
Calib/S ample
Cal A in Calib/S ample Mode, Cal B in Urine Mode
mV Noise
mV Noise
Cal Drift
Cal B/Samp le
Cal A in Calib/S ample Mode, Cal B in Urine Mode
in Sample Mode, Cal B in Urine Mode
Out of Slope/ Machin e Ranges
K+
0
2
2
2
2
2
2
Na+, K+
0
3
3
3
3
3
3
Li+
0
4
4
4
4
4
4
Li+, Na+
0
5
5
5
5
5
5
Li+, K+
0
6
6
6
6
6
6
Li+, K+, Na+
0
7
7
7
7
7
7
Cl-
0
8
8
8
8
8
8
Cl-, Na+
0
9
9
9
9
9
9
Cl-, K+
0
A
A
A
A
A
A
Cl-, Li+
0
B
B
B
B
B
B
Cl-, K+, Na+
0
C
C
C
C
C
C
Cl-, Li+, Na+
0
D
D
D
D
D
D
Cl-, Li+, K+
0
E
E
E
E
E
E
Cl-, Li+, K+, Na+
0
F
F
F
F
F
F
A
The following table displays the causes and the correction measures of some errors. Note that if the correction measure is the replacement of board, bubble detector or other components in the ISE unit, you have to replace the whole ISE unit. You can refer to the chapter 5 for detailed information about some replacement procedures.
Symptom System respond
Problem does
not
Correction
1. No power 2. Communication failure
Turn off the system and then start it again.
3. RS232 cable is disconnected or damaged.
Move the panel of analyzing unit and reconnect or replace cable between the ISE unit and the main control board.
4. Module connector has been damaged.
Replace board.
5. Component failure on board.
Replace board.
7-15
Service Flow Symptom
Problem
Correction
Low Slope
1. Misalignment of electrodes
Remove and replace electrodes <45 mV/decade to reseat properly.
2. Deterioration of calibrator solutions
Replace Calibrant B first and retest. If still low, replace Calibrant A and retest.
3. Deterioration of electrode (low slope)
Replace and test.
Na+, K+, or Li+ <45 mV/decade Cl- <35 mV/decade or High Slope Na+, K+, or Li+ >63 mV/decade
sensing
Single electrode
electrodes
4. Air bubble on reference electrode membrane
Remove electrode, tap to dislodge bubble, replace, and recalibrate.
5. Deterioration of reference electrode
Replace reference and retest.
6. Module or fluid temperatures exceed 37° C. (high slope)
Monitor temperature. Change instrument location if ambient is too great.
1. Deterioration electrode
Replace and test.
Cl- >60 mV/decade
Noise Error Flag
problem
of
sensing
2. Electrical noise spike from environmental source
problem
electrode
electrodes
a) Check for electrical noise coincident with activation. b) Check grounding of Module. c) Component failure on Module board. Replace board.
Noise Error Flag Multiple electrodes
1. Deterioration of reference electrode
Replace reference and retest.
electrode
2. Electrical noise spike from environmental source
a) Check for electrical noise coincident with activation. b) Check grounding of Module. c) Component failure on Module board. Replace board.
Drift Error Flag Single electrode
Drift Error Flag Multiple electrodes
1. May occur when new electrode or new bottle of Calibrant A is installed by system.
Purge the Calibrant A and recalibrate the module. If the electrode is new it may initially drift as it rehydrates over the course of 15 minutes.
2. Deterioration electrode
Replace the problem electrode and test.
of
sensing
1. May occur when new electrode or new bottle of Calibrant A is installed on system.
Purge Calibrant A recalibrate the Module.
2. Deterioration of reference electrode.
Replace reference and retest.
7-16
and
electrode
Service Flow Symptom
Problem
Correction
3. Electrical spike environmental source.
from
a) Check for electrical noise coincident with activation. b) Check grounding of Module. c) Component failure on Module board. Replace the board.
Air in Sample
1. Insufficient sample pipetted into Module sample entry port
a) Host instrument must deliver 70 µL. Increase dispensed volume. b) Operator must place sufficient sample in sample cup to account for all tests programmed. c) Fluid leak in electrodes.
2. Sample not positioned
a) Electrodes not seated sea ted properly. Reseat electrodes. b) Pump tubing aging. Replace. c) Pump tubing obstructed or tubing length is excessive.
Air in Sample Calibrant A
and
1. Sample and Calibrant A are segmented with air.
a) Electrodes are not properly seated or compressed. Check compression plate, spring and seal. Reseat electrodes. b) Ensure that all electrodes and o-rings are in place.
2. Fibrin or salt is plugging the electrode flow path.
a) Use Cleaning procedure for Module. b) Disassemble Module and clean or replace electrode with plugged flow path.
3. Bubble malfunctioning.
detector
4.Waste pump malfunctioning. Air in Calibrant B and Air in Calibrant A
is
Replace bubble detector.
is
Replace waste pump.
1. Calibrant B and Calibrant A are segmented with air.
a) Electrodes are not properly compressed. Check compression plate, spring and seal. b) Ensure that all electrodes and o-rings are in place. c) Ensure tubing is connected properly. d) Replace tubing.
7-17
Service Flow Symptom
Problem
Correction
2. Fibrin or salt is plugging the electrode flow path.
a) Use Cleaning procedure for Module. b) Disassemble Module and clean or replace electrode with plugged flow path.
Air in Calibrant A (no “Air” errors reported for Sample or Calibrant B)
3. Waste pump malfunction.
Replace waste pump, tubing.
4. Bubb Bubble le Detec Detector tor malfunc malfunctio tion. n.
Replac Replace e bubbl bubble e detec detector. tor.
1. Calibrant A pouch is empty.
Replace Calibrant A with a new one, purge and recalibrate.
2. Tubing is disconnected.
Reconnect or replace tubing.
3. Calibrant A pump is not working properly.
a) Check electrical connections. b) Replace pump tubing. c) Replace motor.
4. Tubing is plugged, split or crimped.
7-18
Replace tubing
Mechanical Structure
8 Mechanic Mechanical al Structu tru cture re
Reagent Reagent disk moto r
Synchronous belts
Coder sensors
8-1
Reaction Reaction disk motor
Mechanical Structure
Reagent Reagent prob e assembly
Mixing bar assembly
Reagent Reagent coolin g assembly
8-2
Sample probe assembly
Mechanical Structure
Lamb assembly
Big head of fiber
Fan Fan of Lamb assembly
8-3
Mechanical Structure
Reacti Reacti on dis k TEMP TEMP. control assembly assembly
Optical assembly
Small heads heads of f iber
Reference Reference light assembly
8-4
Mechanical Structure
Washing assemblies
Reagent Reagent syri nge
Sample syringe
8-5
Valves
Pumps
Mechanical Structure
A/D c on ver si on bo ard
Main Main cont rol board
Power drive board
Reagent Reagent refrigeration b oard
8-6
Mechanical Structure
Loading system
Manipulator assembly
8-7
Mechanical Structure
Top of mi xing assembly Top o f sample assembly
Top o f reagent assembly
Bar code scanner (optional)
8-8
Tools and Parts
9 Tools oo ls and Parts Parts 9.1 9.1
Serv Servic ice e Tools oo ls S/N
TOOL
Qu an t i t y
Sp ec if if i c at at i o n
Un i t
1
Allen Key
1 set
1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 5.0
mm
2
Phillips screwdriver
1 set
3*75, 5*75, 6*100
mm
3
Flat-bladed screwdriver screwdrive r
1 set
3*75, 5*75, 6*100
mm
4
Long Allen Key
1 set
2.5, 3.0
mm
5
Tweezers
1 pair
small
/
6
Multimeter Multimete r
1
common
/
7
Movable Wrench
1
common
/
8
Electric Electri c iron
1
30
VA
9
Stainless Stainles s steel
1
0.5 (diameter), (length)
10
Knife
1
common
/
11
Pincers
1
common
/
9.2
15
mm
Parts
S/N
P/N
Name
1
BA30-20-06448
Cable for reagent probe assembly’s elevator motor
2
BA30-20-06449
Cable for reagent probe assembly’s rotary motor
3
BA30-20-06450
Cable for Sample syringe’s motor
4
BA34-20-63601
Cable for Reagent syringe’s motor
5
BA30-20-06452
Cable for sample probe assembly’s elevator motor
6
BA30-20-06453
Cable for sample probe assembly’s rotary motor
7
BA30-20-06454
Cable for mixing bar assembly’s rotary motor
8
BA30-20-06455
Cable for mixing bar assembly’s elevator motor
9
BA30-20-06458
Cable for reagent compartment motor
10
BA30-20-06459
Cable for sample compartment motor
11
BA34-20-63602
Cable for reaction compartment motor
12
BA30-20-06464
Cable for loading system motor
13
BA30-20-06468
Cable for Sample and reagent valves
9-1
Tools and Parts S/N
P/N
Name
14
BA30-20-06476
26-pins sensor cable
15
BA30-20-06477
34-pins sensor cable
16
BA30-20-06480
Detection Detectio n cable for reaction compartment’s fan
17
BA30-20-06481
Cable between CPU board and power drive board
18
BA30-20-06484 BA30-20-06484
Cable for waste and distill water container
19
BA30-20-06485
Cable between CPU board and loading system
20
BA34-20-63651
Cable between CPU board and manipulator
21
BA30-20-06523
Temperature sensor cable of reagent cooling system
22
BA30-20-15291
Cable between CPU board and reagent detection detectio n board
23
BA30-20-15292
Cable between CPU board and sample detection detectio n board
24
BA34-20-63603
Cable between power drive board and mixing motor
25
BA34-20-63631
Reagent probe cable
26
BA30-20-15295
Sample probe cable
27
BA30-20-15296
Mixing bar cable
28
A21-000010---
Tubing
29
BA30-10-15051
Single-way valve
30
0040-10-32303
Tubing
31
0040-10-32304
Connector
32
0040-10-32305
Connector
33
M90-100014---
1/4-28 UNF LOCK NUT-RED NYL LNS-3
34
M90-100015---
1/4-28 UNF LOCK NUT-RED N NYL LNS-4
35
M90-100030---
1/8 Y BARBED FITTING-WHITE FITTING-W HITE NYL Y230-
M90-100050---
BULKHEAD MOUNTED FEMALE LURE RED NYLON CCLR-3
M90-100051---
BULKHEAD MOUNTED FEMALE LURE GREEN NYLON CCLR-4
38
M90-100071---
Tubing
39
BA30-30-15169
5-way connector
40
BA30-10-06666
Step motor
41
BA30-10-15041
synchronous belt (220 XL037)
42
BA30-10-15258
Door’s magnet (couple)
43
BA34-30-63559
Power drive board
44
BA30-30-06312
Reagent cooling control board
45
BA34-30-63613
Sample compartment assembly
36 37
9-2
Tools and Parts S/N
P/N
Name
46
BA30-10-06365
Halogen Lamb (50W)
47
BA30-30-06691
REF pre-amplification pre-amplific ation board
48
BA30-30-15228
340nm amplification amplific ation assembly
49
BA30-10-06367
Filter 340
50
BA30-30-06673
340nm pre-amplification pre-amplifi cation board
51
BA30-30-15229
405nm pre-amplification pre-amplifi cation assembly
52
BA30-10-06368
Filter 405
53
BA30-30-06675
405 pre-amplification pre-amplific ation board
54
BA30-30-15230
450nm pre-amplification pre-amplifi cation assembly
55
BA30-10-06369
Filter 450
56
BA30-30-06677
450 pre-amplification pre-amplific ation board
57
BA30-30-15231
510nm pre-amplification pre-amplifi cation assembly
58
BA30-10-06370
Filter 510
59
BA30-30-06679
510 pre-amplification pre-amplific ation board
60
BA30-30-15232
546nm pre-amplification pre-amplifi cation assembly
61
BA30-10-06371
Filter 546
62
BA30-30-06681
546 pre-amplification pre-amplific ation board
63
BA30-30-15233
578nm pre-amplification pre-amplifi cation assembly
64
BA30-10-06372
Filter 578
65
BA30-30-06683
578 pre-amplification pre-amplific ation board
66
BA30-30-15234
630nm pre-amplification pre-amplifi cation assembly
67
BA30-10-06373
Filter 630
68
BA30-30-15235
630 pre-amplification pre-amplific ation board
69
BA30-30-15235
670nm pre-amplification pre-amplifi cation assembly
70
BA30-10-06374
Filter 670
71
BA30-30-06687
670 pre-amplification pre-amplific ation board
72
BA30-30-15236
700nm pre-amplification pre-amplifi cation assembly
73
BA30-10-06375
Filter 700
74
BA30-30-06689
700 pre-amplification pre-amplific ation board
75
BA30-30-06565
Drive assembly of sample probe
76
BA34-30-63637
Syringe assemply
77
3100-10-49438
100ul Syringe
78
BA30-10-06651
500ul Syringe
79
BA30-10-15115
Step motor
9-3
Tools and Parts S/N
P/N
Name
80
BA30-20-06594
3-way connector for syringe
81
BA30-21-15161
Sensor cable for reagent syringe
82
BA30-21-15162
Sensor cable for sample syringe
83
BA30-30-06634
Mixing bar assembly
84
BA30-21-15159
Cable for mixing bar assembly’s elevator motor
85
BA30-21-15160
Cable for mixing bar assembly’s rotary motor
86
BA30-30-06639
Fans of lamb assembly
87
BA30-30-06641
Washing pool assembly
88
0040-10-32302
Connector
89
0040-10-32303
Washer
90
BA30-21-15311
Valves assembly (KNF)
91
BA30-30-06702
Reaction compartment assembly
92
BA30-30-06715
A/D converter assembly
93
BA30-30-06752
Rocker of reagent probe
94
BA30-30-06304
Liquid detection board
95
BA30-30-06753
Reagent probe assembly
96
BA30-30-06761
Heat-block Heat-bloc k assembly
97
BA30-30-06768
Rocker of sample probe
98
BA30-30-14978
Sample probe assembly
99
BA30-30-06770
Rocker of mixing bar
100
BA34-20-63653
Loading system assembly
101
BA30-21-06488
Pressure sensor and cable for loading system
102
BA30-21-06490 BA30-21-06490
Cable for cuvette-pushing cuvette-pushing limit transducer
103
BA30-21-06491
Cable for cuvette-taking cuvette- taking limit transducer
104
BA30-21-06492 BA30-21-06492
Cable for insufficient-cuvette insufficient-cuvette transducer
105
BA34-20-63648
Board for loading system
106
BA30-21-06489
Cable for no-cuvette no-cuvett e transducer
107
BA30-30-15034
Code sensor of reaction compartment
108
BA30-30-15095
Reagent compartment assembly
109
BA30-30-06780
Reagent disk handle assembly
110
BA30-30-15113
Driving assembly of reagent probe
111 111
BA30-21-15155
Sensor cable for reagent probe’s up-down sensor
112
BA30-21-15156
Sensor cable for reagent probe’s rotary sensor
113
BA30-30-15166
Coder sensor for reagent disk
9-4
Tools and Parts S/N
P/N
Name
114
BA30-30-15167
Coder sensor for sample probe’s sensor
115
BA30-30-15284
Probes connection board
116
BA34-30-63557
Main control board
117
BA33-30-35098
Power supply module
118
BA33-30-35078
PFC Board
119
BA33-30-35082
24V Board
120
BA33-30-35080
12V&5V Board
121
TSB1-20-20399
Fan
122
BA34-20-63655
Manipulator assembly
123
BA33-30-35223
Power socket and switch assembly
124
BA34-20-63622
Reagent temperature control assemply (220V)
125
BA34-20-63614
Cover of sample compartment
126
BA30-30-15029
Cover of reagent compartment
127
BA34-20-63638
Reagent temperature control assemply (110V)
9-5
Maintenance And Test Software
10
Main Maintenance tenance And An d Test Software
10.1 General The maintenance and test software of the BS-300 Chemistry Analyzer is the software that helps maintenance personnel debug and maintain the BS-300 Chemistry Analyzer. Maintenance personnel can, through this software, check whether the probes, mixing bar, disks, fluidic system and photometric system of the BS-300 Chemistry Analyzer work normally, modify parameters, and execute individual commands or combined commands. Run the NewDebug.exe NewDebug.exe file to enter the Maintenance and Test window. See Figure 10-1.
Figure 10-1 Maintenance and Test window
Display Area
As shown in Figure 10-1, there are 4 tabs in the Maintenance and Test window:
Command
PARA an d Speed
Temperature
Photoelectric
10-1
Maintenance And Test Software
Note: The buttons displayed in gray in software interface are not functional.
Following is a brief introduction to the software interface.
Display area
At the bottom of the window is the display area for displaying the frame information and fault/error information. The upper box is for frame information, and the lower one is for fault/error information. There are five types of frame information, including Result, Result , Echo, Echo , Data, Data, Command and Command and Warning. Warning . Select the check box in front of frame information to display relevant frame information. Click the check box again to de-select, and the frame information will not be displayed. Click the Clear button button to clear the information displayed in relevant box. The two Clear buttons buttons are respectively for two boxes.
Setup button
Click the Setup button at up right corner to show a dialog as shown in Figure 10-2, which is used to change communication port and set baud rate and check sum type of the port. Check sum type includes odd, even and none.
Figure 10-2 Serial Port Setup Communication Port
Baud Rate
Check Sum Type
Ab out… ou t… button button
Clicking the Ab the Ab out… ou t… button at up right corner of the interface, you can view the version information of this software. In the following sections, the 4 tabs mentioned above will be detailed respectively.
10-2
Maintenance And Test Software
Note: The units of parameters involved in this software are respectively: Time: 50ms; Sample (reagent) aspirating/dispensing volume: steps of motors; Temperature: ℃. Temperature control coefficient: 0.01.
10.2 Command In the Command Command tab, you can send commands to the analyzing unit and check whether probes, mixing bar, disks, fluidic system, photometric system, ISE unit and built-in bar code scanner work normally.
Figure 10-3 Command tab Command tab Single Command Area
Macroinstruction Area
As shown in Figure 10-3, the Command tab Command tab mainly consists of two parts: Single Command Area and Macroinstruction Area.
10.2 10.2.1 .1 Singl e Command Area Ar ea In Single Command Area, there are 8 sub-tabs:
Main Unit
Reagent Reagent Uni t
Sample Unit
10-3
Maintenance And Test Software
Load Unit
Temp Unit
Photoelectric Photoelectric Unit
ISE Module
Sample Bar Code Module
Check the analyzing unit for their counterparts. There are several buttons in each tab. Click a button and you can send a command to the counterpart on the analyzing unit.
Note: After starting this software, you must click the Download parameters to Analyzing Unit button in the Main Unit Unit sub-tab. Otherwise, the software will not operate the analyzing unit successfully. After clicking the Enable modifying parameters button in the Main Unit Unit sub-tab, you can modify the parameters in the PARA and Speed tab. Speed tab.
Following are instructions to the action implemented by each command.
Main Unit
Command
What it does
Download parameters Analyzing Unit
to
Configures running analyzing unit.
parameters
for
the
Reset all mechanical parts
Moves all units of the analyzing unit to their initial position, and cleans reagent probe, sample probe and mixing bar.
Enable parameters
modifying
Switches on the switch for modifying parameter through this operation before modifying parameter at PARA and Speed tab. Speed tab.
Disable parameters
modifying
Switches off the switch for modifying parameter through this operation after modifying parameter at PARA and Speed tab. Speed tab.
Handshake with Analyzing Unit
Checks that the communication cable works normally.
Send a direct instruction
Edits and sends user-defined commands.
Reagent Reagent Uni t
Command
What it does
Reagent probe wash
Cleans the reagent probe.
Pump, valve (interior) on
Turns on the pump and the valves that control flow of washing the interior of reagent probe.
10-4
Maintenance And Test Software Command
What it does
Pump, valve (interior) off
Turns off the pump and the valves that control flow of washing the interior of reagent probe.
Pump, valve (exterior) on
Turns on the pump and the valve that control flow of washing the exterior of reagent probe.
Pump, valve (exterior) off
Turns off the pump and the valve that control flow of washing the exterior of reagent probe.
Probe above reaction disk
Moves the reagent probe above the reaction disk.
Probe above outer circle
Moves the reagent probe above the outer circle of reagent disk.
Probe above inner circle
Moves the reagent probe above the inner circle of reagent disk.
Probe tip above bottle top (outer circle)
Moves the reagent probe until the probe tip a few millimeters above the top of the reagent bottle in the outer circle of reagent disk.
Probe tip above bottle top (inner circle)
Moves the reagent probe until the probe tip a few millimeters above the top of the reagent bottle in the inner circle of reagent disk.
Probe down to outer circle
Moves the reagent probe down to the reagent bottle in the outer circle of reagent disk.
Probe down to inner circle
Moves the reagent probe down to the reagent bottle in the inner circle reagent disk.
Probe down to reaction disk
Moves the reagent probe down to the cuvette in the reaction disk.
Probe tip above cuvette top
Moves the reagent probe until the probe tip a few millimeters above the top of cuvette in the reaction disk.
Probe down to wash well
Moves the reagent re agent probe to its washing position.
Probe above wash well
Moves the reagent probe to the top of wash well.
Reagent disk rotate
Rotates the reagent disk for specified circles and stop it at specified bottle position. Input rotating circles in the first box and position No. in the second box.
Reagent position
disk
rotate
to
Rotates the reagent disk to specified bottle position. Input position No. in the box.
Reagent syringe init
Moves the reagent syringe to its initial position.
Reagent syringe full
Moves the reagent syringe to its largest travel.
Reagent syringe empty
Empties the reagent syringe.
Reagent unit reset
Cleans the reagent probe and rotates the reagent disk to position No. 1.
10-5
Maintenance And Test Software
Sample Unit
Command
What it does
Sample probe wash
Cleans the sample probe.
Pump, valve (interior) on
Turns on the pump and the valves that control flow of washing the interior of sample probe.
Pump, valve (interior) off
Turns off the pump and the valves that control flow of washing the interior of sample probe.
Pump, valve (exterior) on
Turns on the pump and the valve that control flow of washing the exterior of sample probe.
Pump, valve (exterior) off
Turns off the pump and the valve that control flow of washing the exterior of sample probe.
Probe above wash well
Moves the sample probe to the top of wash well.
Probe above reaction disk
Moves the sample probe to the top of reaction disk.
Probe above outer circle
Moves the sample probe above the outer circle of sample disk.
Probe above inner circle
Moves the sample probe above the inner circle of sample disk.
Probe tip above tube top (outer circle)
Moves the sample probe until the probe tip a few millimeters above the top of the tube in the outer circle of sample disk.
Probe tip above tube top (inner circle)
Moves the sample probe until the probe tip a few millimeters above the top of the tube in the inner circle of sample disk.
Probe tip above cuvette top
Moves the sample probe until the probe tip a few millimeters above the top of cuvette in the reaction disk.
Probe down to reaction disk
Moves the sample probe down to the reaction cuvette in reaction disk.
Probe down to wash well
Moves the position.
Probe down to outer circle
Moves the sample probe down to the tube in the outer circle of sample disk.
Probe down to inner circle
Moves the sample probe down to the tube in the inner circle of sample disk.
Sample disk rotate
Rotates the sample disk for specified circles and stop it at specified tube position. Input rotating circles in the first box and position No. in the second box.
Sample position
disk
rotate
to
sample probe
to its
washing
Rotates the sample disk to specified tube position. Input position No. in the box.
Sample syringe init
Moves the sample syringe to its initial position.
Sample syringe empty
Empties the sample syringe.
Sample syringe full
Moves the sample syringe to its largest travel.
10-6
Maintenance And Test Software Command
What it does
Sample unit reset
Cleans the sample probe and rotate the sample disk to position No. 1.
Load Unit
Command
What it does
Mixer Mixer to the top top of wash wash well well
Moves Moves the the mixing mixing bar bar to the the top top of wash wash well well..
Mixer move into wash well
Moves the mixing bar to its washing position.
Mixer move to the top of reaction disk
Moves the mixing bar to the top of reaction disk.
Mixer disk
Moves the mixing bar to mixing position in reaction disk.
move
into
reaction
Mixer stir for a while
Mixing bar stirs for a specified period.
Mixing bar rotates for a while above wash well
Moves the mixing bar above its wash well and then rotates it for a specified period.
Take new cuvettes
Moves the compartment trolley backwards to get new cuvettes.
Load New cup
Rotates the reaction disk to specified position, then get new cuvettes with the manipulator and put them in reaction disk.
Unload cup
Rotates the reaction disk to specified position, then get the cuvettes in reaction disk with the manipulator and put them in used-cuvette bucket.
Motion Motion of manipulator manipulator fingers
Tightens Tightens or loosen manipulator’s manipulator’s (grabbing mechanism).
fingers
After entering a value of 0-3 to the box, click the button and the fingers will motions. 0 means both fingers tighten; 1 means the upper loosen and the lower tightens; 2 means the upper tightens and the lower loosens; and 3 means both loosen. Manipulator to vertical init. pos.
Moves the position.
Manipulator pos.
init.
Moves the manipulator to horizontal initial position.
To H-pos. H-pos. for taking taking cuvettes cuvettes
Moves the manipul manipulator ator to to horizontal horizontal position position for getting cuvettes.
To V-pos. for taking cuvettes cuvettes
Moves the manipulat manipulator or to vertical position position for getting cuvettes.
To H-position cuvettes
for
placing
Moves the manipulator to horizontal position for placing cuvettes.
To V-position cuvettes
for
placing
Moves the manipulator to vertical position for placing cuvettes.
to
horiz.
10-7
manipulator
to
vertical
initial
Maintenance And Test Software Command
What it does
Move steps vertically
Moves the manipulator for specified steps in vertical way wa y.
Move steps horizontally
Moves the manipulator for specified steps in horizontal way. way.
Loader unit reset
Moves the mixing bar to its initial position, loosen manipulator’ grabbing mechanism and move the manipulator to its initial position.
Temp Uni t
Command
What it does
Temperature control on
Turns on controller(s).
the
specified
temperature
There’re two temperature controllers Reaction Disk and Disk and Reagent Pr eHeat eHeat .
as
Selects the check box(es) in front of a controller(s), click the button, you can turn on the selected controller(s). Temperature control off
Turns off controller(s).
the
specified
temperature
There’re two temperature controllers Reaction Disk and Disk and Reagent Pr eHeat eHeat .
as
Selects the check box(es) in front of a controller(s), click the button, you can turn off the selected controller(s).
Photoelectric Photoelectric Unit
Command
What it does
Reaction disk rotate
Rotates the reaction disk for specified circles and stop it at specified cuvette position. Input rotating circles in the first box and position No. in the second box.
Reaction disk rotate some positions
Rotates the reaction disk for specified cuvette positions, and input number of cuvette positions in the box.
Rotate and measure
Rotates the reaction disk, collects reaction data during rotating, and then stops it at specified cuvette position.
Turn on light
Turns on the light.
Turn off light
Turns off the light.
Test dark current
Turns off the light and measures the dark current at this time. The dark current data will be displayed in the box below.
10-8
Maintenance And Test Software Command
What it does
Test base current
Unloads the first segment of cuvettes in reaction disk with manipulator, then rotates the reaction disk and during rotating measures the light where places the first segment of cuvettes before. The data regarding base current will be displayed in the box below.
ISE Module
Command
What it does
Handshake
The ISE unit shakes hands with the main unit.
Maintain
Executes the maintaining cycle to run the waste pump until the electrode flow path is cleared of fluid.
Detect Bubble
Checks to see if the bubble detector is functioning properly.
Cleaning Period
Removes protein build-up from the ISE Module electrodes.
Calibrate
Calibrates the electrodes of the ISE Module.
Correct Pumps
Calibrates the peristaltic pumps of the ISE Module.
Measure Serum
Measures serum or plasma samples on the ISE Module.
A Purge
Purges Calibrant A solution through the tubing from the pouch of Calibrant A to the ISE Module.
B Purge
Purges Calibrant B solution through the tubing from the pouch of Calibrant B to the ISE Module.
Pumps 200µl of Calibrant A to make it convenient for you to aspirate it from the sample entry port.
Displays the result of bubble detector checking for the last time.
Displays the result of pump correction for the last time.
Displays the software version of the ISE unit.
After clicking this button, the results in mv will also be displayed when calibration, measurement or a sip cycle is finished.
After clicking this button, the results in mv will also be displayed when calibration or measurement is finished.
Disables the effect of clicking or .
Sample Bar Code Module
Command
What it does
Set
Select a symbology from the pull down list and click this button. If Valid is Valid is selected, the symbology will be set as the one that the scanner can recognize. Otherwise, it will be the one that can not be recognized.
10-9
Maintenance And Test Software Command
What it does
Send
Enter an instruction into the edit box and click this button to send it to the analyzing unit.
Laser On
Turns on the laser of the bar code scanner.
Laser Off
Turns off the laser of the bar code scanner.
Handshake
The bar code scanner shakes hands with the main unit.
Scan
After entering the start position No. and the end one into Start Start and End End edit boxes respectively, click this button to scan bar codes on the selected positions.
10.2 10.2.2 .2 Macro Macro ins tru cti on A rea Figure 10-4 Macroinstruction Area Explanation for the macroinstruction
Macroinstruction
Commands
Button
Parameters of the current command
A macroinstruction macroinstruction is a set of commands. When selecting a macroinstruction from the pulldown list in macroinstruction area, you can see there’re various commands and comments regarding the macroinstruction. In addition, parameters of the command currently selected will also be displayed on the interface. In the macroinstruction area, after selecting a macroinstruction from the pulldown list, you can
Click the Run button to run the selected macroinstruction;
10-10
Maintenance And Test Software
Click the Pause button Pause button to suspend the macroinstruction which is running;
Click the Stop button Stop button to stop the macroinstruction which is running.
Note: After all commands in each macroinstruction are executed, the system counts it as one time. When the check box on the right of the macroinstruction pulldown list is selected, the selected macroinstruction will be executed for once; otherwise, the macroinstruction will be executed cyclically.
For each macroinstruction, the commands of it are executed one by one in the sequence in which they are listed. When the set delay expires after a command is executed, the next command is executed immediately. In case of an error in the execution process, the macroinstruction will be stopped. Following are instructions to the action implemented by each macroinstruction. Macroinst Macroinst ruction
Command Command
What What it does
A,B purge
B purge
Purges Calibrant B solution through the tubing from the pouch of Calibrant B to the ISE Module.
A purge
Purges Calibrant A solution through the tubing from the pouch of Calibrant A to the ISE Module.
Reagent probe above wash well
Moves the reagent above its wash well.
Valve on, Pump on
Turns on the pump and valve that control the flow of washing the exterior of reagent probe.
Pump off, Valve off
Turns off the pump and valve that control the flow of washing the exterior of reagent probe.
Sample probe above wash well
Moves the sample above its wash well.
Valve on, Pump on
Turns on the pump and valve that control the flow of washing the exterior of sample probe.
Pump off, Valve off
Turns off the pump and valve that control the flow of washing the exterior of sample probe.
Check exterior fluid path of reagent probe
Check exterior fluid path of sample probe
10-11
probe
probe
Maintenance And Test Software Macroinst Macroinst ruction
Command Command
What What it does
Check interior fluid path of reagent probe
Reagent probe above wash well
Moves the reagent above its wash well.
Valve on, Pump on
Turns on the pump and valves that control the flow of washing the interior of reagent probe.
Pump off, Valve off
Turns off the pump and valves that control the flow of washing the interior of reagent probe.
Sample probe above wash well
Moves the sample above its wash well.
Valve on, Pump on
Turns on the pump and valves that control the flow of washing the interior of sample probe.
Pump off, Valve off
Turns off the pump and valves that control the flow of washing the interior of sample probe.
Test Loader
Reload cup looply
The manipulator gets new cuvettes from the compartment looply, and then put them in the reaction disk.
Test mixing bar in reaction disk
Reaction disk to next cup
The reaction disk rotates and stops. No. of position for reaction disk to stop at will increase.
Mix instruction
The mixing bar moves into reaction cuvette and mixes for a specified period.
Mixer to vertical init
The mixing bar moves to vertical initial position.
Reaction disk to next cup
The reaction disk rotates and stops. No. of position for reaction disk to stop at will increase.
Reagent probe reaction disk
to
The reagent probe moves vertically into bottom of reaction cuvette.
Reagent probe vertical init
to
The reagent probe moves to vertical initial position.
Reaction disk to next cup
The reaction disk rotates and stops. No. of position for reaction disk to stop at will increase.
Check interior fluid path of sample probe
Test reagent probe in reaction disk
Test sample probe in reaction disk
10-12
probe
probe
Maintenance And Test Software Macroinst Macroinst ruction
Urine test diluted)
(auto
Urine test (manually diluted)
Command Command
What What it does
Sample probe reaction disk
to
The sample probe moves vertically into bottom of reaction cuvette.
Sample probe vertical init
to
The sample probe moves to vertical initial position.
Reagent disk in position 50 in next period
Reagent disk rotates to position 50 and in the next period the reagent probe will aspirate urine diluent from the position.
Dispensing 270µl urine diluent
of
Reagent probe aspirates 207µl of urine diluent and dispenses it to a cuvette.
Dispensing 30µl urine sample
of
Sample probe aspirates 30µl of urine sample and dispenses it to a cuvette. The mixing bar stirs it.
Washing mixing bar
Washes the mixing bar.
First period
Sample probe aspirates 70µl of diluted urine to the ISE unit.
Second period
Sample probe aspirates 70µl of diluted urine to the ISE unit.
Third period
Washes the sample probe.
Idle period
Measuring.
Idle period
Measuring.
Idle period
Measuring.
First period
Sample probe aspirates 70µl of diluted urine to the ISE unit.
Second period
Sample probe aspirates 70µl of diluted urine to the ISE unit.
Third period
Washes the sample probe.
Idle period
Measuring.
Idle period
Measuring.
Idle period
Measuring.
10.3 10.3 PARA and Speed In the PARA and Speed tab, Speed tab, you can query, configure or load parameters as well as debug the units.
10-13
Maintenance And Test Software Figure 10-5 PARA and Speed tab Parameters
Buttons
As shown in Figure 10-5, the PARA and Speed tab Speed tab mainly consists of two parts as unit parameter and several operating buttons. Unit Name includes Name includes Main unit, unit , Sample unit , Reagent unit, unit , Photoelectric unit , Temp un it and Load unit. unit . At this screen, you can not only inquire and configure each unit parameters, but also save and read relevant data. Now let’s introduce them respectively.
Inquire unit parameters
1
Select a unit in Unit Name. Name.
2
Click Query button. Query button. For example.
10-14
Maintenance And Test Software
Configure a parameter for a unit
1
Select a unit in Unit Name. Name.
2
Click the Query button. Query button.
3
Edit the parameter if necessary. Otherwise, go to the next step directly. When editing, you can directly select and modify the data to be edited.
4
Select a parameter to be configured. For example.
5
Click the Config button Config button to configure the parameter for relevant unit of the analyzing unit.
Configure all parameters for relevant units
1
Select units in Unit Name. Name.
2
Click the Config All button All button to configure all parameters of the selected unit for relevant units of the analyzing unit.
Save data
1
Select a unit in Unit Name. Name.
2
Click the Save button, Save button, then pop up a dialog.
3
Input a filename and click the Save button, Save button, you can save parameters of the currently selected unit to a .txt file.
Read Data
1
Click Load button, Load button, then pop up a dialog.
2
Select the relevant .txt file and click Open Open button, you can configure parameters in the file for relevant unit of the analyzing unit.
The following contents explain the adjustable parameters.
10-15
Maintenance And Test Software
Reagent unit
Para Paramete meterr
Unit
Default Default
What What the param paramete eterr means means
H-pos. for washing reagent probe
Whole step
42
Position correction of the reagent probe in the horizontal direction when it is above its wash well.
H-pos. of reaction disk
Whole step
-42
Position correction of the reagent probe in the horizontal direction when it is above the reaction disk.
H-pos. of reagent disk inner circle
Whole step
215
Position correction of the reagent probe in the horizontal direction when it is above the inner circle of reagent disk.
H-pos. of reagent disk outer circle
Whole step
165
Position correction of the reagent probe in the horizontal direction when it is above the outer circle of reagent disk.
Position correction of reagent disk inner circle
Micro step
10
The position correction of the reagent disk when it stops at the specified position according to the position No. of its inner circle.
Position correction of reagent disk outer circle
Micro step
10
The position correction of the reagent disk when it stops at the specified position according to the position No. of its outer circle.
Para Paramete meterr
Unit
Default Default
What What the param paramete eterr means means
Bar code position correction of sample disk inner circle
Micro step
18
Position correction of the inner circle of sample disk relative to the outer circle when the bar code scanner scans specified positions.
H-pos. for washing sample probe
Whole step
102
Position correction of the sample probe in the horizontal direction when it is above its wash well.
H-pos. of sample disk inner circle
Whole step
202
Position correction of the sample probe in the horizontal direction when it is above the inner circle of sample disk.
H-pos. of sample disk outer circle
Whole step
165
Position correction of the sample probe in the horizontal direction when it is above the outer circle of sample disk.
Sample unit
10-16
Maintenance And Test Software Para Paramete meterr
Unit
Default Default
What What the param paramete eterr means means
H-pos. of sample probe in reaction disk
Whole step
-106
Position correction of the sample probe in the horizontal direction when it is above the reaction disk.
H-pos. of sample probe in ISE
Whole step
490
Position correction of the sample probe in the horizontal direction when it is above the sample entry port of ISE unit.
Position correction of sample disk inner circle
Micro step
10
The position correction of the sample disk when it stops at the specified position according to the position No. of its inner circle.
Position correction of sample disk outer circle
Micro step
10
The position correction of the sample disk when it stops at the specified position according to the position No. of its outer circle.
V-pos. of sample probe in ISE
Whole step
163
Position correction of the sample probe in the vertical direction when it is above the sample entry port of ISE unit.
Parameter
Unit
Defa efault
Wha What the the para parame mete terr me means
Manipulator h-position for placing cuvettes
Whole step
70
Position correction of the manipulator in horizontal direction when it places a cuvette segment into the reaction disk.
Manipulator h-position for taking cuvettes
Whole step
259
Position correction of the manipulator in horizontal direction when it takes a cuvette segment from the cuvette compartment.
Manipulator v-position for placing cuvettes
Whole step
70
Position correction of the manipulator in vertical direction when it places a cuvette segment into the reaction disk.
Manipulator v-position for taking cuvettes
Whole step
768
Position correction manipulator in vertical when it takes a segment from the compartment.
Mixing bar h-position for washing
Whole step
68
Position correction of the mixing bar in the horizontal direction when it is above its wash well.
Load unit
10-17
of the direction cuvette cuvette
Maintenance And Test Software Parameter
Unit
Defa efault
Wha What the the para parame mete terr me means
Mixing bar h-position for mixing
Whole step
70
Position correction of the mixing bar in horizontal direction when it is above the reaction disk.
Unit
Defa efault
Wha What the the para parame mete terr me means
Photoelectric unit
Parameter Photoelectric parameter 405
gain
/
127
Signal gain coefficient of A/D conversion for wavelength of 405nm.
Photoelectric parameter 450
gain
/
127
Signal gain coefficient of A/D conversion for wavelength of 450nm.
Photoelectric parameter 510
gain
/
127
Signal gain coefficient of A/D conversion for wavelength of 510nm.
Photoelectric parameter 546
gain
/
127
Signal gain coefficient of A/D conversion for wavelength of 546nm.
Photoelectric parameter 578
gain
/
127
Signal gain coefficient of A/D conversion for wavelength of 578nm.
Photoelectric parameter 630
gain
/
127
Signal gain coefficient of A/D conversion for wavelength of 630nm.
Photoelectric parameter 670
gain
/
127
Signal gain coefficient of A/D conversion for wavelength of 670nm.
Photoelectric parameter 700
gain
/
127
Signal gain coefficient of A/D conversion for wavelength of 700nm.
Photoelectric gain parameter reference light
/
127
Signal gain coefficient of A/D conversion for wavelength of reference light.
Photoelectric parameter 340
/
127
Signal gain coefficient of A/D conversion for wavelength of 340nm.
Micro step
25
Correction of the start position when photoelectric collection starts.
gain
Start position of photoelectric collection
10-18
Maintenance And Test Software
10.4 Temperature Figure 10-1 Temperature tab
Reagent Temperature Curve Control Panel
Reaction Tray Temperature Curve Control Panel
The Temperature tab Temperature tab consists of two parts:
Curve of reagent preheated temperature (upper part);
Curve of reaction disk (lower part).
The two temperature curves are controlled respectively through two control panels on the right. Meanings of the parameters on the control panels are as follows. Parameter Parameter
What it does
Value
Displays current temperature value on the temperature curve.
YRange
Temperature range on Y-coordinate of the temperature curve. Input upper limit and lower limit of temperature range in the two boxes.
XRange
X-coordinate point of the temperature curve.
Temp
Intended temperature expected to achieve.
Period
Interval for sampling.
After setting Xrange, Xrange, Yrange and Yrange and Temp on Temp on the control panel, click Confirm button Confirm button to implement them on relevant temperature curve. There are 4 more buttons in each of the two control panels:
Start: Start : After setting Period setting Period,, clicking it, you can start collecting the temperature data;
10-19
Maintenance And Test Software
Stop: Stop : Clicking it, you can stop collecting the temperature data;
Freeze: Freeze : Clicking it, you can freeze the displayed curve;
Resume: Resume: Clicking it, you can resume the frozen curve.
10.5 Photoelectric In the Photoelectric Photoelectric tab, the values of photoelectric data and photoelectric data curve of the selected wavelength are displayed.
Photoelectric tab Figure 10-2 Photoelectric tab Display minimum data of each wavelength
Display maximum data of each wavelength
Display difference between the above two
Display all data Panel Display curve
Through the panel, you can set relevant parameters of photoelectric test data and photoelectric curve. Nine wavelengths (340, 405, 450, 510, 546, 578, 630, 670 and 700) and a base wavelength are listed on the control panel. You can select one of them to display its photoelectric data curve. Parameter Parameter
What it does
MinMax
Displays maximum and minimum for photoelectric data of the selected wavelength.
YRange
Range on the Y-coordinate of the curve. Input upper limit and lower limit for photoelectric data in the two boxes.
XRange
X-coordinate point of the curve.
CupIndex
Specifies the cuvette in certain position of reaction disk to be used to measure photoelectric data.
Dozes-reagent
Reagent dispensed to the specified cuvette.
Dozes-sample
Sample dispensed to the specified cuvette.
Period
Interval for sampling.
10-20
Maintenance And Test Software Parameter Parameter
What it does
Times
The times of sampling.
Mixer Mixer sti r
Select it to stir the reaction mixture; otherwise the reaction mixture will not be stirred.
Reaction Reaction disk rotate
Select it to enable the reaction disk to rotate during measurement with photoelectric data of all wavelengths displayed; otherwise the reaction disk will not be rotated and only photoelectric data of the selected wavelength will be displayed.
Through the control panel, you can select parameters like Wavelength, Wavelength, MinMax, MinMax , YRange, YRange, CupIndex and CupIndex and Period. Period . Moreover, you can select Mixer Mixer sti r as as required. After setting them, click the Confirm button Confirm button to confirm the settings. There are 6 more buttons in the panel:
Start: Start : Clicking it, you can start collecting the temperature data;
Stop: Stop : Clicking it, you can stop collecting the temperature data;
Freeze: Freeze : Clicking it, you can freeze the displayed curve;
Resume: Resume: Clicking it, you can resume the frozen curve;
Save: Save: Clicking it, you can save the photoelectric data of all current wavelengths as a .txt file;
Load: Load : Clicking it, you can read the data in a .txt file saved with photoelectric data.
10-21
Boards Connection Diagrams
Ap A p p end en d i x A
B o ard ar d s Con Co n n ect ec t i o n Diag Di agrr ams am s Manipulator connection board
Power supply assembly
Reagent refrigeration board
Main control board
Power drive board
A/D conversion board
Probes connection board
Feeder connection board
A-1
Boards Connection Diagrams
M42-M43: Refrigeration fan 2100-20-06534
C35: Reagent refrigeration temperature transducer BA30-21-15175
D N G e u l b 2 3 2 5 6 e 0 l b a 0 2 C 0 3 A B e u l b 1
J88
Cable BA30-20-06535
R E 3 I T 3 6 L 6 E P 0 : 1 0 3 2 0 M - 0 0 1 3 2 M
12V GND 12V GND
1
D V N 2 G 1 e d u e l r b 1 2
V 2 1 d e r 1
9 2 5 6 e 0 l b a 0 2 - D C 0 3 N A G B
d e r 2
2
J81
1
2
J79
1
2
J80
1
2
J78
1
2
J82
1
2
J83
1
2
D V N 2 G 1 e d u e l r b 1 2
V 2 1
D N G
V 2 1
e u l b 1
d e r 2
e u l b 1
d e r 2
1
2
1
2
J84
J85
J89 1 red
3 red 4 black
J86
12V
1
2 black
12&5V board BA33-30-35080
12V
2
COOLER
3
12V
4
GND
Reagent refrigeration board BA30-30-06312
GND
COOLER
GNDFAN 12FAN
J74
1
2
3
1
2
3
1
2
3
1
2
3
w o l l e y 1
e u l b 2
d e r 3
w o l l e y 1
e u l b 2
d e r 3
w o l l e y 1
e u l b 2
d e r 3
w o l l e y 1
e u l b 2
d e r 3
N A F
V 2 1
D N G
N A F
V 2 1
D N G
N A F
V 2 1
D N G
N A F
V 2 1
e u l b 2
d e r 1
w o l l e y 3
e u l b 2
d e r 1
w o l l e y 3
e u l b 2
d e r 1
w o l l e y 3
e u l b 2
d e r 1
4 2 5 D 6 e 0 l - N G b a 0 2 C 0 3 A w B o l l e y 3 3
2
1
J73
3
2
1
J72
3
2
1
J71
3
2
1
J75
1
2
1
2
e u l b 1 5 2 5 6 e 0 l b D a 0 2 - N C 0 3 G A B
d e r 2
e u l b 1
d e r 2
V 2 1
D N G
V 2 1
d e r 1
e u l b 2
d e r 1
e u l b 2 2
1
J76
2
1
J77
1
2
e u l b 1
d e r 2
e u l b 2
d e r 1 2
1
J87
1
2
3
k c a l b 1
e t i h w 2
d e r 3
N A F
V 5
e t i h w 2
d e r 3
2
3
0 8 4 6 e 0 l - D b N a 0 2 - G C 0 3 A B k c a l b 1 1
J42 M37-M38: Fans on sides of lamp housing 2100-20-08114
M24-M27: Reaction temperature fan BA30-21-06530
A-5
Fan on rear side of lamp housing
4 yellow
P353
5
5 yellow
4
1
4
1 blue 5
2
6
3
6
12V
Main control board BA34-30-63557
Cable BA30-20-15308
3
2 blue
2
3 white
1
6 orange
Boards Connection Diagrams
M42-M43: Refrigeration fan 2100-20-06534
C35: Reagent refrigeration temperature transducer BA30-21-15175
D N G e u l b 2 3 2 5 6 e 0 l b a 0 2 C 0 3 A B e u l b 1
J88
Cable BA30-20-06535
R E 3 I T 3 L 6 E 6 P 0 : 1 0 3 2 0 M - 0 1 0 2 3 M
12V GND 12V GND
1
D V N 2 G 1 e d u e l r b 1 2
V 2 1 d e r 1
9 2 5 6 e 0 l b a 0 2 - D C 0 3 N A G B
d e r 2
2
J81
1
2
J79
1
2
J80
1
2
J78
1
2
J82
1
2
J83
1
2
D V N 2 G 1 e d u e l r b 1 2
V 2 1
D N G
V 2 1
e u l b 1
d e r 2
e u l b 1
d e r 2
1
2
1
2
J84
J85
J89 1 red
3 red 4 black
J86
12V
1
2 black
12&5V board BA33-30-35080
12V
12V
2
COOLER
3
12V
4
GND
Reagent refrigeration board BA30-30-06312
GND
COOLER
GNDFAN 12FAN
J74
1
2
w o l l e y 1
e u l b 2
4 2 5 D 6 e 0 l - N G b a 0 2 C 0 3 A w B o l l e y 3 3
1
2
3
1
2
3
1
2
3
d e r 3
w o l l e y 1
e u l b 2
d e r 3
w o l l e y 1
e u l b 2
d e r 3
w o l l e y 1
e u l b 2
d e r 3
N A F
V 2 1
D N G
N A F
V 2 1
D N G
N A F
V 2 1
D N G
N A F
V 2 1
e u l b 2
d e r 1
w o l l e y 3
e u l b 2
d e r 1
w o l l e y 3
e u l b 2
d e r 1
w o l l e y 3
e u l b 2
d e r 1
2
3
1
J73
3
2
1
J72
3
2
1
J71
3
2
1
J75
1
2
1
2
e u l b 1 5 2 5 6 e 0 l b D a 0 2 - N C 0 3 G A B
d e r 2
e u l b 1
d e r 2
V 2 1
D N G
V 2 1
d e r 1
e u l b 2
d e r 1
e u l b 2 2
1
J76
2
1
J77
1
2
e u l b 1
d e r 2
e u l b 2
d e r 1 2
1
J87
1
2
3
k c a l b 1
e t i h w 2
d e r 3
N A F
V 5
e t i h w 2
d e r 3
2
3
0 8 4 6 e 0 l - D b N a 0 2 - G C 0 3 A B k c a l b 1 1
J42 M37-M38: Fans on sides of lamp housing 2100-20-08114
M24-M27: Reaction temperature fan BA30-21-06530
A-5
Fan on rear side of lamp housing
Main control board BA34-30-63557
Cable BA30-20-15308
4 yellow
P353
5
5 yellow
4
1
4
1 blue 5
2
6
3
6
3
2 blue
2
3 white
1
6 orange
Boards Connection Diagrams
r o t o m 2 g 5 n 0 5 i x 1 i 0 m 1 C 0 0 D 0 : 9 2 1 - M +
C31: Reagent probe
C32: Sample probe
E B D O N R G P k d c e a r l b 2 1 6 4 5 6 e 0 l b 0 a 2 C 0 3 A B k c a d l b e r 1 2
E B D O N R G P k d c a e r l b 2 1 7 4 5 6 e 0 l b a 0 2 C 0 3 A B k c a d l e r b 2 1
r e c u 1 g d 3 n s 6 i t n 6 a a e t r 0 h 0 e r e 1 r P t 0 : u a 0 r 3 3 e A C p B m e t
r 5 e 1 t 1 a e 5 1 h e r 0 P 1 : 0 9 3 2 A M B
1
J66
2
6 9 2 5 e 1 l b a 0 2 C 0 3 A B
k c a l b 1
d e r 2
1
2
J202
1
2
3
4
1
2
3
4
1
e g n a r o 2
w o l l e y 1 J201
1
n w o r b 3
e l p r u p 4
D N G A n e e r g 5
3
4
5
2
L E V E L _ P R k c a l b 7
V 2 1 d e r 8
6
7
8
2
3
4
2
3
4
D N G A n e e r g 1
O H P _ M A R _ P S e u l b 2
L E V E L _ P S k c a l b 3
V 2 1 d e r 4
1
2
3
4
1
2
3
4
1 n e e r 2 g 9 2 5 e 1 l b a 0 2 C 0 3 A B n e e r g 1
2 e u l b
3 e t i h w
4 d e r
e u l b 2
e t i h w 3
d e r 4
2
3
4
5 9 2 5 e 1 l b a 0 2 C 0 3 A B
O H P _ M A R _ P R e u l b 6
2
Sampel level detection board BA30-30-06304 J65
1 4 9 2 5 e 1 l b a 0 2 C 0 3 A B
1
J66
Reagent level detection board BA30-30-06304 J65
J200
Probes connection board BA30-30-15284
J206
J207
J205
J204
1
2
1
2
1
2
1
1 e u l b 3 0 6 3 e 6 l b a 0 2 C 4 3 A B e u l b 1
2 d e r
1 e u l b 4 0 6 3 e 6 l b a 0 2 C 4 3 A B e u l b 1
2 d e r
2 d e r
d e r 2
1 k c a l 0 b 9 2 5 e 1 l b a 0 2 C 0 3 A B k c a l b 2
1 n e e r 1 g 9 2 5 e 1 l b a 0 2 C 0 3 A B n e e r g 1
2
5 4 J 1
2 2 J 1
d e r 2 2
4 2 J
1
d e r 3 2
3
4 3 J
1
J203 2
3
4
2 e u l b
3 e t i h w
4 d e r
e u l b 2
e t i h w 3
d e r 4
2
3
4
5 3 J
1
Main control board BA34-30-63557
Power drive board BA34-30-63559
A-7
Boards Connection Diagrams
r o t o m 2 g 5 0 n i 5 x 1 i m 0 1 C 0 D 0 : 0 9 2 1 - M +
C31: Reagent probe
C32: Sample probe
E B D O N R G P k c d e a r l b 2 1 6 4 5 6 e 0 l b a 0 2 C 0 3 A B k c a d l b e r 1 2
E B D O N R G P k d c e a r l b 2 1 7 4 5 6 e 0 l b a 0 2 C 0 3 A B k c a d l e r b 2 1
r e c u 1 g d n s 3 i 6 t n 6 a a e t r 0 h 0 e r e r 1 P t 0 : u 3 0 a r 3 e A C p B m e t
r 5 e 1 t 1 a e 5 1 h e r 0 P 1 : 0 9 3 2 A M B
1
J66
2
6 9 2 5 e 1 l b a 0 2 C 0 3 A B
k c a l b 1
d e r 2
1
2
J202
1
2
3
4
1
2
3
4
1
e g n a r o 2
w o l l e y 1 J201
1
n w o r b 3
e l p r u p 4
D N G A n e e r g 5
3
4
5
2
2
L E V E L _ P R k c a l b 7
V 2 1 d e r 8
6
7
8
3
4
2
3
4
D N G A n e e r g 1
O H P _ M A R _ P S e u l b 2
L E V E L _ P S k c a l b 3
V 2 1 d e r 4
1
2
3
4
5 9 2 5 e 1 l b a 0 2 C 0 3 A B
O H P _ M A R _ P R e u l b 6
2
Sampel level detection board BA30-30-06304 J65
1 4 9 2 5 e 1 l b a 0 2 C 0 3 A B
1
J66
Reagent level detection board BA30-30-06304 J65
J200
Probes connection board BA30-30-15284
J206
J207
J205
J204
1
2
1
2
1
2
1
1 e u l b 3 0 6 3 e 6 l b a 0 2 C 4 3 A B e u l b 1
2 d e r
1 e u l b 4 0 6 3 e 6 l b a 0 2 C 4 3 A B e u l b 1
2 d e r
2 d e r
d e r 2
1 k c a l 0 b 9 2 5 1 e l b a 0 2 C 0 3 A B k c a l b 2
1 n e e r 1 g 9 2 5 e 1 l b a 0 2 C 0 3 A B n e e r g 1
2
5 4 J 1
2 2 J 1
d e r 2
4 2 J
2
1
d e r 3 2
4 3 J
3
J203 2
3
4
2 e u l b
3 e t i h w
e u l b 2
e t i h w 3
d e r 4
2
3
4
1
4 d e r
1
2
3
1 n e e 2 r 9 g 2 5 e 1 l b 0 a 2 C 0 3 A B n e e r g 1
2 e u l b
3 e t i h w
4 d e r
e u l b 2
e t i h w 3
d e r 4
2
3
4
5 3 J
1
4
Main control board BA34-30-63557
Power drive board BA34-30-63559
A-7
Boards Connection Diagrams
2 red Cable BA33-21-35151
1 blue 2 red
Cable BA33-21-35152
1 blue
J104
J3 4
MAGU+ MAGU-
3
MAGD+
2
MAGD-
1
Cable J106 BA30-20-06463 4 red 4 red 4 3 blue
3 blue 2 yellow
2 yellow
1 white
1 white
GND MAGD_PHO GND GND
7
3 5
9
2
1
+ D G A M
D G A M
J105
3
MAGU-
2
MAGD+
1
MAGD-
blue red blue
M2: Upper finger magnet BA30-21-06497
M1: Lower finger magnet BA30-21-06498
J101 VCC LOADX_PHO
4 3
GND
2
GND
1
4 red 3 white
Cable BA33-21-35153
2 black
red white black
1 green
C1: Manipulator horizontal position transducer transducer BA30-21-06495
green
J102 VCC LOADY_PHO
Manipulator connection board BA30-30-06316
4 3
GND
2
GND
1
4 red 3 white
Cable BA33-21-35154
2 black
red white black
1 green
C2: Manipulator vertical position transducer transducer BA30-21-06496
green
J108
J43 1
1 U G A M
MAGU+
Main control board BA34-30-63557
URAM_PHO LOADX_PHO DRAM_PHO LOADY_PHO VCC MAGU_PHO
2
+ U G A M
Power drive board BA34-30-63559
red
2 4 6 8
1 2 3 4 5 6 7 8 9 10
Cable BA30-20-06486
1 2 3 4 5 6 7 8 9 10
J107 1 3 5 7 9
2 4 6 8 10
VCC
URAM_PHO LOADX_PHO DRAM_PHO LOADY_PHO VCC MAGU_PHO
4
GND MAGD_PHO GND GND
3
2
1
4 red
3
3 white
GND
2
2 blue
GND
1
1 black
MAGU_PHO
Cable BA33-20-35150
red white blue
C3: Upper electromagnet electromagnet transducer BA30-30-15060
black
J103 VCC MAGD_PHO O H P _ GND M A D D R N N GND D G G
O H P _ M A D D R N U N G G J110
4
J109
4
3
A-8
2
1
4 3 2 1
4 red 3 white 2 blue 1 black
Cable BA33-20-35149
red white blue black
C4: Lower electromagnet transducer BA30-30-15062
Boards Connection Diagrams
2 red Cable BA33-21-35151
1 blue 2 red
Cable BA33-21-35152
1 blue
J104 Power drive board BA34-30-63559 J3
Cable J106 BA30-20-06463 4 red 4 red 4
4
MAGU+ MAGU-
3
MAGD+
2
MAGD-
1
3 blue
3 blue 2 yellow
2 yellow
1 white
1 white
3
MAGU-
2
MAGD+
1
MAGD-
GND MAGD_PHO GND GND
7
3 5
M2: Upper finger magnet BA30-21-06497
blue red
M1: Lower finger magnet BA30-21-06498
blue
J101
2
1
+ D G A M
D G A M
J105
VCC
4 red
4
LOADX_PHO
3
GND
2
GND
1
Cable BA33-21-35153
3 white 2 black
red C1: Manipulator horizontal position transducer transducer BA30-21-06495
white black
1 green
green
J102 VCC
4 red
4
LOADY_PHO
white
2 black
2
C2: Manipulator vertical position transducer transducer BA30-21-06496
black
1 green
1
GND
red
Cable BA33-21-35154
3 white
3
GND
Manipulator connection board BA30-30-06316
green
J108
J43 1
1 + U U G G A A M M
MAGU+
Main control board BA34-30-63557
URAM_PHO LOADX_PHO DRAM_PHO LOADY_PHO VCC MAGU_PHO
2
red
Cable BA30-20-06486
1 2 3 4 5 6 7 8 9 10
2 4 6 8
9
1 2 3 4 5 6 7 8 9 10
J107 1 3 5 7 9
2 4 6 8 10
VCC
URAM_PHO LOADX_PHO DRAM_PHO LOADY_PHO VCC MAGU_PHO GND MAGD_PHO GND GND
4
3
2
4 red
3
3 white
GND
2
2 blue
GND
1
1 black
MAGU_PHO
red
Cable BA33-20-35150
C3: Upper electromagnet electromagnet transducer BA30-30-15060
white blue black
J103 4
VCC MAGD_PHO O H P _ GND M A D D R N N GND D G G
O H P _ M A D D R N N U G G J110
4
1
4
J109
3
2
3 2 1
4 red
red
Cable BA33-20-35149
3 white
C4: Lower electromagnet transducer BA30-30-15062
white
2 blue
blue
1 black
black
1
A-8
Boards Connection Diagrams
Cable BA30-20-06490
4 red 3 white
white
2 black
black
1 green
Main control board BA34-30-63557
1 3
3
2
1
VCC
2 4
5
6
7
8
9
10
1 2 3 4 5 6 7 8 9 10
3
GND
2
GND
1
1 2 3 4 5 6 7 8 9 10
1
VCC PRESSURE_PHO 4 FRONT_PHO BUTTON_S MIDDLE_PHO 3 BACK_PHO VCC GND 2 MIDDLE_PHO GND 1 GND LOADE_PHO J94 GND GND VCC 4
2
3
4
5
6
7
8
9
10
3 white
2 black
GND
1
1 green
J95
DC+ 2
DC1
PRESSURE_PHO
3
GND
2
GND
1
DC+ DCJ97
2
1
J96 1 black
2 red
4
2 red Cable BA30-20-06494
1 blue Cable BA30-20-06464
A-9
DCDC+
red white black
C10: No-cuvette transducer BA30-21-06489
green Cable BA30-20-06488
4 red
C9: Insufficient-cuvette Insufficient-cuvette transducer BA30-21-06492
green Cable BA30-20-06489
2
VCC
black
1 green
GND
red white
2 black
4 red
C8: Cuvette-taking limit transducer BA30-21-06491
green Cable BA30-20-06492
3 white
1
black
1 green
4 red
red white
2 black
3
2
Power drive board BA34-30-63559 DC+ 2 DC- 1
4 red 3 white
LOADE_PHO
J98
J19
Cable BA30-20-06491
J93
J99
Cable BA30-20-06485
4
BACK_PHO Feeder connection board BA30-30-06314
J32 PRESSURE_PHO FRONT_PHO BUTTON_S BACK_PHO VCC MIDDLE_PHO GND LOADE_PHO GND GND
4
C7: Cuvette-pushing limit transducer BA30-21-06490
green J92
J91
red
3 white 2 black 1 green
M5: Feeder motor BA30-21-06494
red white black green
C11: Pressure transducer BA30-21-06488
Boards Connection Diagrams
Cable BA30-20-06490
4 red 3 white
white
2 black
black
1 green 4
3
2
1
VCC BACK_PHO
Main control board BA34-30-63557
Feeder connection board BA30-30-06314
J32 PRESSURE_PHO FRONT_PHO BUTTON_S BACK_PHO VCC MIDDLE_PHO GND LOADE_PHO GND GND
1 3
2 4
5
6
7
8
9
10
1 2 3 4 5 6 7 8 9 10
1 2 3 4 5 6 7 8 9 10
1
2
GND
1
VCC PRESSURE_PHO 4 FRONT_PHO BUTTON_S MIDDLE_PHO 3 BACK_PHO VCC GND 2 MIDDLE_PHO GND 1 GND LOADE_PHO J94 GND GND VCC 4
4
5
6
7
8
9
10
J98
3 white
2
GND
1
1 green
DC+ 2
DC1
4
PRESSURE_PHO
3
GND
2
GND
1
DC+ DCJ97
2
1
J96 1 black
2 red
2 red Cable BA30-20-06494
1 blue Cable BA30-20-06464
A-9
DCDC+
red white black
C10: No-cuvette transducer BA30-21-06489
green Cable BA30-20-06488
4 red
C9: Insufficient-cuvette Insufficient-cuvette transducer BA30-21-06492
green Cable BA30-20-06489
GND
VCC
black
1 green
2 black
red white
2 black
4 red
C8: Cuvette-taking limit transducer BA30-21-06491
green Cable BA30-20-06492
J95
1
black
1 green
4 red
red white
2 black
3 white
2
Power drive board BA34-30-63559 DC+ 2 DC- 1
4 red 3 white
3
LOADE_PHO
J19
3
Cable BA30-20-06491
J93 2
3
4
GND
J99
Cable BA30-20-06485
C7: Cuvette-pushing limit transducer BA30-21-06490
green J92
J91
red
3 white 2 black 1 green
M5: Feeder motor BA30-21-06494
red white black green
C11: Pressure transducer BA30-21-06488
Test Points of Boards
Ap A p p end en d i x B B.1
Test es t Poi Po i n ts o f B o ard ar d s
Test Points on A/D Conversi on Board
TP TP1
TOP/BOTTO TOP/BOTTOM M TOP
Descr ipt ion s The signal at 340nm
TP2
TOP
The signal at 405nm
TP3
TOP
The signal at 450nm
TP4
TOP
The signal at 510nm
TP5
TOP
The signal at 546nm
TP6
TOP
The signal at 578nm
TP7
TOP
The signal at 630nm
TP8
TOP
The signal at 670nm
TP9
TOP
The signal at 700nm
TP10
TOP
The signal of REF
TP11
TOP
The start signals of AD conversion
TP12
TOP
GND
TP13
TOP
The input analog signals of AD conversion
B.2
Test Points on Drive Board
TP
TOP/BOTTO TOP/BOTTOM M
Descr ipt ion s
TP1
TOP
Signal ground (SHIELD) from the main control board
TP2
TOP
5V and 24V ground (GND)
TP3
TOP
5V and 24V ground (GND)
TP4
TOP
12V ground (12VGND)
B.3
Test Points on Reagent Reagent Refri Refri geration Board
TP
TOP/BOTTO TOP/BOTTOM M
Descr ipt ions io ns
TP1
TOP
The temperature switch signal (low)
TP2
TOP
The temperature switch signal (high)
TP3
TOP
Reference ground
TP4
TOP
The switch characteristic of Q1
TP5
TOP
The voltage of the temperature transducer
B.4
Test Points on Main Main Contro l Board
TP
TOP/BOTTO TOP/BOTTOM M
Descr ipt ion s
TP1
BOTTOM&TOP
TxD of Sample Unit
B-1
Test Points of Boards TP
TOP/BOTTO TOP/BOTTOM M
Descr ipt ion s
TP2
BOTTOM&TOP
RxD of Sample Unit
TP3
BOTTOM&TOP
ALE of Sample Unit
TP4
BOTTOM&TOP
TxD of photoelectric photoelectr ic unit
TP5
BOTTOM&TOP
RxD of photoelectric photoelectri c unit
TP6
BOTTOM&TOP
ALE of photoelectric photoelectr ic unit
TP7
BOTTOM&TOP
INT of SPI port
TP8
BOTTOM&TOP
RxD of Main unit
TP9
BOTTOM&TOP
TxD of Main unit
TP10
BOTTOM&TOP
ALE of Main unit
TP11
BOTTOM&TOP
Clock Output of Crystal 16MHZ
TP12
BOTTOM&TOP
TxD of loading/mixing loading/mix ing unit
TP13
BOTTOM&TOP
RxD of loading/mixing loading/mix ing unit
TP14
BOTTOM&TOP
ALE of loading/mixing loading/mix ing unit
TP15
BOTTOM&TOP
GND
TP16
BOTTOM&TOP
VDD 3.3V power
TP17
BOTTOM&TOP
VEE 1.5V power
TP18
BOTTOM&TOP
VTTP +15V power
TP19
BOTTOM&TOP
VTTN -15V power
TP20
BOTTOM&TOP
GNDTT GND of
TP21
BOTTOM&TOP
VCC 5V Power
TP22
BOTTOM&TOP
VPP 12V power
TP23
BOTTOM&TOP
VCCA analog power of 5V
TP24
BOTTOM&TOP
AGND analog GND of 5V
TP25
BOTTOM&TOP
GND
TP26
BOTTOM&TOP
RxD of Sample Unit Extended COM for Bar Code Scanner
TP27
BOTTOM&TOP
TxD of Sample Unit Extended COM for Bar Code Scanner
TP28
BOTTOM&TOP
TxD of Reagent Unit Extended COM for Bar Code Scanner
TP29
BOTTOM&TOP
RxD of Main Unit Extended COM for ISE module
TP30
BOTTOM&TOP
RxD of Reagent Unit Extended COM for Bar Code Scanner
TP31
BOTTOM&TOP
TxD of Main Unit Extended COM for ISE module
TP32
BOTTOM&TOP
Clock Output of Crystal 10MHZ
TP33
BOTTOM&TOP
ALE of temperature unit
TP34
BOTTOM&TOP
TxD of temperature unit
TP35
BOTTOM&TOP
RxD of temperature unit
TP36
BOTTOM&TOP
RxD of Reagent unit
TP37
BOTTOM&TOP
TxD of Reagent unit
TP38
BOTTOM&TOP
ALE of Reagent unit
B-2
±15V
Test Points of Boards
B.5
Test Points Poin ts on PFC PFC Board Bo ard
TP
TOP/BOTTO TOP/BOTTOM M
Descr ipti ip tions ons
TP1
BOTTOM
The switch waveforms of Q3 and Q4
TP2
BOTTOM
The drive waveform of Q5
TP3
BOTTOM
The DC output (390V)
TP4
BOTTOM
The reference voltage of U2 (7.5V)
TP5
BOTTOM
The oscillation oscillat ion waveform of U2 (100KHz)
TP6
BOTTOM
The Vss supply (19V)
TP7
BOTTOM
The VDD supply (14V)
TP8
BOTTOM
The VPP supply (12V)
TP9
BOTTOM
The drive waveform of Q12 (85KHz)
TP10
BOTTOM
The switch waveform of Q12
TP11
BOTTOM
The current waveform of Q12
TP12
BOTTOM
The reference voltage of U8 (5V)
TP13
BOTTOM
The oscillation oscillat ion waveform of U8
TP14
BOTTOM
The 12V(4.5A) supply
B.6
Test Points Poin ts on 5V/1 5V/12V 2V Board Bo ard
TP
TOP/BOTTO TOP/BOTTOM M
Descr ipti ip tions ons
TP1
BOTTOM
The DC output (390V)
TP2
BOTTOM
Input side as reference ground
TP3
BOTTOM
The VDD supply (14V)
TP4
BOTTOM
The oscillation oscilla tion waveform of U301 (100KHz)
TP5
BOTTOM
The reference voltage of U301 (5V)
TP6
BOTTOM
The switch drive waveform
TP7
BOTTOM
The switch waveform
TP8
BOTTOM
The 12V(10A) output
TP9
BOTTOM
The output control voltage of the board
TP10
BOTTOM
The 12V(4.7A) output
TP11
BOTTOM
The supply of FAN (12V)
TP12
BOTTOM
The 5V output of the board
TP13
BOTTOM
Output side as reference ground
TP14
BOTTOM
The switch waveform of U355 (72KHz)
TP15
BOTTOM
Short circuit with the output side reference ground and measure the OVP point of the two 12Vs
TP16
BOTTOM
Short circuit with the output side reference ground and measure the OVP point of the two 5V
B-3
Test Points of Boards
B.7
Test Points Poin ts on 24V 24V Board Bo ard
TP
TOP/BOTTO TOP/BOTTOM M
Descr ipt ion s
TP1
BOTTOM
The DC output (390V)
TP2
BOTTOM
Input side as reference ground
TP3
BOTTOM
The switch waveform
TP4
BOTTOM
The switch drive waveform
TP5
BOTTOM
The current waveform of Q402
TP6
BOTTOM
The VCC supply (14V)
TP7
BOTTOM
The reference voltage of U405 (5V)
TP8
BOTTOM
The oscillation waveform of U405 (100KHz)
TP9
BOTTOM
The power on/off
TP10
BOTTOM
The 24V output
TP11
BOTTOM
Output side as reference ground
TP12
BOTTOM
The voltage waveforms at the two ends of the rectifier rectifi er and fly-wheel diode
TP13
BOTTOM
Short circuit with the output side reference ground and measure the OVP point of the 24V
B-4
P/N: BA34-20-63690 (1.0)