Technical Publications
0459
2155051-2-100 Revis Re vision ion 1
STEREO STER EOTI TIX X 2 om Oper Op erat ator or ma manu nual al do no nott du dupl plic icat atee
Copyrighte 20 2005 05 GE Me Medi dica call Sy Syst stem emss
ATTENTION LES APP APPARE AREILS ILS À RA RAYO YONS NS X SON SONT T DAN DANGER GEREUX EUX À LA FOI FOIS S POU POUR R LE PAT ATIEN IENT T ET POU POUR R LE MAN MANIPU IPULAT LATEUR EUR SI LES MES MESURE URES S DE PRO PROTEC TECTIO TION N NE SON SONT T PAS STR STRICT ICTEME EMENT NTAPPLIQUE PPLIQUEES ES
Bien que cet app Bien appare areil ilsoit con constru struit itselon les nor normes mes de séc sécuri urité téles plu pluss sév sévère ères, s, la sou source rce de ray rayonn onneme ement ntX rep représe résente nteun undanger anger lor lorsqu squee le manipu man ipulat lateur eur est non qua qualif lifié ié ou non nonave averti rti.. Une exp exposi ositio tion n exc excess essive ive aurayonneme ayonnement ntX ent entraî raîne ne des dom dommag mages es à l'o l'orga rganis nisme. me. Parconséquent, onséquent, toute toutess les précau précautions tionsdoive doivent nt être prises prisespour pour évite éviterr que les person personnes nes nonautoriséesou utorisées ou nonqualifiées ualifiées util utilisent isentcet cet apparei appareill créant cré ant ain ainsi si un dan danger ger pou pourr les aut autres res et pou pourr ell elleses-mêm mêmes. es. Avantchaquemanipulation,lespersonnesqualifiéesetautoriséesàseservirdecetappareildoiventserenseignersurlesmesuresdeprotection établies établ iespar laCommissionInter ommission Internati nationale onalede de la la Prot Protectio ectionnRadiologique adiologique,, Anna Annales les26 : Rec Recomman ommandatio dations nsde laCommission ommissionInternationa nternationale lesur la Prot Protectio ectionn Radi Radiologi ologique que et les norme normess nati nationale onaless en vigu vigueur. eur.
WARNING X-RAY EQUIP X-RAY EQUIPMENT MENT IS DAN DANGEROU GEROUS S TO BOTH PA PATIENT TIENT AND OPER OPERATO ATOR R UNLESS UNLE SS MEASUR MEASURES ES OF PROTE PROTECTION CTION ARE STRICT STRICTLY LY OBSER OBSERVED VED
Though this equi Though equipment pment is buil builtt to the high highest eststandardsof elect electrical ricaland andmechanical echanical safety safety,, theuseful x-ray x-raybeam beambecomesa sourc sourcee of dange dangerr in the han hands ds of the una unauth uthori orized zed or unq unqual ualifi ified ed ope operat rator. or. Exc Excess essive ive exp exposu osure re to x-r x-radi adiati ation on cau causes ses dam damage age to hu human man tis tissue sue.. Theref The refore, ore, ade adequa quate te pre precau cautio tions ns mus mustt be tak taken en to prev prevent ent una unauth uthori orized zed or unq unqual ualifi ified ed per person sonss fro from m ope operat rating ing thi thiss equ equipm ipment ent or exp exposi osing ng themse the mselve lvess or oth others ers to its rad radiat iation ion.. Beforeoperation,personsqualifiedandauthorizedtooperatethisequipmentshouldbefamiliarwiththeRecommendationsoftheInternatio nal Commi Commission ssion on Radi Radiologi ological calProtection, rotection, conta contained ined in Anna Annals ls Numb Number er 26 of the ICRP ICRP,, and with appli applicabl cablee natio national nal standa standards. rds.
ATENCION LOS LO S AP APARA ARATOS TOS DE RA RAYOS YOS X SON PEL PELIGR IGROSO OSOS S PARA EL PACI ACIENT ENTE E Y EL OPE OPERAD RADOR OR CUAND CU ANDO O NO SE RES RESPET PETAN AN LAS NOR NORMAS MAS DE PRO PROTEC TECCIO CION N Aunque Aun que est estee apa aparat rato o est estáá con constr struid uido o seg según ún las nor normas mas de seg seguri uridad dadmás est estric rictas tas,, laradiación adiación X con consti stituy tuyee unpeligro eligro alser man manipu ipulad lado o por person per sonas as no aut autori orizad zadas as o nocalifica alificadas das.. Una exp exposi osició ción n exc excesi esiva vaa la rad radiac iación ión X pue puede de cau causar sardaños alorganism rganismo. o. Porconsiguiente,sedebentomartodaslasprecaucionesnecesariasparaevitarquelaspersonasnoautorizadasonocalificadasusenesteapara to,, lo qu to quee ser sería ía un pe peli ligr gro o pa para ra l os de demá máss y pa para ra sí mi mism smas as.. Antes An tes de efe efectu ctuar ar las man manipu ipulac lacion iones, es, las per person sonas as aut autori orizad zadas as y cal califi ificad cadas as en el uso de est estee apa aparat rato, o, deb deben en inf inform ormars arsee sob sobre re las nor normas mas de protección protec ciónfijada fijadass por la Comis Comisión ión Intern Internacion acional al de la Prot Protecció ecciónnRadiológica, adiológica,AnalesNo 26:Recomendació ecomendaciónes nesde laComisiónIntern omisión Internacion acional al sobre la Prot Protecció ecciónn Radi Radiológi ológica ca y normas normasnaci nacionale onales. s.
ACHTUNG RÖNTGENAPP RÖNTG ENAPPAR ARATE ATE SIND EINE GEF GEFAHR AHR FÜR PA PATIENT TIENTEN EN SOWIE SOWIEBEDIE BEDIENUNGS NUNGSPERSO PERSONAL, NAL, WENN DIE GEL GELTENDE TENDEN N SICHER SICHERHEITSV HEITSVORKEH ORKEHRUNGE RUNGEN N NICHTGENAU BEACHT BEACHTET ET WERDE WERDEN N
Dieser Appara Apparatt entsp entspricht richt in sein seiner er Bauwe Bauweise ise stren strengsten gsten elekt elektrisch rischen en und mecha mechanisc nischen hen Siche Sichereitsn reitsnormen ormen,, doch in den Hände Händenn unbe unbefugte fugterr oderr un ode unqua qualif lifizi iziert erter er Pe Person rsonen en wir wird d er zu ein einer er Gef Gefahr ahrenq enquel uelle. le. Übe Übermä rmäßig ßigee Rö Röntg ntgenb enbest estrah rahlun lungg ist für den men mensch schlic lichen hen Org Organi anismu smuss schädlich. DeswegensindhinreichendeVorsichtsmaßnahmenerforderlich,umzuverhindern,daßunbefugteoderunqualifiziertePersonensolcheGeräte bedienen bedie nenoder oder sich selbs selbstt und ander anderee Pers Personen onenderen derenBestrahlung estrahlung ausse aussetzen tzen könn können. en. VorInbetriebnahm nbetriebnahmeediesesApparatssolltesichdas quali qualifizie fizierte rteundbefugteBedienungs edienungsperson personal almitden gelt geltenden endenKrite Kriterien rienfürden gefah gefahrlo rlo senStrahleneinsa trahleneinsatz tzdurch sorgfä sorgfältig ltiges es Studi Studium umdes Hefts Nr.26 derInternationale nternationalenn Kommi Kommission ssionfür fürStrahlenschut trahlenschutzz (ICRP (ICRP)) vertra vertraut utmachen: Empfehlun Empfe hlungen gender der Inter Internatio nationale nalenn Kommi Kommission ssionfür Strah Strahlensc lenschutz hutzund und andere andererr nati nationale onalerr Norme Normenbeh nbehörden. örden.
ATTENZIONE GLI APP APPAR ARECC ECCHI HI A RAG RAGGI GI X SON SONO O PER PERICO ICOLO LOSI SI SIA PER IL PAZI AZIENT ENTE, E, SIA PER L'O L'OPER PERAT ATORE ORE,, A MEN MENO O CHE NON VEN VENGAN GANO O ST STABI ABILIT LITII FA FATTOR TTORII DI SIC SICUR UREZZ EZZA A CON CONTRO TRO LE ESP ESPOSI OSIZIO ZIONI NI E CHE DEV DEVONO ONO ESS ESSERE ERE OS SERVATI SERV ATI CON ESTR ESTREMA EMA RIGOR RIGOROSIT OSITÀ. À. Sebbene que Sebbene questo sto app appara arato to sia cos costru truito ito con confor formem mement ente e agl aglii sta standa ndard rd di sicu sicurezz rezza a ele electt cttric ricii e mec meccan canici ici più ele elevat vati, i, il fas fascio cio rad radioge iogeno no utiledivent utile diventa a una sorgen sorgente te di pericol pericolo o quand quandunque unqueusata usatada person personale ale non autori autorizzato zzato e per giutaincompetente. ncompetente. Unaeccessivaesposi zione radioge radiogena, na, danne dannegia gia sicura sicuramente mente il tessut tessuto o umano umano.. Perciò, Perci ò,dovranno essereprese adegu adeguate ate precau precauzioni zioniper rende rendere re imposs impossibile ibile a person persone e non autor autorizzate izzatee non qualif qualificate, icate,l'uso l'usodi questi appare app arecch cchi, i, ed inol inoltre tre che ess esse e ste stesse sse ed alt altre re evit evitino ino que queste ste rad radiaz iazioni ioni.. Primadell'uso,lepersonequalificateequindiautorizzateall'usodiquestiapparecchi,dovrannoconoscereafondoifattoridiesposizione di sicure sicurezza zza stabil stabiliti, iti,tramite unostudio attent attento o della dellacommiss commissione ioneinternazionale nternazionaleper la protezi protezione oneradiologica(ICRP), al numer numero o 26 degli annalii : raccom annal raccomandazi andazioni oni della dellacommiss commissione ione intern internaziona azionale le sulla sullaprotezio protezione ne radiol radiologica ogica ed altrenormative standa standard rd di livell livello o nazion nazionale. ale.
STEREOTIX 2
GE Medical Systems REV 1
om 2155051–2–100
REGULATORY REGULAT ORY REQUIREMENTS REQUIREMENT S
This product complies with the regulatory requirements of the following:
D
Council Directive 93/42/EEC concerning medical devices when it bears the following CE marking of conformity. 0459
For a system, the location of the CE marking label is described in the system manual. European registered place of business: GE Medical Systems Europe Quality Assurance Manager BP 34 F 78533 BUC CEDEX France
Tel: +33 1 30 70 40 40
D
Green QSD 1990 Standard issued by MDD (Medical Devices Directorate, Department of Health, UK).
D
Medical Device Good Manufacturing Practice Manual issued by the FDA (Food and Drug Administration, Department of Health, USA).
D
Underwriters’ Laboratories, Inc. (UL), an independent testing laboratory.
D
Canadian Standards Association (CSA).
D
International Electrotechnical organization, when applicable.
D
General Electric Medical Systems is ISO 9001 certified.
D
The original document was written in English.
i
Commission
(IEC),
international
standards
STEREOTIX 2
GE Medical Systems REV 1
om 2155051–2–100
ELECTROMAGNETIC COMPATIBILITY (EMC) This equipment complies with IEC60601–1–2 Edition 2 EMC standard for medical devices. This equipment generates, uses, and can radiate radio frequency energy. energy. The equipment may cause radio frequency interference to other medical and non-medical devices and radio communications. To provide reasonable protection against such interference, interference, this product complies with radiated emissions as per CISPR11 Group 1, Class A standard limits. Detailed requirements and recommendations about power supply distribution and installation are listed in the Pre–Installation Manual (pim) shipped with your system. However, there is no guarantee that interference will not occur in a particular installation. If this equipment is found to cause interference (which may be determined by turning the equipment on and off), the user (or qualified service personnel) should attempt to correct the problem by one or more of the following measure(s):
D
reorient or relocate the affected device(s)
D
increase the separation between the equipment and the affected device
D
power the equipment from a source different from that of the affected device
D
consult the point of purchase or service representative for further suggestions
The manufacturer is not responsible for any interference caused by using other than recommended interconnect cables or by unauthorized changes or modifications to this equipment. Unauthorized changes or modifications could void the users’ authority to operate the equipment. All interconnect cables to peripheral devices must must be shielded and properly grounded. Use of cables not properly shielded and grounded may result in the equipment causing radio frequency interference. Do not use devices which intentionally transmit RF S ignals (Cellular Phones, Transceivers, or Radio Controlled Products) in the vicinity of this equipment as it may cause performance outside the published specifications. Recommended separation distances are detailed in the Pre–Installation Manual (pim) shipped with your system. Keep the power to this type of devices turned off when near this equipment. The medical staff in charge of this equipment is required to instruct technicians, patients, and other people who may be around this equipment to comply fully with the above requirement. Further data and recommendations for meeting Electromagnetic Compatibility requirements for a typical installation are given in the Pre–Installation Manual (pim) shipped with your system. Note that the magnetic field of an MRI device located nearby may cause a risk of interference. Magnetic field amplitude limits are specified in the Pre–Installation Manual (pim) shipped with your system.
ii
STEREOTIX 2
GE Medical Systems REV 1
om 2155051–2–100
RECYCLING:
Machines or accessories at end-of-life: The elimination of machines and accessories must be in accordance with national regulations for waste processing. All materials and components that could pose a risk to the environment must be removed from the end–of–life machines and accessories (examples: dry and wet cell batteries, transformer oil, etc...). Please consult your local GEMS representative before discarding these products.
Packing materials: The materials used to pack our equipment are recyclable. They must be collected and processed in accordance with the regulations in force for the country where the machines or accessories are unpacked.
iii
STEREOTIX 2
GE Medical Systems REV 1
om 2155051–2–100
Blank page
iv
STEREOTIX 2
GE Medical Systems
REV 1
om 2155051-2-100
SERVICE OFFICES IN EUROPE
For service issues, contact the Service office in your country or in France.
AUSTRIA GE GesmbH Medical Systems Austria Prinz Eugen Strasse 8/8 A–1140 WIEN
TEL: FAX: TLX:
0660 8651 toll free +43 1 505 38 74 136314
BELGIUM GE Medical Systems Benelux Gulkenrodestraat 3 B–2160 WOMMELGEM
TEL: FAX: TLX:
+32 3 320 12 11 +32 3 320 12 59 72722
TEL: FAX:
+45 45 51 00 55 +45 42 42 59 89
TEL:
+33 1 30 70 40 40
TEL: FAX: TLX:
+49 69 9760 7380 +49 69 7682 237 412002 GEGF D
TEL: FAX:
+30 1 93 24 582 +30 1 93 58 414
TEL: FAX: TLX:
+39 39 20 881 +39 39 73 37 86 3333 28
TEL: FAX:
+31 304 79711 +31 304 11702
TEL: FAX:
+48 2 625 59 62 +48 2 615 59 66
DENMARK GE Medical Systems Danmark Skovlytoften 4 DK–2840 HOLTE FRANCE GE Medical Systems 283 rue de la Minière BP34 F–78533 BUC CEDEX GERMANY GE Medical Systems Deutschland GmbH & Co. KG Praunheimer Landstrasse 50 D–60488 FRANKFURT AM MAIN GREECE GE Medical Systems Hellas 41 Nikolaou Plastira Street GR–171 21 NEA SMYRNI ITALY GE Medical Systems Italia Via Monte Albenza 9 I–20052 MONZA NETHERLANDS GE Medical Systems Nederland B.V. Atoomweg 512 NL–3542 AB UTRECHT POLAND GE Medical Systems Polska Krzywickiego 34 P–02–078 WARSZAWA
v
STEREOTIX 2
GE Medical Systems
REV 1
om 2155051-2-100
PORTUGAL GE Medical Systems Portuguesa S.A. Rua Sa da Bandeira, 585 Apartado 4094 P–4002 PORTO CODEX
TEL: FAX: TLX:
+351 2 2007696/97 +351 2 2084494 22804
TEL: FAX: TLX:
+7 095 956 7037 +7 502 220 3259 613020 GEMED SU
TEL: FAX:
+34 1 663 25 00 +34 1 663 25 01
SWEDEN GE Medical Systems PO–BOX 1243 S–16428 KISTA
TEL: FAX: TLX:
+46 87 50 57 00 +46 87 51 30 90 12228 CGRSWES
SWITZERLAND GE Medical Systems (Schweiz) AG Sternmattweg 1 CH–6010 KRIENS
TEL: FAX:
+41 41 425577 +41 41 421859
TEL: FAX:
+90 212 75 5552 +90 212 211 2571
TEL: FAX:
+44 753 874000 +44 753 696067
RUSSIA GE VNIIEM Mantulinskaya UI. 5A 123100 MOSCOW
SPAIN GE Medical Systems España Avda. de Europa 22 Parque Empresarial la Moraleja E–28100 ALCOBENDAS
TURKEY GE Medical Systems Turkiye A.S. Mevluk Pehliran Sodak Yilmaz Han, No 24 Kat 1 Gayretteppe ISTANBUL UNITED KINGDOM IGE Medical Systems Coolidge House 352 Buckingham Avenue SLOUGH Berkshire SL1 4ER
vi
STEREO STE REOTIX TIX 2
GE Medical Systems REV 1
om 2155051-2-100
CONTENTS CHAPTER
1
TITLE REGULATORY REQUIREMENTS SERVICE OFFICES IN EUROPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . TABLE OF CONTENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . FOREWORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . REVISION HISTORY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
i v v ii x xi
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-1
SECTION 1 - PRESENTATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-3
SECTION 2 - DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1-3
2-1 2-2 2-3 2
PAGE
Puncture Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-7 Controller Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-9 Deviation Readout Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-11
OPERATING MODES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-1
SECTION 1 - THEORY OF OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2-3
SECTION 2 - INSTALLING STEREORIX 2 ON SENOGRAPHE DMR . . . . . . .
2-5
2-1
2-2
2-3 2-4
STEREOTIX 2 installed on Senographe DMR . . . . . . . . . . . . . . . . . . . . . . . . 2-1-1 Removing Ac Accessories fr f rom Se S enographe DM D MR . . . . . . . . . . . . . . . . . 2-1-2 Preparing Senographe DMR . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1-3 Securing ST STEREOTIX 22 to to Se Senographe DM D MR . . . . . . . . . . . . . . . . . 2-1-4 Starting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Exposure with Phantom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2-1 Positioning the Phantom . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2-2 Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2-3 Film . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using th the Co Controller Un Unit an and th the Pu Puncture De Device . . . . . . . . . . . . . . . . . . . . 2-3-1 Using the Controller Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3-2 Using the Puncture Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Selecting a Second Point . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
vii
2-5 2-7 2-7 2-9 2-9 2-11 2-11 2-13 2-15 2-17 2-17 2-21 2-22
STEREO STE REOTIX TIX 2
GE Medical Systems REV 1
om 2155051-2-100
CONTENTS (C (CON ONT' T'D) D) CHAPTER
3
TITLE
EXAM PREPARATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-1
SECTION 11 - - MA MAKING A A PU PUNCTURE ON ON A A PA PATIENT . . . . . . . . . . . . . . . . . . . .
3-3
1-1
Types of Breast Puncture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1-1 Cytologic Puncture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1-2 Histology Puncture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1-3 Preoperative Localization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3-3 3-3 3-3 3-4 3-4
SECTION 22 - - MA M AKING A A PU P UNCTURE US U SING ST S TEREOTIX 2 . . . . . . . . . . . . . .
3-7
1-2
2-1
PO P OSITIONING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1-1 Positioning STEREOTIX 2 on Senographe DMR . . . . . . . . . . . . . . 2-1-2 Positioning the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Operating Difficulties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2-1 X-ray Image . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2-2 Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2-3 Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Making Stereoscopic Exposures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3-1 Positioning the Breast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3-2 Exposure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3-3 Film Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Using th the Co Controller Un Unit an and th the Pu Puncture De Device . . . . . . . . . . . . . . . . . . . . 2-4-1 Using the Controller Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4-2 Using the Puncture Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Selecting a Second Suspect Area (second lesion to be punctured) . . . . . . . . .
3-7 3-7 3-11 3-12 3-12 3-12 3-12 3-13 3-13 3-14 3-14 3-15 3-15 3-16 3-18
MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-1
SECTION 1 - CLEANING AND DISINFECTION . . . . . . . . . . . . . . . . . . . . . . . . . .
4-3
SECTION 2 - PREVENTIVE MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-5
SECTION 3 - MESSAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-5
2-2
2-3
2-4 2-5 4
PAGE
viii
STEREO STE REOTIX TIX 2
GE Medical Systems REV 1
om 2155051-2-100
CONTENTS (C (CON ONT' T'D) D) CHAPTER
5
6
TITLE
PAGE
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-1
SECTION 1 - PHYSICAL SPECIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-3
SECTION 2 - ELECTRICAL & TECHNICAL SPECIFICATION . . . . . . . . . . . . . .
5-3
2-1 2-2 2-3
Technical Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Radiographic Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Electrical Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-3 5-5 5-5
SECTION 3 - AMBIENT CONDITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-5
SECTION 4 - MEANING OF SYMBOLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-6
REGULATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
6-1
ix
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
FOREWORD This manual is provided for STEREOTIX 2 operators. It is designed to supply all the information required for the correct use of this equipment.
x
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
REVISION HISTORY REV 1
DATE
TYPE OF MODIFICATION
June 2005
New release based on document 2155051-100 rev. 1, with the following changes: - EURge 16438: Regulatory Requirements section updated for EMC installation recommandations IEC 601-1-2 Ed. 2. - WEEE Compliance: Chapter 5 Specifications, added Section 4 Meaning of Symbols, including symbol and text for waste of electrical and electronic equipment.
LIST OF EFFECTIVE PAGES PAGE
REVISION
Title page
1
i thru xii
1
1-1 thru 1-12
1
2-1 thru 2-22
1
3-1 thru 3-18
1
4-1 thru 4-8
1
5-1 thru 5-6
1
6-1 thru 6-2
1
PAGE
REVISION
xi
PAGE
REVISION
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
Blank page
xii
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 1 - INTRODUCTION N O I T C U D O R T N I
1-1
STEREOTIX 2
GE Medical Systems REV 1
N O I T C U D O R T N I
om 2155051-2-100
ILLUSTRATION 1-1 STEREOTIX 2
1
3 2 X 0.0
XI
I O
X Y Z
Y 0.0
Z 0.0
0
L1
0
-
XA
L2
L3 +
XB YAB
1-2
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
SECTION 1 PRESENTATION The three-dimensional stereotaxy, STEREOTIX 2, is an optional unit for the Senographe DMR. This stereotaxy device computes the coordinates of a point reproduced on two stereotactic views taken at angles of +15 _ and -15_ on the same film. The device allows the physician to locate in the three directions (X, Y and Z) of a radiographic image with suspected lesion previously detected on mammograms. STEREOTIX 2 permits cytologic and histologic puncture and/or preoperative localization of mammary lesions, irrespective of their position in the breast. SECTION 2 DESCRIPTION See Illustrations 1-1 and 1-2. The STEREOTIX 2 assembly consists of three units, interconnected by cable:
Note:
D
Puncture Device (see Illustration 1-1, 1 ),
D
Controller Unit (Illustration 1-1, 2 ),
D
Deviation Readout Unit (Illustration 1-1, 3 ). The CE marking label is located on the back side of the Puncture Device.
1-3
N O I T C U D O R T N I
STEREOTIX 2
GE Medical Systems REV 1
N O I T C U D O R T N I
om 2155051-2-100
ILLUSTRATION 1-2 ACCESSORIES
1
2
0
9
1
4
1
1
1-4
0
8
1
2
2
25
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
To following accessories are supplied: D
Precision test phantom (Illustration 1-2, 1 ),
D
Six needle-guides (sterilized and one-time use) with the following internal diameters (Illustration 1-2, 2 ).
TABLEAU 1-1 NEEDLE-GUIDES
Needle Needle-guide
0.8 0.9
1-5
0.9 1.0
Diameter (mm) 1.1 1.2 1.2 1.4
1.6 1.8
2.0 2.25
N O I T C U D O R T N I
STEREOTIX 2
GE Medical Systems REV 1
N O I T C U D O R T N I
om 2155051-2-100
ILLUSTRATION 1-3 PUNCTURE DEVICE
8 1
2
7
3
6
5
4
1-6
STEREOTIX 2
GE Medical Systems REV 1
2-1
om 2155051-2-100
Puncture Device See Illustration 1-3 The Puncture Device is a self-contained removable assembly attached to the film take-up arm of the Senographe DMR in place of the magnification plate.
Rear locating rods 7 , and side fasteners 6 secure this assembly to theSenographe (see Chapter 2, Para 2-1-3, Securing STEREOTIX 2 to Senographe DMR. The Instrument Unit comprises: D
Plate 4 used to position the breast and which can receive an 18/24-cm cassette. Two metal letters, R and L (right and left) and a metal locating inverted T are included in the magnification plate. The cassette moves from right to left and from left to right in the tunnel of the magnification plate.
D
Manual compression arm 1 equipped with aperture compression paddle 3 . To compress, use the knobs 5 located on either side of the compression arm. The window opens to 54 mm on the X-axis, and to 43 mm on the Y-axis.
Note:
When the compression paddle leaves the park position (High position) to compress the breast, the Up/Down and rotation movements of the Senographe DMR examination arm are inhibited. To tilt the tube arm support, use the locking lever located at the rear of the column. D A manually-adjusted needle-holder
2 can be moved in three directions
(X, Y and Z) via knobs 8 located on either side of the needle-holder. Needle-holder travel is: 50 mm ±0.5 mm along X-axis 40 mm ±0.5 mm along Y-axis 70 mm ±0.5 mm along Z-axis The needle-holder receives the sterilized, disposable needle-guides. Note:
There is a needle-holder for each needle size. The needle-guide contains a hole bored to a diameter suited to its needle (see Table 1-1).
1-7
N O I T C U D O R T N I
STEREOTIX 2
GE Medical Systems REV 1
N O I T C U D O R T N I
om 2155051-2-100
ILLUSTRATION 1-4 CONTROLLERUNIT
1
2
3
4
I
X
Z
Y
O
5
X Y Z
6
7
0
L1
0
-
19
8
L2
9
L3 +
12
11
10
13
14 20
XI
X
I
Y
O
Z
0
L1
0
-
L2
L3 +
XA
XB
YAB
18
17
16
1-8
15
STEREOTIX 2
GE Medical Systems REV 1
2-2
om 2155051-2-100
Controller Unit See Illustration 1-4. The Controller Unit consists of: D
Control and Display Console 1 used to compute the coordinates from a point marked on the film, and to select and/or adjust the needle length. See Illustration 1-4
2
ON switch.
3
OFF switch.
4
X coordinate readout. Y coordinate readout. Z coordinate readout.
D
5
RESET.
6
ZERO SELECT.
7
No. 1 needle length.
8
No. 2 needle length.
9
No. 3 needle length.
10
Increment in needle length by 1 mm.
11
Needle length readout.
12
Decrement needle length by 1 mm.
Viewbox 18 which can receive 18 x 24 cm format film. The viewbox is equipped with two cross-hair cursors, X and Y, which can be moved via three knobs, XA, XB and YAB, located on the top of the unit on either side of the light plate. A line drawn on the light plate serves as a horizontal reference and is used to position the film.
13
Left cross-hair cursor movement knob, X axis (absissa).
14
Right cross-hair cursor movement knob, X axis (absissa).
15
Right and left cross-hair cursor movement knob, Y axis (ordinate).
16 and 17 Note:
Right and left cross-hair cursor which move on X and Y axes.
On an exposure, the ordinate is identical for the same point appearing on the righthand view and on the lefthand view.
1-9
N O I T C U D O R T N I
STEREOTIX 2
GE Medical Systems REV 1
N O I T C U D O R T N I
om 2155051-2-100
ILLUSTRATION 1-5 DEVIATION READOUT UNIT
X 0.0
Y 0.0
1-10
Z 0.0
STEREOTIX 2
GE Medical Systems REV 1
2-3
om 2155051-2-100
Deviation Readout Unit See Illustration 1-5 This unit is installed on the Senographe DMR column readout or on the STEREOTIX 2 Controller Unit. It displays the deviation between the X, Y and Z coordinates of the needle-holder before it is positioned, and the X, Y and Z coordinates of the suspect exposure (lesion to be punctured). While the operator is turning the needle-holder knobs or adjusting the needlelength, the Deviation Readout Unit shows the variation in the deviation in real time.
1-11
N O I T C U D O R T N I
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
Blank page N O I T C U D O R T N I
1-12
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 2 - OPERATING MODES
G N I S T E A D R O E M P O
2-1
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
Blank page
G N I S T E A D R O E M P O
2-2
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 2 - OPERATING MODES SECTION 1 THEORY OF OPERATION STEREOTIX 2 computes the coordinates of a suspect x-ray image from two stereotactic views of the same point taken at angles of +15 _ and -15_ on the same film. The SID value is known. According to the length of the needle selected, the STEREOTIX also computes the difference between the X, Y and Z coordinates of the suspect exposure on the stereoscopic exposures and the X, Y and Z coordinates of the needle-holder beforeit is positioned. The deviation found between the respective coordinates of the image and the needle-holder along with the variations of this deviation are displayed on the Deviation Readout Unit. Note:
To vary the deviation between the X, Y and Z coordinates of the image and of the needle-holder, adjust the needle-holder positioning knobs and change the needle length selection. Having selected the needle length (with or without adjustment), the operator moves the needle-holder via the knobs along the three axes to reduce the value displayed on the Deviation Readout Unit to zero. The needle-holder is now positioned over the lesion to be punctured in the breast. Place the needle-guide in the needle-holder and insert the puncture product.
2-3
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 2-1 STEREOTIX 2 INSTALLEDON SENOGRAPHE DMR
0° - 15°
+ 15°
- 27°
+ 27°
G N I S T E A D R O E M P O
2-4
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
SECTION 2 INSTALLING STEREORIX 2 ON SENOGRAPHE DMR 2-1
STEREOTIX 2 installed on Senographe DMR See Illustration 2-1 Note:
The Senographe DMR Examination Arm carries Silk-screened user markings with the following color code: Blue:
Top holes and left middle hole, anchoring, SID: 510 mm Tube-arm support tilt: +/- 15_
Orange:
Tube-arm support tilt: +/- 27 _
2-5
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 2-2 PREPARINGSENOGRAPHE DMR TO RECEIVE STEREOTIX 2
7
G N I S T E A D R O E M P O
1
2
510
3
560 610
4 1 2 3
5
6
2-6
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
2-1-1 Removing Accessories from Senographe DMR See Illustration 2-2 1. Remove face shield 1 . 2. Remove compression paddle 2 by sliding it sideways. 3. Remove Bucky 4 as follows: a.
Open flap located 3 on rear side of film-holder assembly 5 .
b.
Lift the lever.
c.
Hold Bucky grid by its sides and remove.
2-1-2 Preparing Senographe DMR See Illustration 2-2 1. Move compression paddle to maximum High position.
WARNING
THE TURN-KNOBS 7 USED FOR THE MANUAL COMPRESSION ADJUSTMENT IN STANDARD MAMMOGRAPHY (SEE SENOGRAPHE DMR OPERATOR MANUAL) MUST NOT BE OPERATED WHEN THE STEREOTIX 2 IS BEING USED. 2. Via two levers 6 , located one on either side of receptor-holder, reduce SID to 510 mm (SID 510 = Blue color code). Push lever 6 first to engage the mechanism, then turn lever 6 frontward to release mechanism and change SID.
WARNING
SID510 IS RESERVED FOR THE INSTALLATION OF STEREOTIX 2. AT SID 510 MM AND WITHOUT THE STEREOTIX 2, THE X-RAY EXPOSURE IS INHIBITED AND A MESSAGE APPEARS ON THE CONSOLE READOUT. 3. Place receptor-holder photocell in frontal position (blue color code).
WARNING
THE PHOTOCELL MUST BE PLACED IN THE FRONTAL POSITION (BLUE COLOR CODE) WHEN THE STEREOTIX 2 IS BEING USED. FAILURE IN DOING SO MAY RESULT IN INADEQUATE FILM QUALITY AND UNWANTED EXPOSURE.
2-7
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 2-3 SECURING STEREOTIX 2 IN POSITION; CONTROLLER UNIT START-UP
1
2 G N I S T E A D R O E M P O
3
4
5
1
2
3
6 I O
X Y Z
0
L1
0
-
XA
L2
L3 +
XB YAB
2-8
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
2-1-3 Securing STEREOTIX 2 to Senographe DMR See Illustration 2-3 1. Move compression paddle of Instrument Unit to low position. 2. Hold Instrument Unit, and engage pins 1 of rear section in holes marked on Senographe DMR receptor-holder arm 2 (top holes 3 andleftmiddlehole
4 : Blue color code). 3. Lock lateral fasteners 5 by folding them down on sides of receptor-holder arm.
2-1-4 Starting Up See Illustration 2-3 Press Note:
I , 6 to switch on Controller Unit.
The Controller Unit autotests at switch-on. The automatic movement of the cross-hair cursors is part of this procedure.
2-9
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 2-4 INSTALLING THE PHANTOM; TUBE-ARM SUPPORT ANGLES
0° - 15°
+ 15°
- 27°
+ 27°
G N I S T E A D R O E M P O
1 1 2
3
2-10
STEREOTIX 2
GE Medical Systems REV 1
2-2
om 2155051-2-100
Exposure with Phantom See Illustration 2-4 The following accuracy test must be made before using STEREOTIX 2 on a patient.
2-2-1 Positioning the Phantom The phantom 3 is positioned under the compression paddle 2 so that the holes face the opening in the paddle. Compress the phantom by turning the knobs 1 located one on either side of the compression arm of the Instrument Unit. Note:
When the compression paddle leaves the park (High) position, the Up/Down and tilting movements of the Senographe DMR are inhibited. To move the Tube-Arm Support, press the trigger lever located at its rear (see Senographe DMR om, Ill 1-3). This releases the Tube-Arm Support from the receptor-holder arm. The silk-screened indicators show the position of the tube-arm support when it is tilted to either side of the Examination Arm. The color code is: Blue: Orange:
+/-15 _ +/-27_
Four angle shifts are possible: +15_ thru -15_ to make stereoscopic views. +27_ thru -27_ to install the needle (exposure is inhibited).
CAUTION
When the tube-arm support is being tilted to +/-15_ (color code: Blue), exposure is permitted only with the STEREOTIX 2 installed. In all other cases, exposure is inhibited and a warning message appears on the Control Console of the Senographe DMR.
WARNING
WHEN THE X-RAY TUBE ARM IS BEING UNCOUPLED FROM THE RECEPTOR ARM SUPPORT, DO NOT ALLOW OBSTACLES BETWEEN THE TWO ARMS, AND KEEP YOUR HANDS ON THE LEVERS DURING THE WHOLE MOVEMENT.
2-11
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 2-5 EXPOSURE AND FILM
- 15°
0°
G N I S T E A D R O E M P O
+ 15°
1
2
3
4
2-12
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
2-2-2 Exposure See Illustration 2-5 THE AOP (AUTOMATIC OPTIMIZATION PARAMETERS) MODE CANNOT BE USED WITH THE STEREOTIX 2 INSTALLED.
WARNING
1. Place an 18/24 film cassette 3 in film-holder plate of Instrument Unit 4 . Note:
The same film cassette will be used to make the two exposures, left and right. 2. The tube and the film cassette must be positioned on the same side. You can start on either side. We start on the righthand side. D
FIRST EXPOSURE
1. Insert film cassette into righthand side Y of film-holder plate. 2. Tilt tube 1 to +15_ on righthand side (color code: Blue) Y by pressing lever
2 of the Tube-Arm Support and by pushing the arm gently. When lock is engaged, release lever. 3. Select rad parameters. 4. Make an exposure. SECOND EXPOSURE 1. Push film cassette to lefthand side of film-holder plate F.
D
2. Tilt tube to -15_ on lefthand side (color code: Blue) F by pressing lever of the tube-arm support and by pushing the arm gently. When lock is engaged, release lever. 3. Select rad parameters. 4. Make an exposure. 5. Remove film cassette and develop film.
WARNING
THE PHANTOM SHOULD NOT BE MOVED.
2-13
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
Blank page
G N I S T E A D R O E M P O
2-14
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
2-2-3 Film The film comprises two exposures, left and right.
FILM
ILLUSTRATION 2-6
G N I S T E A D R O E M P O
L
RL
R
Each exposure contains: D
Inverted T reference mark.
D
Letters L and R (left and right). Projection of phantom copper plugs. Projection of compression paddle aperture.
D D
Note:
The projection of the copper disks contained in thephantom should appear in the projection of the compression paddle aperture.
2-15
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 2-7 USING THE CONTROLLER
6
X
.
XI
G N I S T E A D R O E M P O
I O
2
Y
1
Z
.
.
X Y
0
L1
Z
0
-
L2
L3
2
+
XA
XB YAB
3 2
4
X
6
Y
.
XI
I O
X Y Z
Z
.
0.0 0.0 0.0
.
0
L1
0
-
L2
L3 +
5 XA
XB YAB
2-16
STEREOTIX 2
GE Medical Systems REV 1
2-3
om 2155051-2-100
Using the Controller Unit and the Puncture Device
2-3-1 Using the Controller Unit 1. Press RESET 1 twice. The X, Y and Z 6 readouts go off. 2. Slide film 4 along light plate 3 and under cross-hair cursors, and secure it with pressure rollers. 3. Position film with inverted Ts of film superposed on horizontal line of coordinates of light plate coordinates. 4. Position cross-hair cursors on inverted T of each exposure by turning XA, XB, and YAB knobs 2 (located on either side of light plate). Note:
The Y-ordinate is identical for a same point on the lefthand and righthand exposures. 5. Press ZERO SELECT key 5 to validate location of inverted Ts. The X, Y and Z readouts 6 light, each displaying 0.0.
2-17
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 2-8 USING THE CONTROLLER UNIT (continued)
3
X
XI
G N I S T E A D R O E M P O
Y
.
I O
X Y Z
Z
.
0.2 13.7 32.9
.
0
L1
0
-
L2
L3
3
+
XA
XB YAB
*
* 3
X
Y
Z
- 8.5 - 17.5 -20.6
1 XI
I O
X Y Z
0.2 13.7 32.9
0 0
L1 -
XA
L2
80
L3 +
XB
2 YAB
*
*
2-18
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
See Illustration 2-8 6. Position cross-hair cursors XA, XB and YAB on each of the selected disk projections (one per exposure, lesion to be punctured, by turning the XA, XB, and YAB knobs
WARNING
Note:
3 .
THE USER MUST ACCURATELY SELECT THE SAME POINT SEEN ON THE TWO EXPOSURES. THE XA, XB, AND YAB DEVIATIONS MAY LEAD TO DEVIATIONS OF ONE THRU SEVERAL MILLIMETERS WHEN THE NEEDLE IS BEING INSERTED IN THE PHANTOM OR IN THE LESION. The Y-ordinate is identical for a given point appearing on the lefthand and righthand exposures. The computer reads the deviation between the original coordinates (marked with an inverted T on film) and those of the lesion to be punctured. The deviation values are displayed on the X-Y-Z readout of the Controller Unit. 7. Press L1, L2 or L3 1 to select needle length. 8. Needle length can be adjusted in steps of 1 mm via the + key or - key 2 located on either side of the needle length readout. Hold one of the keys down for fast incrementation or decrementation. 9. Release the key to validate the length selected and to hold it in memory even when power is removed from the STEREOTIX 2.
Note:
Needle lengths are preprogrammed at installation: L1 = 70 mm: L2 = 80 mm; L3 = 90 mm. The user can modify this program via the + key or - key 2 . Releasing the key validates the length selected and holds it in memory even during STEREOTIX 2 switch-off. The needle length can be set to a depth of 60 mm thru 120 mm via its key (L1, L2 or L3, 1 ). The readout will not display less than 60 or more than 120.
2-19
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 2-9 USING THE PUNCTURE DEVICE
- 27°
G N I S T E A D R O E M P O
4
1
3
2
X
Y
Z
- 8.5 - 17.5 - 20.6
2-20
X
Y
Z
0.0
0.0
0.0
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
2-3-2 Using the Puncture Device See Illustration 2-9 The Deviation Readout Unit shows the deviation between the needle-holder 3 and the point selected in the phantom 2 (lesion to be punctured). 1. Turn the X, Y, and Z 1 of the needle-holder until the deviation shown by the readout is reduced to zero at X, Y, and Z. The needle-holder is now opposite the point selected in the phantom (point to be punctured). 2. Move the tube-arm support 4 to the righthand or lefthand side at + or - 27 _ (color code: Orange) by pressing the trigger lever of the tube-arm support and by pushing the arm gently. When the lock is engaged in the notch, release the lever. The space above the needle-holder is thereby free and facilitates the installation of the needle and needle-holder.
CAUTION
At +27_ and - 27_ tilt, exposure is inhibited and a warning message appears on the Senographe DMR Control Console readout. 3. Insert needle-guide of suitable diameter in needle-holder 3 . 4. Slide needle into needle-guide. Check by eye that the needle adjusts easily and correctly in the the three axes, X, Y, and Z.
WARNING
WHEN THE X-RAY TUBE ARM IS BEING UNCOUPLED FROM THE RECEPTOR ARM SUPPORT, DO NOT ALLOW OBSTACLES BETWEEN THE TWO ARMS, AND KEEP YOUR ON THE LEVERS DURING THE WHOLE MOVEMENT.
2-21
G N I S T E A D R O E M P O
STEREOTIX 2
GE Medical Systems REV 1
2-4
om 2155051-2-100
Selecting a Second Point To check that STEREOTIX 2 is operating correctly, the operator can select a second point in the phantom, using the same reference exposure as above: 1. Mark the first point, and leave the film in position on the light plate of the Controller Unit. It is not necessary to reposition the reference marks on the inverted T marks of the film. 2. Hit RESET once to cancel needle and previous point coordinates.
G N I S T E A D R O E M P O
3. Position cross-hair cursors XA, XB, and YAB on each copper disk selected (one per exposure), seen on lefthand and righthand exposures by turning knobs XA, XB, and YAB located on either side of light plate.
WARNING
Note:
THE OPERATOR MUST ACCURATELY SELECT THE SAME POINT SEEN ON THE TWO EXPOSURES. THE XA, XB, AND YAB DEVIATIONS MAY LEAD TO DEVIATIONS OF SEVERAL MILLIMETERS WHEN THE NEEDLE IS BEING INSERTED IN THE PHANTOM OR IN THE LESION.
The Y-ordinate is identical for a given point appearing on the lefthand and righthand exposures. The computer reads the deviation between the original coordinates (marked with X on film) and those of the lesion to be punctured. The deviation values are displayed on the X-Y-Z readout of the Controller Unit. 4. Press L1, L2 or L3 to select needle length. The needle length appears on the corresponding readout. 5. Apply Para 2-3-2: Using the Puncture Device.
2-22
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 3 - EXAM PREPARATION
N O I T M A A R X A E P E R P
3-1
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
Blank page
N O I T M A A R X A E P E R P
3-2
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 3 - EXAM PREPARATION SECTION 1 MAKING A PUNCTURE ON A PATIENT 1-1
Types of Breast Puncture
1-1-1 Cytologic Puncture Cytologic puncture means takinga sampleof thecells from a suspect area in thebreast, whether this zone is palpable or non-palpable in the exposure. The diameter of the needle used for the puncture is usually 0.8 mm thru 1.2 mm. The sample is called the 'material'. A anatomopathology examination is made on the material to determine the type of cells, and to confirm or invalidate its pathological character. Two techniques can be used to obtain a sufficient quantity of cells: a.
Aspiration by a syringe with its needle inserted in the suspect area.
b.
Needle left inserted in the suspect area for a few minutes allowing the cells to move up the needle by capillary action.
1-1-2 Histology Puncture Histology puncture takes a small fragment of tissuefrom thesuspect area in thebreast, whether this area is palpable or not in the clinical examination. This sample is called a 'core sample'. A anatomopathology examination is made on the material to determine thetype of cells, andto confirm of invalidate its pathological character. The diameter of the needle used for the puncture is usually 0.8 mm thru 2.0 mm. The needle is always equipped with a localization wire and has a cutting edge at its distal part or along the length of the needle to cut through the tissue. The core sample is held in the needle by aspiration (syringe).
3-3
N O I T M A A R X A E P E R P
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
1-1-3 Preoperative Localization Preoperative localization entails inserting an x-ray-opaque metal wire which emerges from the skin. The wire allows the surgeon to accurately locate the lesion in the breast, as well as to reduce the size of the tissue sample to be removed and so preserve theesthetic form of thebreast. Localization can be preceded by a cytologic or histologic puncture using a needle of 0.9 thru 1.4 mm in diameter. The needle is equipped with a metal wire whose distal point is in the shape of a hook. (The wire may be shipped in the needle or packaged separately.) When the needle is removed from the breast, the wire catches in the lesion and remains in place. 1-2
Preparation The following model list should be adapted to the requirements and techniques of each user. The decision to apply local anaesthetic before the puncture is made by the radiologist. The following items are arranged on a tray (the Senographe DMR accessory tray may be used):
N O I T M A A R X A E P E R P
D
Sterilized gloves (if required by the radiologist).
D
D
Sterilized compresses. Disinfectant. Local anaesthetic (if required). Sparadrap or equivalent. Local anaesthetic needle (if required). Local anaesthetic syringe (if required).
D
Puncture needle as required (cytology, histology, preoperative localization).
D D D D
Note:
The operator must decide the needle length, measured from the point to the bottom of the guard. D D
Syringe, with socket matching the needle (concerns cytologic or histology puncture when the syringe is not supplied with a needle). Vacuum pumping device of size suited to syringe (concerns cytologic or histology puncture when the gun is not supplied with a needle).
3-4
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100 D
Two needle-guides to suit the needle diameter. THE INTERNAL DIAMETER OF THE NEEDLE-GUIDE MUSTALWAYS BE GREATER THAN THE INTERNAL DIAMETER OF THE NEEDLE, EG:
WARNING
NEEDLE = 0.8 MM INSERTED IN NEEDLE-GUIDE = 0.9 MM D D
Short, flexible adaptor to connect syringe to puncture needleif thesyringe cannot be positioned vertically on the needle. Sample spreading equipment or storage bottle for tissue sample.
N O I T M A A R X A E P E R P
3-5
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 3-1 POSITION OF STEREOTIX 2 ON SENOGRAPHE DMR; OBLIQUE VIEW
N O I T M A A R X A E P E R P
3-6
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
SECTION 2 MAKING A PUNCTURE USING STEREOTIX 2 2-1
POSITIONING
2-1-1 Positioning STEREOTIX 2 on Senographe DMR See illustrations 3-1, 3-2 and 3-3. Depending on the positionof thesuspect x-ray image on the standard mammography exposures, the operator decides the tilt of the Senographe DMR examination arm. For example, if the lesion is distinct and accessible at an oblique angle of 60 _, the operator tilts the examination arm to 60_ so that the patient can be positioned identically for the stereotactic localization exposures. Sometimes, as in preoperative localization, patient positioning depends on the incision to be made by the surgeon. Operator and surgeon need to agree on a suitable positioning for installing the localization wires. 1. Disassemble Senographe DMR accessories. 2. Reduce SID. 3. Place photocell in frontal position (see Chap 2, Para 2-1: STEREOTIX 2 installed on Senographe DMR). 4. Place STEREOTIX 2 on Positioning Arm of Senographe DMR, and close fasteners. 5. Tilt column to selected angular shift using tilt control knobs located above and below of left- and righthand levers of Senographe DMR. Note:
WARNING
The STEREOTIX 2 compression paddles must be in park (High) position. Otherwise, the Up/Down movement and Positioning Arm tilt of the Senographe DMR are inhibited). THE ARM OF THE SENOGRAPHE DMR MUST NOT BE ANGLED MORE THAN +90° OR 90° (i.e. arm horizontal on either side) WHEN STEREOTIX 2 IS USED.
3-7
N O I T M A A R X A E P E R P
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 3-2 POSITION OF STEREOTIX 2 ON SENOGRAPHE DMR; CRANIOCAUDAL VIEW
N O I T M A A R X A E P E R P
3-8
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 3-3 POSITION OF STEREOTIX 2 ON SENOGRAPHE DMR; LATERAL VIEW
N O I T M A A R X A E P E R P
3-9
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 3-4 POSITIONING THE PATIENT
N O I T M A A R X A E P E R P
3-10
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
2-1-2 Positioning the Patient See Illustration 3-4 Mammary puncture with STEREOTIX 2 is always carried out with the patient in the seated position. Use the examination seat of the Senographe DMR. Adjusting the Chair: 1. Set chair height and depth of back of chair so that patient is sitting comfortably with her back straight and firmly against the chair back. 2. Set height of footstand so that patient feet are resting without effort. 3. Set height of arm rests to support patient elbows if required by exam position selected. Note:
The patient should be seated in a comfortable position in order to remain motionless for the duration of the exam. This condition is essential for the accuracy of the puncture. N O I T M A A R X A E P E R P
3-11
STEREOTIX 2
GE Medical Systems REV 1
2-2
om 2155051-2-100
Operating Difficulties It is essential to proceed in a methodical manner for the preparation and running of mammary puncture examinations with the STEREOTIX 2. Observe the following recommendations to ensure the accuracy of the system is maintained.
2-2-1 X-ray Image The suspect image must be of sufficient radiographic quality to be correctly displayed on the localization exposures. This does not concern the size of the image, but its visibility, so that lesions of a size smaller than 0.5 mm can be punctured. The physician must be able to accurately determine the same point to be punctured on each of the two localization exposures, left and right. A deviation of a few millimeters in t he X , Y and Z axes may lead to errors in accuracy when inserting the needle into the breast. 2-2-2 Patient Women undergoing a mammography examination often feel apprehensive because of the examination and the ensuing diagnosis. To reassure the patient, receive her in a warm room and provide a simple but precise explanation of the procedure.
N O I T M A A R X A E P E R P
The patient should be comfortably seated, and well supported by the chair, in order to remain motionless throughout the examination. While the films are being processed, an operator should remain with the patient to ensure that she does not move. The breast should be firmly compressed without, however, causing discomfort to the patient. 2-2-3 Equipment The equipment should be prepared before the start of the examination and placed beside the physician. The accessory tray of the Senographe DMR can be used. The operator must know exactly the length of the needle being used in order to program this length on the STEREOTIX 2. The length is measured from the point to the bottom of the guard. When installing the localization hook, it is recommended to place the point of the needle 5 mm to 10 mm below (0) zero of the Z-coordinate (Z = depth) so that the hook opens below the lesion and surrounds it. This should then to be approved by the physician.
3-12
STEREOTIX 2
GE Medical Systems REV 1
2-3
om 2155051-2-100
Making Stereoscopic Exposures
2-3-1 Positioning the Breast The breast is positioned under the compression paddle so that the area to be examined is opposite the aperture in the paddle (manual and x-ray localization is required). Compress breast by turning compression knobs (compression must be firm). Note:
WARNING Note:
To ensure that the area to be examined is opposite the aperture of the compression paddle, it is recommended to use an exposure with a 0 _ angular shift, an 18 x 24 cm format cassette placed at the center of the cassette-holder of the STEREOTIX 2.
THE AOP (AUTOMATIC OPTIMIZATION PARAMETERS) MODE CANNOT BE USED WITH STEREOTIX 2. When the compression paddle leaves the park (High) position, the Up/Down and rotation movements of the Senographe DMR are inhibited. To move the tube-arm support, press the trigger lever located at its rear. The silk-screened indicators show the position of the tube-arm support when it is tilted to either side of the Positioning Arm. The color code is: Blue: +/-15_ Orange: +/-27_ Four angle shifts are possible: +15_ thru -15_ to make stereoscopic exposures. +27_ thru -27_ to install the needle (exposure is inhibited).
CAUTION
When the tube-arm support is tilted to +/-15_ (color code: Blue), exposure is permitted only with the STEREOTIX 2 installed. In all other cases, exposure is inhibited and a warning message appears on the Control Console of the SENOGRAPHE DMR.
WARNING
WHEN THE X-RAY TUBE ARM IS BEING UNCOUPLED FROM THE RECEPTOR ARM SUPPORT, DO NOT ALLOW OBSTACLES BETWEEN THE TWO ARMS, AND KEEP YOUR HANDS ON THE LEVERS DURING THE WHOLE MOVEMENT.
3-13
N O I T M A A R X A E P E R P
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
2-3-2 Exposure THE AOP (AUTOMATIC OPTIMIZATION PARAMETERS) MODE CANNOT BE USED WITH STEREOTIX 2.
WARNING
Place an 18/24 film cassette in the film-holder plate of the Instrument Unit. Note:
The same film cassette will be used to make the two exposures, left and right. The tube and the film cassette must be positioned on the same side. You can start on either side. Here, we start on the righthand side. D
First Exposure:
1. Insert film cassette in righthand side, for example, of film-holder plate. 2. Tilt tube to +15_ on righthand side (color code: Blue) by pressing lever of the tube-arm support and by pushing the arm gently. When lock is engaged, release lever.
N O I T M A A R X A E P E R P
3. Select rad parameters. 4. Make the exposure. D
Second Exposure:
1. Push film cassette to lefthand side of film-holder plate. 2. Tilt tube to -15_ on lefthand side (color code: Blue) by pressing lever of the tube-arm support and by pushing the arm gently. When lock is engaged, release lever. 3. Select rad parameters. 4. Make an exposure. 5. Remove film cassette and develop film.
WARNING
WHILE THE FILM IS BEING DEVELOPED, THE PATIENT MUST REMAIN STATIONARY AND BE ACCOMPANIED BY ONE OF THE OPERATORS.
2-3-3 Film Description The film contains two exposures, left and right. Each view carries an inverted T reference mark, the letters L or R (left or right), the projection of the radiographic image with suspected lesion, and the projection of the compression paddle aperture.
3-14
STEREOTIX 2
GE Medical Systems REV 1
2-4
om 2155051-2-100
Using the Controller Unit and the Puncture Device
2-4-1 Using the Controller Unit 1. Press key 1 to switch on controller. 2. Press RESET twice to cancel the parameters of the previous exam and to avoid an operating error. The X, Y and Z readouts go off. 3. Slide film along light plate and under cross-hair cursors, and secure it with pressure rollers. 4. Position film with inverted Ts of film superposed on horizontal line of light plate coordinates. 5. Position cross-hair cursors on inverted T of each exposure by turning XA, XB, and YAB knobs (located on either side of light plate). Note:
The Y ordinate is identical for a same point on the lefthand and righthand exposures. 6. Press ZERO SELECT key to validate location of inverted Ts. The X, Y and Z readouts light, displaying 0, 0, 0. 7. Position XA, XB and YAB cross-hair cursors on each of the selected disk projections (one per exposure, lesion to be punctured, by turning the XA, XB, and YAB knobs.
WARNING
Note:
THE USER MUST ACCURATELY SELECT THE SAME POINT SEEN ON THE TWO EXPOSURES. THE XA, XB, AND YAB DEVIATIONS MAY LEAD TO DEVIATIONS OF ONE THRU SEVERAL MILLIMETERS WHEN THE NEEDLE IS BEING INSERTED IN THE LESION.
The Y-ordinate is identical for a given point appearing on the lefthand and righthand exposures. The computer reads the deviation between the original coordinates (marked with an X on film) and those of the lesion to be punctured. The deviation values are displayed on the X-Y-Z readout of the Controller Unit. 8. Press L1, L2 or L3 to select needle length. The needle length appears on the corresponding readout. 9. Release L1, L2 or L3 to confirm and store the length selected even after power is removed from STEREOTIX 2.
3-15
N O I T M A A R X A E P E R P
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
10. Needle length can be adjusted in steps of 1 mm via the + key or - key located on either side of the needle length readout. Holding one of the keys down provides fast incrementation or decrementation. Note:
Needle lengths are preprogrammed at installation: L1 = 70 mm: L2 = 80 mm; L3 = 90 mm. The user can modify this program via the + key or - key located on either side of the needle length readout. Releasing the key validates the length selected and holds it in memory even during STEREOTIX 2 switch-off. The needle length can be set to a depth of 60 mm thru 120 mm via the L1, L2 or L3 key. The readout will not display less than 60 or more than 120.
2-4-2 Using the Puncture Device The deviation readout shows the deviation between the needle-holder and the point selected in the breast (lesion to be punctured). 1. Turn the X, Y, and Z of the needle-holder until the deviation shown by the readout is reduced to zero at X, Y, and Z. The needle-holder is now opposite the point selected in the breast (lesion to be punctured).
N O I T M A A R X A E P E R P
2. Move the tube-arm support to the righthand or lefthand side at + or - 27 _ (color code: Orange) by pressing the lever of the Tube-arm Support and by pushing the arm gently. When the lock is engaged, release the lever. The space above the needle-holder is thereby free andfacilitates the installation of the needle-holder and the needle.
CAUTION
At +27_ and -27_ tilt, exposure is inhibited and a warning message appears on the Senographe DMR Control Console readout.
3-16
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
3. The radiologist disinfects the skin of the patient. 4. Insert needle-guide of suitable diameter in needle-holder. Note:
The tray containing the equipment required for the puncture must be ready and beside the radiologist. The tray provided on the accessory support of the Senographe DRM can be used. 5. Slide needle into needle-guide. When it reaches the surface of the skin, push it firmlyinto thebreast. Thebottomof theneedle-guard should touch theupper edge of the needle-guide.
WARNING
WHATEVER THE TYPE OF PUNCTURE PERFORMED, TEST EXPOSURES MUST BE MADE WITH THE NEEDLE IN PLACE IN THE BREAST. 6. Insert film cassette and make exposures as described in Para 2-3-2, Exposure. 7. The radiologist inspects the exposure; the tip of the needle should be at the center of the suspected lesion. 8. The examination progresses according to the type puncture being made and the technique of the radiologist. 9. Following removal of the needle, and especially in the event of presurgical localization, the radiologist applies a dry sterilized compress over the skin.
WARNING
Note:
WARNING
WHEN SURGICAL LOCALIZATION IS COMPLETE AND AFTER REMOVAL OF THE NEEDLE, THE SURGEON ENSURES THAT: 1. THE COMPRESSION PAD IS RELEASED SLOWLY, AND 2. THAT THE LOCALIZATION WIRE EMERGING FROM THE SKIN IS MOVED CAREFULLY SO AS NOT TO MOVE THE WIRE HOOK FROM THE CENTER OF THE LESION. A biopsy gun hoder is delivered with the systemin order to allow use of such gun to perform a breast biopsy.
WHEN A BIOPSY GUN IS USED, IT IS MANDATORY TO TAKE INTO ACCOUNT THE STRIKE MARGIN (DEPENDENT ON THE TYPE OF NEEDLE AND TYPE OF GUN USED) TO AVOID HITTING THE TRAY.
3-17
N O I T M A A R X A E P E R P
STEREOTIX 2
GE Medical Systems REV 1
2-5
om 2155051-2-100
Selecting a Second Suspect Area (second lesion to be punctured) If two radiographic images with suspected lesions appear on the same exposure, and if the projection of these two images is within the aperture of the compression paddle, a puncture can be made of each suspect area without moving the patient: 1. Use same reference exposure as above, if desired. 2. Mark the first point, and leave the film in position on the light plate of the Controller Unit. 3. Hit RESET ONCE to cancel needle and previous lesion coordinates. 4. Position cross-hair cursors XA, XB, and YAB at center of each of suspect areas selected for second selection (one per exposure), by turning knobs XA, XB, and YAB located on either side of light plate.
WARNING N O I T M A A R X A E P E R P
Note:
THE OPERATOR MUST ACCURATELY SELECT THE SAME POINT SEEN ON THE TWOEXPOSURES. THEXA, XB, ANDYAB DEVIATIONS MAY LEAD TO DEVIATIONS OF SEVERAL MILLIMETERS WHEN THE NEEDLE IS BEING INSERTED IN THE LESION.
The Y-ordinate is identical for a given point appearing on the lefthand and righthand exposures. The computer reads the deviation between the original coordinates (marked with an X on the film) and those of the lesion to be punctured. The deviation values are displayed on the X-Y-Z readout of the Controller Unit. 5. Press L1, L2 or L3 to select needle length. The needle length appears on the corresponding readout. 6. Apply Para 2-4-2, Using the Puncture Device.
3-18
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 4 - MAINTENANCE
E C N A N E T N I A M
4-1
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
Blank page
E C N A N E T N I A M
4-2
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 4 - MAINTENANCE SECTION 1 CLEANING AND DISINFECTION Adequate cleaning and disinfection is necessary to prevent disease transmission between patients. Be sure to thoroughly clean and disinfect equipment surfaces that contact the patient and all equipment surfaces likely to become soiled from patient contact. Mammography equipment surfaces (Table 4-1) that normally contact the patient's intact skin need only low or intermediate level disinfection. However, when dealing with patients having exposed wounds, lesions or other conditions where breached skin may contact the equipment surface, a high-level disinfection may be warranted. TABLE 4-1 CLEANING
Description Bucky plates Compression paddles Magnification plate tummy shield face shield other plastic or painted surfaces
Operation Use mild soap or a disinfectant Use mild soap or a disinfectant Use mild soap or a disinfectant Use mild soap or a disinfectant Use mild soap or a disinfectant Use mild soap
Improper cleaning methods or the use of certain cleaning and disinfecting agents can damage the equipment, cause poor imaging performance or increase the risk of electric shock. To avoid possible injury or equipment damage:
CAUTION D
D D
Do not use harsh detergents, abrasive cleaners, strong concentrations of alcohol or any products containing methyl alcohol. If skin prep products contain high alcohol concentrations, allow sufficient drying time before applying compres sion. Do not expose equipment parts to steam or high temperature sterilization. Never allow liquids to enter the internal parts of the equipment. If you become aware of liquid entry, disconnect the electrical supply and have the equipment checked by qualified service personnel before returning it to use.
Equipment Cleaning Instructions Equipment parts may be washed with mild soap in lukewarm water. Removable parts that do not contain electrical components such as the compression paddles may be removed from the equipment and immersed. Equipment surfaces that enclose electrical components maybe wiped with soft dampened cloth, taking care notto allow liquids to enter the equipment. Surfaces should be scrubbed as needed using a soft sponge, gauze or cloth to remove all visible residue.
4-3
E C N A N E T N I A M
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
Scrubbing with a soft bristle brush (such as a toothbrush) may be necessary to reach corners or to remove material that has dried onto the surface. Rinse all surfaces with clean water to remove visible soap residue, taking care to prevent liquid entry to internal equipment parts. Dry surfaces with a soft cloth to remove any visible residue. Low or Intermediate Level Disinfection Before each use, equipment surfaces that contact the patient (Bucky, compression paddle, cassette holder or magnification device) should be subjected to low or intermediate level disinfection with a suitable disinfecting or sanitizing agent. Use an EPA registered disinfectant and be sure to follow the label instructions and precautions for use, storage and disposal. High Level Disinfection In the event of equipment contacting breached skin or being used with infected or immune compromised patients, equipment surfaces that contact the patient should be subjected to high level disinfection. Use only FDA cleared and EPA registered high-level disinfecting agents and be sure to follow the label instructions and precautions for use, storage and disposal. Cidex liquid chemical germicide is recommended for this purpose.
E C N A N E T N I A M
Recommended Cleaning Agents The following products have been used on GE equipment without causing equipment damage. This list is not exclusive nor intended to certify these product in any way. GE does not warrant any of these products, please contact the manufacturer of the cleaning agent directly with any questions you may have. T-Spray disinfectant detergent, mfg. by Pharmaceutical Innovations, Inc., Newark, NJ, USA EPA Reg. No. 1839-83-65071, EPA Est. No. 65071-NJ-01 Do Not Use any of the following chemicals at or above the concentrations given: Methanol - all concentrations, D Ethanol - 50% or greater, D Sodium hypochlorite (bleach) - 5% or greater, D abrasive cleansers, D organic solvents, D Other chemicals that may react with Lexan (tm), epoxy resins or paint. If you are unsure, contact the manufacturer of the cleaning agent directly. D
4-4
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
SECTION 2
PREVENTIVE MAINTENANCE STEREOTIX 2 operational reliability ensures accurate measurement over a long period. However, it is recommended to check overall calibration regularly using the Plexiglas phantom supplied with the system. Checking frequency depends on the frequency of use, but should be at least monthly. A GEMS-E Service Engineer will carry out the annual planned maintenance during one of the two annual site calls scheduled for the SENOGRAPHE DMR. SECTION 3
MESSAGES Three types of messages may be displayed for the user on the gantry or on the Control Console: D
I xx:
D
E yy: S zz:
D
Note:
Information message: fault message which does not inhibit the exposure. Error message: fault inhibits the exposure. Stop message: exposure is stopped due to incorrect use of equipment.
A 15-character message is displayed on the Readout located on top of the Column. The first three characters are identification characters. The remaining 12 characters constitute the message. Example:
I 76 FAULT
A message displayed on the Control Console has 40 characters. The first three characters are identification characters. The next 28 characters comprise the message. The seven remaining characters are also identification characters. Example:
I 76 DETECTOR FAILURE
158/011
When calling the Field Service of GE CGR, it is essential to note each message in its entirety (2 for the Readout messages, and 3 for the Control Console messages). Listed below by type are all the messages which can appear on the Column Readout and on the Control Console, with the corresponding corrective action. For a description of the messages for the SENOGRAPHE DMR and STEREOTIX 2, see the SENOGRAPHE DMR om, Chapter 4, Section 3.
4-5
E C N A N E T N I A M
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
TABLEAU 4-2 ERROR MESSAGESDISPLAYED ON READOUT AND ON CONTROL CONSOLE
Message No.
E C N A N E T N I A M
Message on Column Readout
Message on Control Console
Explanation/meaning of message
Corrective Action
I 50
LOCK JAMMED
None
LOCK PROBLEM: In spite of command, lock release does not operate.
1.Try again to release lock. 2.If fault continues, note ALL messages and contact GEMS-E Field Service
I 51
HOT LOCK
None
Lock temperature too high (used for too long period and/or toofrequently).
-Wait for lock to cool.
I 52
ARM UNLOCKED
TUBE HOUSING ARM UNLOCKED
Tube housing arm in unsta ble position because lock is badlylatched.
1.Pivot arm to latch correctly. 2.If fault continues, note ALL messages and contact GEMS-E Field Service.
I 53
UNLOCKED SID
UNLOCKED IMAGE RE CEPTOR
Image receptor is incorrectly locked.
1.Check that image receptor is correctly locked in a given SID position. 2.If fault continues, note ALL messages and contact GEMS-E Field Service.
I 74
LOCK FAILURE
ARM LOCK FAILURE
Arm lock is not functioning correctly.
1.Switch OFF, and switch ON. 2.If fault continues, note ALL messages and contact GEMS-E Field Service.
I 80
STEREOTIX
None
Movements requested (ele vator or rotation) cannot be made by interlock with Ster eotix 2option.
1.Connectcontrollerof Stereo tix 2 to Senographe. 2.Connectpositionertocontroller. 3.Switchcontrolleron. 4.Move Stereotix 2 paddle up to end of travel. 5.If fault continues, note ALL messages and contact GEMS-E Field Service.
4-6
STEREOTIX 2
GE Medical Systems REV 1 Message No.
om 2155051-2-100 Message on Column Readout
Message on Control Console
Explanation/meaning of message
Corrective Action
E 51
COLLIM FAIL
LATERAL COLLIMATOR FAILURE
Lateral collimator is faulty, so 1.Switch OFF. a good format cannot be en 2.Switch ON. sured. 3.If fault continues, note ALL messages and inform Exposure is inhibited. GEMS-E Field Service.
E 53
COLLIM FAIL
REAR COLLIMATOR FAILURE
Rear collimator is faulty, so a good format cannot be en sured. Exposure is inhibited.
1.Switch OFF. 2.Switch ON. 3.If fault continues, note ALL messages and inform GEMS-E Field Service.
S 50
None
INCORRECT X-RAY IN CIDENCE
Position of Tube-arm Support does not permit exposure: -Positionnotlocked -Outside ±15° range (Stereotix 2 not installed) -Outside ±27° range (Stereotix 2 installed)
-Correct the position of the tube carrying arm.
S 51
None
INCORRECT SID
Image receptor is at SID 510 mm and Stereotix 2 notinstalled.
-Change the focal distance.
S 52
None
DET. CELL DISTANCE NOT OK
Stereotix 2 is installed and the image receptor is not at SID 510 mm.
-Change SID to 510 mm.
4-7
E C N A N E T N I A M
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
Blank page
E C N A N E T N I A M
4-8
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 5 - SPECIFICATIONS
S N O I T A C I F I C E P S
5-1
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
ILLUSTRATION 5-1 INSTRUMENTUNIT & CONTROLLER UNIT: SPACEREQUIREMENTS
Z Y X
300
435
425
370 S N O I T A C I F I C E P S
320
210 130
480
5-2
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
SECTION 1
PHYSICAL SPECIFICATION See Illustration 5-1 TABLEAU 5-1 SPACE REQUIREMENTS & WEIGHT
Dimensions
Description Length mm (ft)
Weight
Width mm (ft)
Height mm (ft)
Illustration
daN (lb)
ControllerUnit withoutReadout
480
(1.57)
320
(1.05)
130
(0.43)
8.2
(18.1)
1-4
ControllerUnit with Readout
480
(1.57)
370
(1.22)
210
(0.69)
8.2
(18.1)
1-4 1-5
Instrument Unit
435
(1.43)
425
(1.39)
300
(0.98)
8.5
(18.7)
1-3
SECTION 2
ELECTRICAL & TECHNICAL SPECIFICATION 2-1
Technical Specification NEEDLE-HOLDER TRAVEL: D D D
50 mm ± 0.5 mm, on X axis, 40 mm ± 0.5 mm, on Y axis, 70 mm ± 0.5 mm, on Z axis.
ACCURACY: D
Needle: ± 1 mm, in X, Y and Z axes.
COMPRESSION PADDLE APERTURE: D D
54 mm in X axis, 43 mm in Y axis.
TYPE OF MATERIALS IN CONTACT WITH SKIN: D
D
Compression paddle: - Lexan polycarbonate, Film cassette holder: - Impax.
5-3
S N O I T A C I F I C E P S
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
A
ILLUSTRATION 5-2 PRECISIONTEST PHANTOM
É
É
30 25
50 40
15
4 R=1
30
5
4 R=3
13,5
5 2
0 3
4 5
5 1 5
5 3
5
43
Edge (gravure)
10 S N O I T A C I F I C E P S
7
57 20
Five holes 0.9 thru 1.3 mm dia.
A 5 (PM)
0.1
Five copper plugs/hole + 0. 1 mm
SECTION A-A
5-4
3 2
0 0 1
STEREOTIX 2
GE Medical Systems REV 1
2-2
om 2155051-2-100
Radiographic Specifications D
Film cassette to DIN standard: - 18 x 24 mm.
D
Film: - 18 x 24 mm.
D A and B exposure formats:
- 85 x 75 mm. D
Measurement surface: - 54 x 43 mm.
D
Resolution: - ± 1 mm.
D
Precision test phantom (see Illustration 5-2) - 50 x 100 x 57 mm (n x L x W). Five holes drilled for accuracy test: - Ø 0.9 à 0.13 mm, - Depths: 5, 10, 20, 30 and 40 mm.
2-3
Electrical Specification Power supply: 27 VAC, supplied by SENOGRAPHE DMR gantry.
SECTION 3
AMBIENT CONDITIONS TABLEAU 5-2 AMBIENT LIMIT CONDITIONS
RELATIVE HUMIDITY withoutcondensation
80%
TEMPERATURE In operation
Storage and transportation
Min.
Max.
Min.
Max.
15° C (59° F)
40° C (104° F)
-30° C (-22° F)
60° C (140° F)
5-5
ALTITUDE Storage and In operation transporta tion 3000 m (9842')
3000 m (9842')
S N O I T A C I F I C E P S
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
SECTION 4 MEANING OF SYMBOLS
Protective earth (ground)
Earth (ground)
Dangerous voltage
Type B equipment
This symbol indicates that waste electrical and electronic equipment must not be disposed of as unsorted municipal waste and must be collected separately. Please contact an authorized representative of the manufacturer for information concerning the decommissioning of your equipment.
S N O I T A C I F I C E P S
5-6
STEREOTIX 2
GE Medical Systems REV 1
om 2155051-2-100
CHAPTER 6 - REGULATIONS The STEREOTIX 2 complies to the following regulations: D
UL 187 X-ray equipment
D
CSA 22.2, No. 114
Radiation Emitting Equipment
D
IEC 601-1
Medical Electrical Equipment Part 1: General requirements for safety
D
IEC 601-2-7
Medical Electrical Equipment Part2: Particular requirements for the safety of high voltage generators of diagnostic x-ray generators.
Code of Federal regulation Performance standards for ionizing radiation emitting products. D EMC (Electromagnetic compatibility): D Emission: 1. CISPR11 class A: Electromagnetic disturbances. D
21 CFR, Part 1020.30, sub. J
Immunity 1. IEC 801-2
D
Electrostatic discharge (conducting parts:3kV, non-conducting parts:8kV).
2. IEC 801-3
Radiated electromagnetic fields (3 V/m, from 26MHz to1GHz).
3. IEC 801-4
Elecrical fast transient/burst (2kV on power supply cables, 500 V interconnection cables).
4. IEC 801-5
Shock waves: (2kV common mode, 1kV differential mode).
D
CE Marking according to council directive 93/42/EEC.
S N O I T A L U G E R
6-1