Contents ABSTRACT ...................................................................................................................................................... 2 INTRODUCTION ............................................................................................................................................. 3 OBJECTIVES ................................................................................................................................................... 4 THEORY ......................................................................................................................................................... 5 MATERIAL AND APPARATUS ....................................................................................................................... 12 PROCEDURES ............................................................................................................................................... 15 RESULT AND CALCULATIONS ...................................................................................................................... 18 DISCUSSION ................................................................................................................................................ 25
CONCLUSION ............................................................................................................................................... 28 RECOMMENDATION ................................................................................................................................... 29 REFFERENCE ................................................................................................................................................ 30 APPENDIX .................................................................................................................................................... 31
ABSTRACT
SOLTEQ® Flowmeter Measurement Apparatus (Model: (Model: FM101) the combination combination of a few flowmeter is designed to amount a flow of an incompressible fluid such as water. By doing this experiment, the flow rates of the the water we will will measure the flow rates rates by comparing the pressure different in three types of flow meter which are orifice, rotameter, and venturi meter. Meanwhile, the loss coefficient can be determined when fluid flows through a 90◦ elbow by
measuring flow rate on manometers. The actual flowrates for the water measure by setup fixed f ixed volume divided by time. Then Then by plotting the graph graph we can compare the flow rates rates each meter and do further discussion. By the end, venturi meter is more accurate compare to orifice meter which the flow rates of venturi meter is closer to the actual value of the flow rates. On the whole, our experiment was successfully done because we has achieved the objective of the experiment.
INTRODUCTION
The apparatus used in this experiment, SOLTEQ flowmeter measurement (Model : FM101) apparatus is premeditated together with a hydraulic bench and a water supply. This allow student to gain more knowledge for themselves with classic methods of flow measurement an example of incompressible incompressible fluid. By operating this equipment student will be allowed to compare between different types of flow measurement devices, specifically by using a venturi device, orifice device and rotameter. Later, the data recorded will be used to study against measurement obtained from the hydraulic bench. Other feature of the apparatus includes a 90 degree elbow with pressure tappings. This feature allows students to calculate the total head loss and loss coefficient when fluid flows through the 90 degrees elbow. To conclude, the apparatus allows the following experiment to be peform, a comparison of flow measurement using venturi, orifice, rotameter and bench and determination of total head loss and loss coefficient of fluid flow through a 90 degree elbow.
OBJECTIVES
For this experiment, the objectives are given below : 1. To observe and compare the flow of the flow measurement between venture, orifice, rotameter and bench. 2. To determine the total head head loss and loss coefficient coefficient of fluid flow through through a 90 degree elbow. 3. To determine the total head loss of fluid flow. 4. To verify Bernoulli’s equation.
THEORY
Venturi Meter
The venturi meter consists of a venturi tube and a suitable differential pressure gauge. The venturi tube has a converging portion, a throat and a diverging portion as shown in the figure below. The function of the converging portion is to increase the velocity of the fluid and lower its static pressure. A pressure difference difference between inlet and and throat is thus developed, where pressure difference is correlated with the rate of discharge. The diverging cone serves to change the area of the stream back to the entrance area and convert velocity head into pressure head.
Figure 1: Venturi Meter
Assume incompressible flow and no frictional losses, from Bernoulli’s Equation
p1
v 1
2
2g
Z 1
p 2
v 2
2
2g
Z 2 ………………….…………………………...(1)
Use of the continuity continuity Equation Q = A1V 1 = A2V 2, equation (1) becomes p p 1
2
2 V 2
A 1 Z Z 2g A 1
2
2
1
2
………………….…………………...(2)
A Ideal Q A V A 1 A
2
2
2
2
1
2
1 / 2
p p Z Z 2g 1
2
1
2
1/ 2
…………...…(3)
However, in the case of real fluid flow, the flow rate will be expected to be less than that given by equation (2) because of frictional effects and consequent head loss between inlet and throat. In metering practice, this non-ideality is accounted by insertion of an experimentally determined coefficient, Cd,termed as the coefficient of discharge. With Z 1 = Z2 , equation (3) becomes
A Actual Q C d A 1 A 2
2
1
2
1 2
p p 2 g 1
1 2
2
……………..…. (4)
Hence,
At 2 q C d At 1 A
1 2
2 g P P /
12
1
2
…………………....…. (5)
Where, = = = = = = = = =
C d D2 D1 At A g
P1 P2
Coefficient of discharge (0.98) Throat diameter = 16 mm Inlet diameter = 26 mm Throat area = 2.011 x 10-4 m2 Inlet area = 5.309 x 10-4 m2 9.81 m/s2 Density of water = 1000 kg/m3 Inlet pressure (Pa) Throat pressure (Pa)
Rotameter
The rotameter is a flow meter in which a rotating free float is the indicating element. Basically, a rotameter consists of a transparent tapered vertical tube through which fluid flow upward. Within the tube is placed a freely suspended “float” of pump-bob shape. When there is
no flow, the float rests on a stop at the bottom end. As flow commences, the float rises until
upward and buoyancy forces on it are balanced by its weight. The float rises only a short distance if the rate of flow is small, and vice versa. The points of equilibrium can be noted as a function of flow rate. With a well-calibrated marked glass tube, the level of the float becomes a direct measure of flow rate.
Scale Tapered tube
Flow
Figure 1: The Rotameter
Orifice Meter
The orifice for use as a metering device in a pipeline consists of a concentric squareedged circular hole in a thin plate, which is clamped between the flanges of the pipe as shown in the figure below.
A2
A1
Figure 3: Orifice Meter
Pressure connections for attaching separate pressure gauges are made at holes in the pipe walls on both side of the orifice plate. The downstream pressure tap is placed at the minimum pressure position, which is assumed to be at the vena contracta. The centre of the inlet
pressure tap is located between one-half and two pipe diameters from the upstream side of the orifice plate, usually a distance of one pipe diameter is employed. Equation (4) for the venturi meter can also be applied to the orifice meter where
A Q C A 1 Actual d A 2
2
1
2
1 2
1 2
p p 2 g 1
2
………………. (6) ……………….
The coefficient of discharge, Cd in the case of the orifice meter will be different from that for the case of a venturi meter.
A Q C d At 1 t A
2
1 2
2 g h h
1 2
7
8
…………………………….(7)
Where, C d D7 D8 At A (h7 – h8)
= = = = = =
Coefficient of discharge (0.63) Orifice diameter = 16 mm Orifice upstream diameter = 26 mm Orifice area = 2.011 x 10-4 m2 Orifice upstream area = 5.309 x 10-4 m2 Pressure difference across orifice (m)
90o elbow
Figure below shows fluid flowing in a pipeline where there is some pipe fitting such as bend or valve, and change in pipe diameter. Included in the figure is the variation of piezometric head along the pipe run, as would be shown by numerous pressure tappings at the pipe wall.
Figure 2 : Piezometric head along a pipeline
If the upstream and downstream lines of linear friction gradient are extrapolated to the plane of fitting, a loss of piezometric piezometric head,
h, due to the fitting is found. found. By By introducing introducing the the
velocity heads in the upstream and downstream runs of pipe, total head loss, H can be determined in which
H h
V 1
2
2g
V 2
2
2g
………………………………………………………………(8)
Energy losses are proportional to the velocity head of the fluid as it flows around an elbow, through an enlargement or contraction of the flow section, or through a valve. Experimental values for energy losses are usually expressed in terms of a dimensionless loss coefficient K , where
K
H 2
V 1 / 2g
or
H 2
V 2 / 2g
……………………………..…………………………………(9)
For results of better accuracy, long sections of straight pipe are required to establish with certainty the relative positions of the linear sections of the piezometric lines. However, in a compact apparatus as described in this manual, only two piezometers are used, one placed upstream and the other downstream of the fitting, at sufficient distances as to avoid severe disturbances. These piezometers measure the piezometric head loss,
h’ between the tapping.
Thus h h'h f ……………………………..………………………………………(10)
L V Where h f 4 f D g 2 2
Δh f = f L D V
= = = =
friction head loss which would be incurred in fully developed flow along the run of pipe between the piezometer tappings friction factor distance between the piezometer, measured along the pipe center line pipe diameter average velocity of fluid flow in pipe
The friction head loss is estimated by choosing a suitable value of friction factor, f for fully developed flow along a smooth pipe. The method used in this manual to determine the friction factor is the Prandtl equation 1
f
4 log Re
f
0.4
…………………………………………………………(11)
Typical values derived from this equation are tabulated in the table below:
x 104
0.5
1.0
1.5
2.0
2.5
3.0
3.5
x 10-3
9.27
7.73
6.96
6.48
6.14
5.88
5.67
Re, f ,
In determination of the fraction factor, f , it is sufficient to establish the value of f at at just one typical flow rate, as about the middle of the range of measurement due to the fact that f
varies only slowly with Re, and the friction loss is generally fairly small in relation to the measured value of h’. Characteristic Characteristic of flow f low through elbow and at changes in diameter
90 o Elbow
Figure below shows flow round a 90o elbow which has a constant circular cross section. D
V
R
Figure 3 : 3 : 90o Elbow
The value of loss coefficient K is is dependent on the ratio of the bend radius, R to the pipe inside diameter D. As this ratio increase, the value of K will will fall and vice versa. H KV 2 / 2 g …………………………………………………..……………(12)
Where, K V g
= = =
Coefficient of losses Velocity of flow 9.81 m/s2
MATERIAL AND APPARATUS
Part Identification Identification
5 1 6
2
3 7 8 4
9
Figure 4: Part Identification Diagram
1. Manometer Tubes
6. Rotameter
2. Discharge Valve
7. 90° Elbow
3. Water Outlet
8. Orifice
4. Water Supply
9. Venturi
5. Staddle Valve
Sketch of apparatus and devices
Discharge valve Rotameter
90 elbow
Venturi Meter
Orifice
Water Supply
Figure 5: 5: Sketch of apparatus and devices
Specification of dimensions
Venturi meter A
B
C
D
E
F
Figure 6: Specification Specification of the Venturi Meter
Tapping A = 26 mm Tapping B = 21.6 mm Tapping C = 16 mm Tapping D = 20 mm Tapping E = 22 mm Tapping F = 26 mm
Orifice
G
H
Figure 7: Specification Specification of the Orifice Plate
Orifice upstream diameter (G) = 26 mm Orifice diameter (H) = 16 mm
PROCEDURES
General Start-up Procedures
1.
The flow control valve of hydraulic bench is fully closed and the discharge valve are fully open.
2.
The discharge hose is properly directed to volumetric tank of fibreglass before starting up system. The volumetric drain valve also had been ensure left open to allow flow discharge back into sump tank.
3.
Once step (2) is confirmed start up the pump supply from hydraulic bench. the bench valve are slowly open. At this point, the observation show the water flowing from the hydraulic bench through to the flow apparatus and discharge through into the volumetric tank of hydraulic bench and then drained back into sump tank of hydraulic bench.
4.
The flow control valve then open fully. When the flow in the pipe is steady and there is no trapped bubble, the bench valve were closed to reduce the flow to the maximum measurable flow rate.
5.
The water level in the manometer board will begin to display different level of water heights. If the water level in the manometer board is too high where it is out of visible point, adjust the water level by using the straddle valve. With the maximum measurable flow rate, retain maximum readings on manometer.
6.
The flow discharge valve were slowly reduce to reduce the flow until it fully closed
7.
Water level in the manometer board begin to level into a straight level. This level maybe at the lower or maybe at the higher end of the manometer board range.
8.
“Trapped Bubbles” in the glass tube or plastic transfer tube been lookout. The
bubble removed from the system for better accuracy.
Demonstration of the operation and characteristic of three different basic types of flowmeter.
Procedures:
1.
The apparatus placed on bench, the inlet pipe connected to bench supply and outlet pipe into volumetric tank.
2.
The bench valve fully closed and the discharge valve fully opened, the pumped switched on from the hydraulic bench.
3.
The bench valve closed when the flow in pipe is steady and there is no trapped bubble to reduce the fflow low to the maximum measurable flow rate.
4.
The water level in the manometer board adjusted by using the air bleed screw. Maximum readings retained on manometers with the maximum measurable flow rate.
5.
The readings on manometers (A - J), rotameter and measured flow rate recorded.
6.
Step 5 is repeated for different flow rates. The flow rates can be adjusted by utilizing both bench valve and discharge valve.
7.
To demonstrate similar flow rates at different system static pressures, adjust bench and flow control valve together. Adjusting manometer levels as require
Determination of the loss coefficient when fluid flows through a 90 d egree elbow
Procedures:
1.
The apparatus placed on the bench, inlet pipe connected to bench supply and outlet pipe into volumetric tank.
2.
The bench valve fully closed and the discharge valve fully opened, The pump supply from hydraulic bench switch on.
3.
The bench valve were slowly opened until it is fully opened.
4.
When the flow in the pipe is steady and there is no trapped bubble, the bench valve closed to reduce the flow to the maximum measurable flow rate.
5.
The manometer board adjusted by using the air bleed screw. Retain maximum readings on manometers with the maximum measurable flow rate.
6.
The readings on manometers manometers (I and J) and measured flow rate were recorded.
7.
Step 6 is repeated for different flow rates. The flow rates adjusted by utilizing both bench valve and discharge valve.
8. 9.
The tables data were complete The graph H against
V s
2
2g
for 90 degree elbow to determine the coefficient of
losses were plotted.
General Shut-down Procedures
1.
The water supply valve and venturi discharge valve closed.
2.
water supply pump was turned off.
3.
The water from the unit drained off when not in use.
RESULT AND CALCULATIONS
NO. 1 2 3 4 5
A 281 284 302 328 360
B 280 282 298 319 347
MANOMETER READING ( mm ) C D E F G 277 278 279 280 280 265 277 276 277 277 262 283 289 294 293 256 297 308 315 313 250 313 329 344 342
NO.
ROTAMETER ( L/MIN )
VOLUME (L)
TIME ( MIN )
FLOWRATE,Q ( L/MIN )
1
4
2
0.58
3.45
2
8
2
0.27
7.41
3
12
2
0.17
11.76
4
16
2
0.13
15.38
5
20
2
0.10
20.00
NO 1 2 3 4 5
NO 1 2 3 4 5
Time Flow Rate, Q t ( m3/s ) 5.75 10-5 1.23 10-4 1.96 10-4 2.56 10-4 3.33 10-4
×× ×× ×
Variable Area Head Loss ,Ha ( m ) -0.001 0.001 -0.006 -0.011 -0.016
Rotameter Flow Rate, Q a ( m3/s ) 6.67 10-5 1.33 10-4 2.00 10-4 2.67 10-4 3.33 10-4
×× ×× ×
Orifice Plate Head Loss ,Ho ( m ) 0.008 0.036 0.089 0.157 0.252
H 272 241 205 158 92
I 275 258 240 220 190
J 275 255 238 216 183
FLOWRATE USING BERNOULLI EQUATION ORIFICE Venturi 3.83 10-5 5.42 10-5 ( m3/s ) ( m3/s ) 8.35 10-5 1.15 10-4 ( m3/s ) ( m3/s ) 1.21 10-4 1.81 10-4 ( m3/s ) ( m3/s ) 1.62 10-4 2.40 10-4 ( m3/s ) ( m3/s ) 2.01 10-4 3.04 10-4 ( m3/s ) ( m3/s )
× × × × ×
× × × × ×
Timed Flow Rate Squared,Q t2 3.30 10-9 1.51 10-8 3.84 10-8 6.55 10-8 1.11 10-7
×× ×× ×
Venturi Meter Head Loss ,Hv ( m ) 0.004 0.019 0.040 0.072 0.110
Q,flowrate (L/min) orifice venturi 2.30 3.25 5.01 6.90 7.26 10.86 9.72 14.40 12.10 18.24
actual 3.45 7.41 11.78 15.38 20.00
rotameter 4.00 7.98 12.00 16.02 19.98
Comparism of flowrates between flowmeter flowmeter l 25 a u t c a 20 Q , e r t 15 w m w o 10 l f d e t a 5 l u c l a c 0
actual orifice venturi rotameter
0
5
10
15
20
25
Q rotameter (L/min)
The graph shows shows the comparism of the flow rates between the flowmeters, Based Based on the experiment from trial no 1 untill 5, the graph that indicates the higher the flow rates of the rotameter, the higher the flowrates of the venture and orifice. However ,since the coefficient of discharged for orifice meter is smaller than venture meter, the graph shows that venture meters shows that its flow rates calculated using Bernoulli’s equation is nearer to the actual value of flow rates.
Determination of loss coefficient when fluid flows through a 90° elbow NO
VOLUME
TIME
FLOWRATE.Q
DIFFERENTIAL
V
V 2/2g
(L)
( sec )
( L/min )
10 10 10 10 10
175.00 79.70 52.25 38.73 29.13
3.43 7.53 11.48 15.49 20.60
1 2 3 4 5
∆ℎ′
PIEZOMETER PIEZOMETER HEAD, ( ( mm ) ELBOW ( h i-h j ) 1 3 2 4 7
( mm/s )
( mm )
5.7143*10-3 3.7642*10-2 3.8278*10-2 1.0328*10-1 2.4030*10-1
1.6642*10-9 7.2218*10-8 7.4679*10-8 5.4367*10-7 2.9431*10-6
Graph of delta h against v 2/2g 10
) m m ( d a e H r e t e m o z e i P l a i t n e r e f f i D
9
y = 1.3x 1.3x - 0.5 R² = 0.7972
8 7 6 5 4 3 2 1 0 0
1
2
3
4
5
6
7
8
v2/2g x 10-6 (mm)
The graph shows that the differential pizometer head is increasing as the v 2/2g increasing.
∆ℎ′
Therefore,
is increasing linearly with the v 2/2g. From the graph, we obtained can obtained is
the coefficient of losses which is the slope of the graph. NO
Variable Area % Flow Rate Error
Orifice Plate % Flow Rate Error
Venturi Meter % Flow Rate Error
1 2 3 4 5 AVERAGE
16.00 8.13 2.04 4.30 0 6.09
-5.74 -6.50 -7.65 -6.25 -9.6 -7.15
-33.39 -32.11 -38.27 -36.72 -39.64 -36.03
CALCULATIONS
Experiment 1
Timed Flow rate, Q t ( m3/s )
=
=
= (0.002m3/34.8s ) = 5.75
×
10-5 m3/s
Variable Area Flow Rate, Q a ( m3/s )
=
= =
6 46 ×
= 6.67
10-5 m3/s
Orifice Plate Flow Rate, Q o ( m3/s )
= Cd A2
√ 2−−
= ( 0.63 ) ( 2.01
= 5.42
×
× √ −(−2(9...8××.8) 10-4 )
10-5( m3/s )
Venturi Meter Flow rate, Q v ( m3/s )
= Cd A2
√ 2−− × √ −(−2(9...8××.4)
= ( 0.98 ) ( 2.01
= 3.83
×
10-4 )
10-5( m3/s )
Velocity of the water , v(mm/s)
, ∆
=
V = 1 mm/ 175 s
V = 5.7143 x 10 -3mm/s Sample calculation for v 2/2g = 5.7143 x 10-3mm/s / 2(9810mm/s2)
= 1.6642*10-9 mm
Sample calculation variable Area % Flow Rate Error
=
=
− × 6.6 × . ×−. × × 100
100
= 16%
Orifice Plate % Flow Rate Error
=
=
− × .42 × . ×−. × × 100
100
= -5.74%
Venturi Meter % Flow Rate Error
=
=
− × 3.83× .−× . × × 100
100
= -33.39%
Variable Area Head Loss ( H a )
= hD – hE = 0.277m – 0.278m = - 0.001m
Orifice Plate Head Loss ( H o )
= hF – hH
= 0.280m – 0.272m = 0.008m
Venturi Meter Head Loss ( H v )
= hA – hC = 0.281m – 0.277m = 0.004m
Timed Flow Rate Squared ( Q t2 )
= ( Timed Flow Rate )2 = =
5.75 × 10−− . ×
2
DISCUSSION
Based on the experiment, we successfully obtained the flow rate measurement by comparing of pressure drop by using three basic types of flow measuring measuring techniques which which is Orificemeter, Rotameter and Orifice Meter. Flow rates from the rotameter used as the factor to gain to flow rates for venturi meter and the orifice meter. Besides, we also determined the actual flow rates for the water using the constant volume and the time taken for the water to reach the specific volume experiment.
Comparism of flowrates between flowmeter flowmeter 25
l a u t c a 20 Q , e r t 15 w m w o l f 10 d e t a 5 l u c l a c
actual orifice venturi rotameter
0 0
5
10
15
20
25
Q rotameter (L/min)
Referring to the graph, the data from all the flow meter show the same trend which increase lineary. As the flow rates increases, all reading from the all types of the flow meter increase. Though use rotameter as the reference, in case the actual value for flow rates that we got turns from the rotameter flow rates. Besides, the flowrates for venturi meter and orifice also turn from the actual flow rates. Among the factor could be due to the friction and the no slip condition as water flows through each of the flowmeters. Q rotameter > Q actual > Q venturi> Q orifice
The graph shows that when comparing between orifice and venturi, venturi meter is more accurate since the flow rate obtained from the orifice meter is closer to the actual value of the flow rate. Due to its streamlined design, its gradual contraction and expansion prevent flow separation and whirling, whirling, and minor minor friction losses on the inner inner wall surfaces. The The meter consist on streamlined shape and the system has a steady change in diameter. The flow streamline does not have drastically change in diameter like orifice and do not obstructed by a float like rotameter. Thus pressure disturbed are less likely. While, for orifice, it has the simplest design and it occupies minimal space. The sudden change in the flow causing causing the swirl and the velocity velocity increase, the vena contracta contracta decreases. The smaller the vena contracta gets the greater the pressure difference, thus higher energy and pressure loss. In overall orifice meter have highest minor loss coefficient, while venture meter has the lowest. For rotameter, it should be used if large pressure drop is acceptable, when comparing between the three flow meter, the energy loss for rotameter are the higher than venturi and orifice. If large pressure change, the most suitable are venture meter cause it will damage the pipes. However its large and expensive to produce. To install it in resident pipeline it requirebig space. Vice versa, orifice meter is very inexpensive for it is just a flat plate and a thin orifice plate and because of that easy to install. The discharge coefficient are the minimum required flow calculated based on the design flow rate and expected pressure drop across the valve, and valve must have a flow coefficient higher than calculated value. The value of constant are different because discharge coefficient valve for liquid are smaller than gas due the expansion of the gas. For 90° elbow, the observation we made from fluid flowing, the element disturb the smooth flow of the fluid and cause additional loss of pressure(head loss) because of the flow mixing and separation the element induced. Thus the experiment is carried out to resolute minor losses which usually represent by discharge coefficient and resistant coefficient.
After successfully doing the investigation for the coefficient discharge, the relationship
can be show based on the graph Δ ′ versus v2/2g .
Graph of delta h against v 2/2g 10
) m m ( d a e H r e t e m o z e i P l a i t n e r e f f i D
9 8 7 6 5 y = 1.3x 1.3x - 0.5 R² = 0.7972
4 3 2 1 0 0
1
2
3
4
5
6
7
8
v2/2g x 10-6 (mm)
The graph above shows the differential of piezometer head increases as the value of
v2/2g increases. Therefore, Δ ′ is increasing linearly with the v 2/2g. From the slope of the graph, we obtained the loss coefficient which was the gradient k = 1.3. As for the result, there were error in plotting the graph. As we can observe the slope slightly change at v 2/2g = 3. This shouldn’t happen as the graph should be linearly increase.
For better accuracy result, long sections of the pipe are require to establish with certainty each position of the point of the linear sections of piezometer line. A sufficient distance to avoid severe disturbance because in this experiment only use 2 piezometer as the equipment already compact.
CONCLUSION
Theoretically Theoretically state that flowrate that is quantity of a gas or liquid moving through a pipe which passes per unit time; usually represented by the symbol Q. The SI unit is m3/s (cubic metres per second). In this test, flowrate was measured by using orifice meter, venturi meter and the rotameter. This measuring technique operation and characteristics are to be determined by comparing pressure drop that will be calculated that related to the velocity of the fluid in the pipe using the Bernoulli and Continuity equations. The data recorded show that orificemeter has higher pressure drop and cannot be recovered due to flow rates increases at the opening of the orifice plate and the energy lost are not much. For the rotameter the energy value that loss higher than venture and orifice, due to large drop in pressure due to friction. For venturi meter, the value are closer to the actual flow rate due to the lower pressure drop that came from its shape which are streamlined and do not have boundary- layer separation and thus form drag
is assumed negligible. negligible. Although Although it has a converging converging and
diverging and diverging part which we know will cause pressure loss at the converging part, but it properly designed until some percentage percentage of pressure loss is covered back at at the diverging part. This is good when handling high pressure and energy recovery. To conclude venturi meter was choosen as the more accurate then rotameter and orifice meter.
RECOMMENDATION
For recommendation, as student based on the experiment there are a few ways of improving way to increase the level of safety among the student. For the safety precaution and maintenance, It is important to drain all water from the apparatus when not in use. The apparatus should be stored properly to prevent damage. Any manometer tube, which does not fill with water or slow fill, f ill, indicates that tapping or connection of the manometer is blocked. To remove the obstacle, disconnect the flexible connection tube and blow through. The apparatus should not be exposed to any shock and stresses. Always wear protective clothing, shoes, helmet and goggles throughout the laboratory session. Always run the experiment after fully understand the unit and procedures. Next, to increase the efficiency of the experiment data make sure to remove the air bubbles to increase the accuracy and avoid reading error. Depress the stadle valve at the top right side of manometer board. Depress staddle staddle valve ligthtly to allow fluid and trapped air to escape out. Allow enough time for bleeding to allowing all bubbles escape. Besides that, when taking readings from the manometer, make sure the eye level perpendicular to eyes to avoid parallax error. Removing parallax error to get the accurate data and result. Addition the water level in the manometer board must be monitor which if the level rise to high until out of visible point, adjust the water level by using the staddle valve. With the maximum measurable flow rate, retain the maximum reading on manometer . Lastly, to increase the accuracy, the experiment should be conducted at least three times to get average. This will reduce the deviation data from the result.
REFFERENCE
Web
1) http://hyperphysics.phy-astr.g http://hyperphysics.phy-astr.gsu.edu/hba su.edu/hbase/pber.html se/pber.html 2) http://www.omega.com/ http://www.omega.com/prodinfo/flow prodinfo/flowmeters.html meters.html 3)http://www2.emersonprocess 3)http://www2.emersonprocess.com/en-us/bra .com/en-us/brands/rosemount/ nds/rosemount/flow/dp-flowflow/dp-flowproducts/compact-orifice-flowme products/compact-orifice-flowmeters/pages/ind ters/pages/index.aspx ex.aspx 4) http://www.lmnoeng.c http://www.lmnoeng.com/venturi.php om/venturi.php 5) http://www.lmnoeng.c http://www.lmnoeng.com/orifice.ph om/orifice.php p 6) https://www.scribd. https://www.scribd.com/doc/9625975 com/doc/96259752/SOLTEQ-Flowm 2/SOLTEQ-Flowmeter-Measurement-Appara eter-Measurement-Apparatus tus
APPENDIX